CIPM MRA COMPARISON REPORTS
EURAMET.M.T-S6
Torque measurements
Published on 20/02/2026 - © 2026 B. Lefranc
et al
Citation B. Lefranc
et al 2026 CIPM MRA Comparison reports
Abstract
This report describes the supplementary comparison EURAMET.M.T-S6 for
torque value from 10 N·m to 500 N·m between PTB (Germany) and LNE
(France) in both clockwise and anticlockwise direction. In this
comparison exercise, five torque transducers were used as transfer
standards. The results of the transducers at each torque step, expressed
as relative deviations between the measurement results of LNE and PTB,
were combined using the generalized weighted mean. Two evaluations were
considered: the first with the relative deviations supposed uncorrelated
and the second with the relative deviations supposed correlated. Both
evaluations supports the uncertainty claims of LNE for this comparison.