CIPM MRA COMPARISON REPORTS

SUPPLEMENTARY COMPARISON

SIM.AUV.V-S1

Secondary calibration of accelerometer sensitivity magnitude following ISO 16063-21:2003
M.A. González Durán , A. Ruiz Rueda , J.P. Ayala Breña , D. Nordmann , L. Klaus , T. Bruns
Published on 06/11/2026 - © 2026 M.A. González Durán et al
Citation M.A. González Durán et al 2026 CIPM MRA Comparison reports 09002

Abstract

The results of the SIM supplementary comparison in the secondary vibration calibration of the complex charge sensitivity (magnitude) concerning steady-state sinusoidal vibration are presented. This regional supplementary comparison followed the global CCAUV.V-K5 comparison over the frequency range from 10  Hz to 10  kHz. Measurement results reported by the two participants for magnitude of the complex charge sensitivity of back-to-back and single ended accelerometers were compared. A link to the CCAUV.V-K5 with the respective degrees of equivalence are reported.

Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.

The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).