CIPM MRA COMPARISON REPORTS

KEY COMPARISON

BIPM.RI(I)-K1 (PTB 2024)

Measurement of air kerma for Cobalt 60
C. Kessler , S. Pojtinger
Published on 20/02/2026 - © 2026 C. Kessler et al
Citation C. Kessler et al 2026 CIPM MRA Comparison reports 06004

Abstract

A new direct key comparison of the standards for air kerma of the Physikalisch-Technische Bundesanstalt (PTB), Germany, and the Bureau International des Poids et Mesures (BIPM) was carried out in the 60Co radiation beam of the BIPM in February 2024. The comparison result, evaluated as a ratio of the PTB and the BIPM standards for air kerma, is 1.0004 with a combined standard uncertainty of 2.5 parts in 103. The results for an indirect comparison made at the same time are consistent with the direct comparison result at the level of 3.4 parts in 103. The result of the direct comparison agrees with the direct comparison carried out in 2014 at the level of the combined standard uncertainty, when updated with the changes implemented to the standards at each laboratory. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.