CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Selected graphs of equivalence for eight of the frequency values
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Selected graphs of equivalence for eight of the frequency values
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Degrees of equivalence: Di and Ui
Results are presented under A4 printable format in Summary Results (PDF file).
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Degrees of equivalence: Di and Ui
Results are presented under A4 printable format in Summary Results (PDF file).
Metrology area, Sub-field | Acoustics, Ultrasound, Vibration, Vibration |
Description | Calibration of accelerometers at low frequencies |
Time of measurements | 2012 - 2013 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
References | |
Measurand | Magnitude and phase of complex voltage sensitivity |
Parameters | Mandatory frequency range: 1 Hz to 80 Hz Covered frequencies (Hz) 0.1; 0.125; 0.16; 0.2; 0.25; 0.315; 0.4; 0.5; 0.63; 0.8; 1; 1.25; 1.6; 2; 2.5; 3.15; 4; 6.3; 8; 10; 12.5; 16; 20; 25; 31.5; 40; 63; 80; 100; 125; 160; 200 |
Transfer device | Accelerometer |
Comparison type | Key Comparison |
Consultative Committee | CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration) |
Conducted by | EURAMET (European Association of National Metrology Institutes) |
Comments | Results published on 24 June 2019 AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, and EURAMET.AUV.V-K3 results are linked to those of CCAUV.V-K3 Low-frequency acceleration, complex sensitivity
|
Pilot institute |
LNE
Laboratoire national de métrologie et d'essais France |
Contact person | Claire BARTOLI +33 1 30 69 13 76 |
Pilot laboratory | |
---|---|
LNE |
Laboratoire national de métrologie et d'essais, France, EURAMET |
CEM |
Centro Español de Metrologia, Spain, EURAMET |
CMI |
Czech Metrology Institute, Czechia, EURAMET |
GUM |
Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET |
HBK-DPLA |
Hottinger Brüel & Kjaer A/S - Danish Primary Laboratory for Acoustics Section, Denmark, EURAMET |
INRIM |
Istituto Nazionale di Ricerca Metrologica, Italy, EURAMET |
METAS |
Federal Institute of Metrology, Switzerland, EURAMET |
MIKES |
VTT Technical Research Centre of Finland Ltd, Centre for Metrology/Mittatekniikan keskus, Finland, EURAMET |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
SP |
Technical Research Institute of Sweden from 2017 Research Institutes of Sweden AB, Sweden, EURAMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR,M, and of its expanded uncertainty, UR,M (k = 2), is explained in Section 8 of the Final Report and is based on a weighted mean.
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR,M, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the Final Report. |
Pair-wise degrees of equivalence may be calculated taking into account possible correlations. |
Linking of APMP.AUV.V-K3 to CCAUV.V-K3
The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA, having participated in both comparisons.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons.
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
The computation of the key comparison reference value at each frequency, xR,P, and of its expanded uncertainty, UP,M (k = 2), is explained in Section 8 of the Final Report and is based on a weighted mean.
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR,P, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the Final Report. |
Pair-wise degrees of equivalence may be calculated taking into account possible correlations. |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Sections 7.1 and 7.2 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the AFRIMETS.AUV.V-K3 Final Report.
APMP.AUV.V-K3
The laboratories' individual measurement may be found in Section 9 of the APMP.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
The laboratories' individual measurement may be found in Sections 7.1 and 7.2 of the Final Report.