CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3

DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz

Selected graphs of equivalence for eight of the frequency values

 

DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz

Selected graphs of equivalence for eight of the frequency values

 

 

CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3

DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz

Degrees of equivalence: Di and Ui

 

 

 

 

 

Results are presented under A4 printable format in Summary Results (PDF file).

 

DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz

Degrees of equivalence: Di and Ui

 

 

 

Results are presented under A4 printable format in Summary Results (PDF file).

 

APMP.AUV.V-K3.1
Metrology area, Sub-field Acoustics, Ultrasound, Vibration, Vibration
Description Vibration acceleration
Time of measurements 2018
Status Approved for equivalence
Final Reports of the comparisons
References

APMP.AUV.V-K3.1 Technical Protocol

Measurand Acceleration sensitivity:
Voltage sensitivity [mV /(m/s2]
Phase shift in °

Parameters Frequency
Transfer device Single-ended (SE) servo accelerometer (SA 704) with a signal conditioner (MSA-I), power supply for the conditioner.
Comparison type Key Comparison
Consultative Committee CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration)
Conducted by APMP (Asia Pacific Metrology Programme)
Pilot institute CMS
ITRI Center for Measurement Standards
Chinese Taipei
Contact person Tsung-Hsien TU

+886-574-3791
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
CMS

ITRI Center for Measurement Standards, Chinese Taipei, APMP

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

NIM

National Institute of Metrology, China, APMP

NIMT

National Institute of Metrology (Thailand), Thailand, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

NPLI

CSIR National Physical Laboratory of India, India, APMP

CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3

DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz

 

CCAUV.V-K3

The computation of the key comparison reference value at each frequency, xR,M, and of its expanded uncertainty, UR,M (k = 2), is explained in Section 8 of the Final Report and is based on a weighted mean.

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR,M, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the Final Report.
Pair-wise degrees of equivalence may be calculated taking into account possible correlations.

 

Linking of APMP.AUV.V-K3 to CCAUV.V-K3

The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons. 

Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3

The linking was made via NMISA, having participated in both comparisons. 

Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3

The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. 

DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz

The computation of the key comparison reference value at each frequency, xR,P, and of its expanded uncertainty, UP,M (k = 2), is explained in Section 8 of the Final Report and is based on a weighted mean.

 

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR,P, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the Final Report.
Pair-wise degrees of equivalence may be calculated taking into account possible correlations.

CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3

DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz

 

CCAUV.V-K3

The laboratories' individual measurement may be found in Sections 7.1 and 7.2 of the CCAUV.V-K3 Final Report.

AFRIMETS.AUV.V-K3

The laboratories' individual measurement may be found in Section 7 of the AFRIMETS.AUV.V-K3 Final Report.

APMP.AUV.V-K3

The laboratories' individual measurement may be found in Section 9 of the APMP.AUV.V-K3 Final Report.

EURAMET.AUV.V-K3

The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.

 

DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz

The laboratories' individual measurement may be found in Sections 7.1 and 7.2 of the Final Report.