CIPM MRA COMPARISON REPORTS

KEY COMPARISON

CCQM-K173.2

Assay of sodium carbonate
M. Máriássy , Z. Hanková , P.P. Borges , S. Pereira Sobral , D. Nagy , Z. Nagyné Szilágyi , T. Orwa , S. Kachila
Published on 05/01/2026 - © 2026 M. Máriássy et al
Citation M. Máriássy et al 2026 CIPM MRA Comparison reports 08002

Abstract

The CCQM key comparison CCQM-K173.2 is a subsequent comparison to CCQM-K173. It was based on a request of INMETRO for a bilateral comparison; KEBS and BFKH joined also to demonstrate their capability to measure the amount content of bases. The institutes could use a method of their choice, although the use of coulometry or titrimetry with potentiometric determination of the endpoint was expected. The result of SMU was used as a link to the original comparison CCQM-K173.

The results of KEBS and BFKH were in good agreement with the reference value; a probable reason for the slight deviation of INMETRO’s result was suggested.

Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.

The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).