CIPM MRA COMPARISON REPORTS

SUPPLEMENTARY COMPARISON

EURAMET.M.FF-S16

Calibration of piston operated volumetric instruments
L. Mićić , E. Batista , M. Benkova , U. Turnšek , Ü. Akçadağ , E. Borovac , C. Vámossy , E. Malejczyk , E. Smits , L.L. Grue ... , P. Wennergren , A.A. Al-shahrani , J. Mudreša , I. Florin
Published on 31/08/2026 - © 2026 L. Mićić et al
Citation L. Mićić et al 2026 CIPM MRA Comparison reports 07009

Abstract

This report summarizes the results of the measurements performed as a Supplementary Comparison (EURAMET.M.FF-S16) on piston - operated volumetric instruments among DMDM, IPQ, UME, IMBiH, MIRS, BFKH, CMI, GUM, FORCE, RISE, VSL, SASO-NMCC, BMM and INM. The comparison measurements have been performed between January 2022 and November 2023. In this comparison, "delivered" volumes of the micropipette in range (0,2 - 10) µL, 50 mL piston burette and 1 mL syringe were measured.

Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.

The final report has been peer-reviewed and approved for publication by the CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).