CIPM MRA COMPARISON REPORTS

KEY COMPARISON

CCT-K9.3

Realizations of the ITS-90 from 83.805 8 K to 692.677 K
R. Caballero , E. Gómez , M. Quezada , I. Chelidze , M. Zirakadze , L.F. Chaves , E. Ejigu
Published on 28/09/2026 - © 2026 R. Caballero et al
Citation R. Caballero et al 2026 CIPM MRA Comparison reports 03006

Abstract

This report shows the results of a  comparison of temperature standards between five national metrology institutes (CEM, CESMEC, GEOSTM, LACOMET and NMISA) in a schema were the pilot, CEM, is acting as the link to the CCT-K9 key comparison reference value, via CEM's participation in EURAMET.T-K9.

The measurands of this comparison were the W(t90) ratios of standard platinum resistance thermometers, used as transfer standards, measured at the zinc freezing point (Zn FP, 419.527 °C), tin freezing point (Sn FP, 231.928 °C), gallium melting point (Ga MP, 29.7646 °C), mercury triple point (Hg TP, -38.8344 °C) and argon triple point (Ar TP, -189.344 2 °C).

Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.

The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).