CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Selected graphs of equivalence for eight of the frequency values
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Selected graphs of equivalence for eight of the frequency values
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Degrees of equivalence: Di and Ui
Results are presented under A4 printable format in Summary Results (PDF file).
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Degrees of equivalence: Di and Ui
Results are presented under A4 printable format in Summary Results (PDF file).
Metrology area, Sub-field | Acoustics, Ultrasound, Vibration, Vibration |
Description | Vibration acceleration |
Time of measurements | 2010 - 2011 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
References | |
Measurand | Acceleration sensitivity Norminal value: 136 mV/(m/s2) |
Transfer device | Two quartz-flexure servo accelerometers of single-ended type and a signal conditioner |
Comparison type | Key Comparison |
Consultative Committee | CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration) |
Conducted by | APMP (Asia Pacific Metrology Programme) |
Comments | Key comparison APMP.AUV.V-K3 was first declared as APMP.AUV.V-S1, then upgraded at the 9th meeting of the CCAUV, 2013, as a "V-K3" key comparison. Results published on 07 December 2016 Updated on 24 June 2019 AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, and EURAMET.AUV.V-K3 results are linked to those of CCAUV.V-K3 Low-frequency acceleration, complex sensitivity
|
Pilot institute |
NIM
National Institute of Metrology China |
Contact person | Qiao SUN +86 10 645 24 623 |
Pilot laboratory | |
---|---|
NIM |
National Institute of Metrology, China, APMP |
CMS |
ITRI Center for Measurement Standards, Chinese Taipei, APMP |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
NIMT |
National Institute of Metrology (Thailand), Thailand, APMP |
NMIA |
National Measurement Institute, Australia, Australia, APMP |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
NMISA |
National Metrology Institute of South Africa, South Africa, AFRIMETS |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR,M, and of its expanded uncertainty, UR,M (k = 2), is explained in Section 8 of the Final Report and is based on a weighted mean.
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR,M, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the Final Report. |
Pair-wise degrees of equivalence may be calculated taking into account possible correlations. |
Linking of APMP.AUV.V-K3 to CCAUV.V-K3
The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA, having participated in both comparisons.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons.
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
The computation of the key comparison reference value at each frequency, xR,P, and of its expanded uncertainty, UP,M (k = 2), is explained in Section 8 of the Final Report and is based on a weighted mean.
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR,P, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the Final Report. |
Pair-wise degrees of equivalence may be calculated taking into account possible correlations. |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3 and EURAMET.AUV.V-K3
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Sections 7.1 and 7.2 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the AFRIMETS.AUV.V-K3 Final Report.
APMP.AUV.V-K3
The laboratories' individual measurement may be found in Section 9 of the APMP.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
The laboratories' individual measurement may be found in Sections 7.1 and 7.2 of the Final Report.