CCAUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.1 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.2 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.4 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.8 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 1.6 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 3.15 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 6.3 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 12.5 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 25 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 40 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Selected graphs of equivalence are available for eight of the frequency values, namely 0.1 Hz, 0.2 Hz, 0.4 Hz, 0.8 Hz, 1.6 Hz, 3.15 Hz, 6.3 Hz, 12.5 Hz, 25 Hz and 40 Hz
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.1 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.2 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.4 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
Enlarged Graph
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.8 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 1.6 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 3.15 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
Enlarged graph
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 6.3 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
Enlarged graph
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 12.5 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
Enlarged graph
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 25 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
Enalrged graph
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 40 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
Enlarged graph
CCAUV.V-K3 and APMP.AUV.V-K3.1
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Selected graphs of equivalence are available for eight of the frequency values, namely 0.1 Hz, 0.2 Hz, 0.4 Hz, 0.8 Hz, 1.6 Hz, 3.15 Hz, 6.3 Hz, 12.5 Hz, 25 Hz and 40 Hz
CCAUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.1 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | -0.62 | 0.66 | ||
PTB | -0.20 | 0.67 | ||
BKSV-DPLA | 0.34 | 0.67 | ||
NMISA | 4.21 | 0.86 | ||
CENAM | 0.22 | 0.95 | ||
NMIA | 0.28 | 0.23 | ||
NMIJ | -0.10 | 0.14 | ||
A*STAR | -0.45 | 0.75 | ||
CMI | 2.73 | 2.42 | ||
CMS | -0.74 | 0.51 | ||
KRISS | 0.64 | 1.9 | ||
NMISA | -0.53 | 1.2 | ||
NPLI | -0.57 | 1.5 |
CCAUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.2 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | -0.23 | 0.67 | ||
PTB | -0.15 | 0.39 | ||
BKSV-DPLA | 0.00 | 0.53 | ||
GUM | 1.31 | 1.24 | ||
METAS | -0.20 | 0.41 | ||
NMISA | 0.99 | 0.82 | ||
INMETRO | 0.12 | 0.32 | ||
CENAM | 0.09 | 0.67 | ||
NMIA | 0.04 | 0.25 | ||
NMIJ | -0.07 | 0.17 | ||
A*STAR | 0.05 | 0.53 | ||
CMI | 0.77 | 0.75 | ||
CMS | -0.44 | 0.51 | ||
KRISS | 0.10 | 0.75 | ||
NMISA | -0.12 | 1.2 | ||
NPLI | -0.20 | 1.5 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.4 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | -0.07 | 0.26 | ||
LNE | 0.20 | 0.40 | ||
PTB | -0.11 | 0.26 | ||
BKSV-DPLA | -0.01 | 0.54 | ||
GUM | 0.40 | 0.68 | ||
METAS | 0.02 | 0.42 | ||
NMISA | 0.22 | 0.82 | ||
INMETRO | -0.01 | 0.33 | ||
CENAM | -0.01 | 0.68 | ||
NMIA | 0.03 | 0.26 | ||
NMIJ | -0.01 | 0.19 | ||
A*STAR | 0.05 | 0.36 | ||
NMISA | -0.08 | 0.21 | ||
INMETRO | 0.08 | 0.21 | ||
CMI | 0.41 | 0.59 | ||
RISE | -0.77 | 1.51 | ||
CEM | 0.02 | 0.72 | ||
CMS | -0.28 | 0.43 | ||
KRISS | -0.01 | 0.43 | ||
NMISA | -0.02 | 0.70 | ||
NPLI | -0.09 | 1.4 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.8 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | 0.04 | 0.26 | ||
LNE | 0.04 | 0.40 | ||
PTB | -0.08 | 0.26 | ||
BKSV-DPLA | 0.01 | 0.54 | ||
GUM | 0.17 | 0.68 | ||
METAS | 0.07 | 0.42 | ||
NMISA | -0.06 | 0.40 | ||
INMETRO | -0.03 | 0.33 | ||
CENAM | -0.26 | 0.40 | ||
NMIA | 0.01 | 0.26 | ||
NMIJ | -0.01 | 0.19 | ||
KRISS | 0.01 | 0.11 | ||
A*STAR | 0.06 | 0.36 | ||
NMISA | -0.05 | 0.21 | ||
INMETRO | 0.05 | 0.21 | ||
NIM | 0.14 | 0.22 | ||
NMIA | -0.05 | 0.22 | ||
NMISA | -0.21 | 0.38 | ||
KRISS | 0.04 | 0.38 | ||
CMS | 0.89 | 3.30 | ||
NMIJ | -0.27 | 1.08 | ||
CMI | 0.19 | 0.58 | ||
RISE | -0.14 | 0.58 | ||
INRiM | -0.44 | 0.58 | ||
CEM | 0.00 | 0.71 | ||
CMS | 0.00 | 0.43 | ||
KRISS | 0.14 | 0.57 | ||
NMISA | 0.02 | 0.70 | ||
NPLI | 0.18 | 1.0 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 1.6 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | 0.02 | 0.26 | ||
LNE | 0.05 | 0.40 | ||
PTB | -0.06 | 0.26 | ||
BKSV-DPLA | 0.00 | 0.54 | ||
GUM | 0.13 | 0.54 | ||
METAS | 0.08 | 0.40 | ||
NMISA | -0.04 | 0.40 | ||
INMETRO | -0.05 | 0.33 | ||
CENAM | -0.16 | 0.40 | ||
NMIA | 0.02 | 0.26 | ||
NMIJ | 0.00 | 0.19 | ||
KRISS | 0.00 | 0.11 | ||
A*STAR | 0.05 | 0.39 | ||
NMISA | -0.04 | 0.21 | ||
INMETRO | 0.04 | 0.21 | ||
NIM | 0.08 | 0.22 | ||
NMIA | -0.02 | 0.22 | ||
NMISA | -0.04 | 0.38 | ||
KRISS | 0.03 | 0.52 | ||
CMS | 1.39 | 3.18 | ||
NMIJ | -0.27 | 0.66 | ||
CMI | 0.22 | 0.58 | ||
RISE | -0.06 | 0.45 | ||
INRiM | -0.30 | 0.57 | ||
CEM | -0.03 | 0.71 | ||
MIKES | -0.08 | 0.71 | ||
CMS | -0.02 | 0.43 | ||
KRISS | 0.13 | 0.43 | ||
NMISA | 0.05 | 0.43 | ||
NPLI | 0.19 | 1.0 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 3.15 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | 0.03 | 0.26 | ||
LNE | 0.03 | 0.40 | ||
PTB | -0.08 | 0.26 | ||
BKSV-DPLA | 0.00 | 0.54 | ||
GUM | 0.11 | 0.54 | ||
METAS | 0.09 | 0.39 | ||
NMISA | -0.06 | 0.40 | ||
INMETRO | -0.04 | 0.33 | ||
CENAM | -0.10 | 0.40 | ||
NMIA | 0.01 | 0.26 | ||
NMIJ | 0.00 | 0.19 | ||
KRISS | 0.02 | 0.11 | ||
A*STAR | 0.07 | 0.40 | ||
NMISA | -0.03 | 0.21 | ||
INMETRO | 0.03 | 0.21 | ||
NIM | 0.10 | 0.22 | ||
NMIA | -0.02 | 0.22 | ||
NMISA | -0.14 | 0.38 | ||
KRISS | -0.02 | 0.38 | ||
CMS | 0.41 | 0.70 | ||
CMI | 0.18 | 0.58 | ||
RISE | 0.01 | 0.45 | ||
INRiM | -0.21 | 0.57 | ||
CEM | -0.02 | 0.71 | ||
MIKES | 0.01 | 0.71 | ||
CMS | -0.06 | 0.43 | ||
KRISS | -0.01 | 0.43 | ||
NMISA | 0.02 | 0.43 | ||
NPLI | 0.17 | 1.0 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 6.3 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | -0.02 | 0.26 | ||
LNE | 0.00 | 0.40 | ||
PTB | -0.13 | 0.26 | ||
BKSV-DPLA | -0.04 | 0.54 | ||
GUM | 0.10 | 0.40 | ||
METAS | 0.04 | 0.32 | ||
NMISA | -0.08 | 0.40 | ||
INMETRO | -0.07 | 0.33 | ||
CENAM | -0.10 | 0.40 | ||
NMIA | -0.02 | 0.26 | ||
NMIJ | -0.03 | 0.19 | ||
KRISS | 0.06 | 0.11 | ||
A*STAR | 0.08 | 0.42 | ||
NMISA | -0.01 | 0.21 | ||
INMETRO | 0.01 | 0.21 | ||
NIM | 0.10 | 0.22 | ||
NMIA | -0.03 | 0.22 | ||
NMISA | -0.06 | 0.38 | ||
KRISS | -0.09 | 0.38 | ||
CMS | 0.60 | 0.70 | ||
CMI | 0.17 | 0.57 | ||
RISE | 0.04 | 0.45 | ||
INRiM | -0.26 | 0.57 | ||
CEM | -0.02 | 0.70 | ||
MIKES | 0.01 | 0.70 | ||
CMS | -0.06 | 0.43 | ||
KRISS | -0.01 | 0.43 | ||
NMISA | -0.01 | 0.43 | ||
NPLI | 0.16 | 1.0 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 12.5 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | 0.07 | 0.26 | ||
LNE | 0.05 | 0.40 | ||
PTB | -0.07 | 0.12 | ||
BKSV-DPLA | 0.04 | 0.54 | ||
GUM | 0.14 | 0.40 | ||
METAS | 0.08 | 0.50 | ||
NMISA | 0.04 | 0.40 | ||
INMETRO | -0.06 | 0.33 | ||
CENAM | -0.07 | 0.40 | ||
NMIA | 0.05 | 0.26 | ||
NMIJ | 0.04 | 0.19 | ||
KRISS | -0.01 | 0.12 | ||
A*STAR | 0.13 | 0.43 | ||
NMISA | 0.02 | 0.21 | ||
INMETRO | -0.02 | 0.21 | ||
NIM | 0.10 | 0.22 | ||
NMIA | -0.05 | 0.22 | ||
NMISA | -0.06 | 0.37 | ||
KRISS | -0.12 | 0.52 | ||
CMS | 0.71 | 0.57 | ||
CMI | 0.25 | 0.56 | ||
RISE | 0.14 | 0.43 | ||
INRiM | -0.19 | 0.56 | ||
CEM | 0.03 | 0.69 | ||
MIKES | 0.03 | 0.69 | ||
CMS | 0.03 | 0.43 | ||
KRISS | 0.08 | 0.43 | ||
NMISA | 0.01 | 0.43 | ||
NPLI | 0.23 | 1.0 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 25 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | -0.10 | 0.26 | ||
LNE | 0.01 | 0.40 | ||
PTB | 0.01 | 0.11 | ||
BKSV-DPLA | -0.04 | 0.68 | ||
GUM | 0.13 | 0.40 | ||
METAS | 0.03 | 0.49 | ||
NMISA | 0.07 | 0.40 | ||
INMETRO | -0.07 | 0.33 | ||
CENAM | -0.08 | 0.40 | ||
NMIA | 0.02 | 0.36 | ||
NMIJ | 0.02 | 0.19 | ||
KRISS | -0.38 | 0.16 | ||
A*STAR | 0.07 | 0.42 | ||
NMISA | 0.04 | 0.21 | ||
INMETRO | -0.04 | 0.21 | ||
CMI | 0.31 | 0.56 | ||
RISE | 0.09 | 0.43 | ||
INRiM | -0.15 | 0.56 | ||
CEM | 0.04 | 0.70 | ||
MIKES | 0.01 | 0.70 | ||
CMS | -0.19 | 0.43 | ||
KRISS | -0.07 | 0.43 | ||
NMISA | -0.36 | 0.43 | ||
NPLI | -0.05 | 1.2 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 40 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(mV/(m/s2)) |
(mV/(m/s2)) |
(mV/(m/s2)) |
||
NIM | 0.09 | 0.27 | ||
LNE | 0.02 | 0.41 | ||
PTB | 0.16 | 0.12 | ||
BKSV-DPLA | 0.46 | 0.69 | ||
GUM | 0.12 | 0.41 | ||
METAS | 0.02 | 0.49 | ||
NMISA | 0.12 | 0.41 | ||
INMETRO | -0.16 | 0.34 | ||
CENAM | -0.11 | 0.41 | ||
NMIA | 0.02 | 0.27 | ||
NMIJ | 0.00 | 0.27 | ||
KRISS | -0.22 | 0.12 | ||
A*STAR | 0.12 | 0.40 | ||
NMISA | 0.12 | 0.22 | ||
INMETRO | -0.12 | 0.21 | ||
CMI | 0.25 | 0.57 | ||
RISE | 0.03 | 0.43 | ||
INRiM | -0.11 | 0.57 | ||
CEM | 0.03 | 0.70 | ||
MIKES | -0.08 | 0.70 | ||
CMS | 0.07 | 0.43 | ||
KRISS | -0.09 | 0.43 | ||
NMISA | -0.40 | 0.43 | ||
NPLI | -0.02 | 1.3 |
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Degrees of equivalence: Di and Ui
Lab i | 0.1 Hz | 0.125 Hz | 0.16 Hz | 0.2 Hz | 0.25 Hz | 0.315 Hz | 0.4 Hz | 0.5 Hz | |||||||||||||||
Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | ||||||||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | ||||||||||||||||
NIM | -0.62 | 0.66 | -0.41 | 0.67 | -0.27 | 0.67 | -0.23 | 0.67 | -0.15 | 0.68 | -0.11 | 0.68 | -0.07 | 0.26 | 0.05 | 0.26 | |||||||
LNE | - | - | - | - | - | - | - | - | - | - | - | - | 0.20 | 0.40 | 0.10 | 0.40 | |||||||
PTB | -0.20 | 0.67 | -0.12 | 0.67 | -0.13 | 0.67 | -0.15 | 0.39 | -0.13 | 0.40 | -0.12 | 0.40 | -0.11 | 0.26 | -0.13 | 0.26 | |||||||
BKSV-DPLA | 0.34 | 0.67 | 0.13 | 0.67 | 0.06 | 0.67 | 0.00 | 0.53 | -0.02 | 0.54 | -0.01 | 0.54 | -0.01 | 0.54 | -0.03 | 0.54 | |||||||
GUM | - | - | - | - | - | - | 1.31 | 1.24 | 0.88 | 1.23 | 0.58 | 1.23 | 0.40 | 0.68 | 0.27 | 0.68 | |||||||
METAS | - | - | - | - | - | - | -0.20 | 0.41 | -0.10 | 0.41 | -0.02 | 0.41 | 0.02 | 0.42 | 0.03 | 0.42 | |||||||
NMISA | 4.21 | 0.86 | 2.53 | 0.84 | 1.64 | 0.84 | 0.99 | 0.82 | 0.63 | 0.82 | 0.37 | 0.82 | 0.22 | 0.82 | 0.00 | 0.40 | |||||||
INMETRO | - | - | - | - | - | - | 0.12 | 0.32 | 0.07 | 0.33 | 0.03 | 0.33 | -0.01 | 0.33 | -0.03 | 0.33 | |||||||
CENAM | 0.22 | 0.95 | 0.23 | 0.95 | 0.17 | 0.95 | 0.09 | 0.67 | 0.02 | 0.68 | -0.03 | 0.68 | -0.01 | 0.68 | -0.21 | 0.40 | |||||||
NMIA | 0.28 | 0.23 | 0.17 | 0.23 | 0.11 | 0.23 | 0.04 | 0.25 | 0.04 | 0.26 | 0.04 | 0.26 | 0.03 | 0.26 | -0.01 | 0.26 | |||||||
NMIJ | -0.10 | 0.14 | -0.06 | 0.14 | -0.04 | 0.14 | -0.07 | 0.17 | -0.04 | 0.18 | -0.02 | 0.18 | -0.01 | 0.19 | -0.04 | 0.19 | |||||||
KRISS | - | - | - | - | - | - | - | - | - | - | - | - | - | - | 0.03 | 0.11 | |||||||
A*STAR | -0.45 | 0.75 | -0.16 | 0.69 | 0.02 | 0.68 | 0.05 | 0.53 | 0.07 | 0.54 | 0.07 | 0.54 | 0.05 | 0.36 | 0.05 | 0.36 | |||||||
NMISA | - | - | - | - | - | - | - | - | - | - | - | - | -0.08 | 0.21 | -0.04 | 0.21 | |||||||
INMETRO | - | - | - | - | - | - | - | - | - | - | - | - | 0.08 | 0.21 | 0.04 | 0.21 | |||||||
NIM | - | - | - | - | - | - | - | - | - | - | - | - | - | - | 0.19 | 0.22 | |||||||
NMIA | - | - | - | - | - | - | - | - | - | - | - | - | - | - | -0.09 | 0.21 | |||||||
NMISA | - | - | - | - | - | - | - | - | - | - | - | - | - | - | -0.28 | 0.37 | |||||||
KRISS | - | - | - | - | - | - | - | - | - | - | - | - | - | - | 0.12 | 0.66 | |||||||
CMS | - | - | - | - | - | - | - | - | - | - | - | - | - | - | 0.59 | 3.71 | |||||||
NMIJ | - | - | - | - | - | - | - | - | - | - | - | - | - | - | 0.72 | 2.74 | |||||||
CMI | 2.73 | 2.42 | 1.79 | 1.60 | 1.34 | 1.21 | 0.77 | 0.75 | 0.61 | 0.62 | 0.45 | 0.61 | 0.41 | 0.59 | 0.25 | 0.58 | |||||||
RISE | - | - | - | - | - | - | - | - | - | - | -1.24 | 2.45 | -0.77 | 1.51 | -0.49 | 1.10 | |||||||
INRiM | - | - | - | - | - | - | - | - | - | - | - | - | - | - | -0.33 | 0.58 | |||||||
CEM | - | - | - | - | - | - | - | - | - | - | - | - | 0.02 | 0.72 | -0.03 | 0.71 | |||||||
MIKES | - | - | - | - | - | - | - | - | - | - | - | - | - | - | - | - | |||||||
CMS | -0.74 | 0.51 | -0.60 | 0.51 | -0.45 | 0.51 | -0.44 | 0.51 | -0.37 | 0.51 | -0.34 | 0.51 | -0.28 | 0.43 | 0.04 | 0.43 | |||||||
KRISS | 0.64 | 1.9 | 0.34 | 1.5 | 0.23 | 1.2 | 0.10 | 0.75 | 0.10 | 0.75 | 0.05 | 0.51 | -0.01 | 0.43 | 0.11 | 0.57 | |||||||
NMISA | -0.53 | 1.2 | -0.31 | 1.2 | -0.10 | 1.2 | -0.12 | 1.2 | -0.06 | 0.75 | -0.03 | 0.75 | -0.02 | 0.70 | 0.01 | 0.70 | |||||||
NPLI | -0.57 | 1.5 | -0.39 | 1.5 | -0.27 | 1.5 | -0.20 | 1.5 | -0.15 | 1.5 | -0.12 | 1.5 | -0.09 | 1.4 | 0.18 | 1.0 |
Degrees of equivalence: Di and Ui (continued)
Lab i | 0.63 Hz | 0.8 Hz | 1 Hz | 1.25 Hz | 1.6 Hz | 2 Hz | 2.50 Hz | 3.15 Hz | |||||||||||||||
Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | ||||||||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | ||||||||||||||||
NIM | 0.08 | 0.26 | 0.04 | 0.26 | 0.05 | 0.26 | 0.03 | 0.26 | 0.02 | 0.26 | -0.02 | 0.26 | 0.02 | 0.26 | 0.03 | 0.26 | |||||||
LNE | 0.12 | 0.40 | 0.04 | 0.40 | 0.02 | 0.40 | 0.05 | 0.40 | 0.05 | 0.40 | 0.04 | 0.40 | 0.03 | 0.40 | 0.03 | 0.40 | |||||||
PTB | -0.10 | 0.26 | -0.08 | 0.26 | -0.06 | 0.26 | -0.06 | 0.26 | -0.06 | 0.26 | -0.05 | 0.26 | -0.08 | 0.26 | -0.08 | 0.26 | |||||||
BKSV-DPLA | 0.01 | 0.54 | 0.01 | 0.54 | -0.02 | 0.54 | 0.00 | 0.54 | 0.00 | 0.54 | -0.01 | 0.54 | -0.02 | 0.54 | 0.00 | 0.54 | |||||||
GUM | 0.24 | 0.68 | 0.17 | 0.68 | 0.16 | 0.54 | 0.14 | 0.54 | 0.13 | 0.54 | 0.13 | 0.54 | 0.12 | 0.54 | 0.11 | 0.54 | |||||||
METAS | 0.07 | 0.42 | 0.07 | 0.42 | 0.09 | 0.40 | 0.08 | 0.40 | 0.08 | 0.40 | 0.10 | 0.39 | 0.08 | 0.39 | 0.09 | 0.39 | |||||||
NMISA | -0.04 | 0.40 | -0.06 | 0.40 | -0.04 | 0.40 | -0.03 | 0.40 | -0.04 | 0.40 | -0.03 | 0.40 | -0.05 | 0.40 | -0.06 | 0.40 | |||||||
INMETRO | 0.00 | 0.33 | -0.03 | 0.33 | -0.03 | 0.33 | -0.04 | 0.33 | -0.05 | 0.33 | -0.05 | 0.33 | -0.05 | 0.33 | -0.04 | 0.33 | |||||||
CENAM | -0.26 | 0.40 | -0.26 | 0.40 | -0.19 | 0.40 | -0.15 | 0.40 | -0.16 | 0.40 | -0.12 | 0.40 | -0.12 | 0.40 | -0.10 | 0.40 | |||||||
NMIA | 0.00 | 0.26 | 0.01 | 0.26 | 0.01 | 0.26 | 0.02 | 0.26 | 0.02 | 0.26 | 0.02 | 0.26 | 0.01 | 0.26 | 0.01 | 0.26 | |||||||
NMIJ | -0.02 | 0.19 | -0.01 | 0.19 | 0.00 | 0.19 | 0.00 | 0.19 | 0.00 | 0.19 | 0.00 | 0.19 | 0.00 | 0.19 | 0.00 | 0.19 | |||||||
KRISS | 0.01 | 0.11 | 0.01 | 0.11 | 0.01 | 0.11 | 0.01 | 0.11 | 0.00 | 0.11 | 0.01 | 0.11 | 0.00 | 0.11 | 0.02 | 0.11 | |||||||
A*STAR | 0.11 | 0.36 | 0.06 | 0.36 | 0.04 | 0.39 | 0.06 | 0.39 | 0.05 | 0.39 | 0.06 | 0.39 | 0.06 | 0.40 | 0.07 | 0.40 | |||||||
NMISA | -0.07 | 0.21 | -0.05 | 0.21 | -0.01 | 0.21 | -0.03 | 0.21 | -0.04 | 0.21 | -0.03 | 0.21 | -0.03 | 0.21 | -0.03 | 0.21 | |||||||
INMETRO | 0.07 | 0.21 | 0.05 | 0.21 | 0.01 | 0.21 | 0.03 | 0.21 | 0.04 | 0.21 | 0.03 | 0.21 | 0.03 | 0.21 | 0.03 | 0.21 | |||||||
NIM | 0.15 | 0.22 | 0.14 | 0.22 | 0.07 | 0.22 | 0.07 | 0.22 | 0.08 | 0.22 | 0.06 | 0.22 | 0.06 | 0.22 | 0.10 | 0.22 | |||||||
NMIA | -0.05 | 0.22 | -0.05 | 0.22 | -0.04 | 0.22 | -0.03 | 0.22 | -0.02 | 0.22 | -0.06 | 0.22 | -0.06 | 0.22 | -0.02 | 0.22 | |||||||
NMISA | -0.25 | 0.37 | -0.21 | 0.38 | -0.08 | 0.37 | -0.06 | 0.38 | -0.04 | 0.38 | -0.08 | 0.38 | -0.07 | 0.38 | -0.14 | 0.38 | |||||||
KRISS | 0.12 | 0.66 | 0.04 | 0.38 | 0.06 | 0.66 | 0.01 | 0.38 | 0.03 | 0.52 | 0.10 | 0.38 | 0.06 | 0.38 | -0.02 | 0.38 | |||||||
CMS | 0.97 | 3.72 | 0.89 | 3.30 | 1.35 | 3.18 | 1.04 | 3.17 | 1.39 | 3.18 | 0.69 | 3.16 | 0.37 | 3.02 | 0.41 | 0.70 | |||||||
NMIJ | -0.21 | 1.08 | -0.27 | 1.08 | -0.18 | 1.22 | -0.32 | 0.94 | -0.27 | 0.66 | -0.21 | 0.94 | - | - | - | - | |||||||
CMI | 0.20 | 0.58 | 0.19 | 0.58 | 0.16 | 0.58 | 0.22 | 0.58 | 0.22 | 0.58 | 0.16 | 0.58 | 0.20 | 0.58 | 0.18 | 0.58 | |||||||
RISE | -0.26 | 0.71 | -0.14 | 0.58 | -0.11 | 0.45 | -0.08 | 0.45 | -0.06 | 0.45 | -0.03 | 0.45 | 0.04 | 0.45 | 0.01 | 0.45 | |||||||
INRiM | - | - | -0.44 | 0.58 | -0.57 | 0.57 | -0.44 | 0.57 | -0.30 | 0.57 | -0.22 | 0.58 | -0.21 | 0.57 | -0.21 | 0.57 | |||||||
CEM | -0.02 | 0.72 | 0.00 | 0.71 | -0.03 | 0.71 | -0.03 | 0.71 | -0.03 | 0.71 | -0.03 | 0.71 | -0.02 | 0.71 | -0.02 | 0.71 | |||||||
MIKES | - | - | - | - | -0.11 | 0.71 | -0.08 | 0.71 | -0.08 | 0.71 | -0.11 | 0.71 | -0.07 | 0.71 | 0.01 | 0.71 | |||||||
CMS | 0.00 | 0.43 | 0.00 | 0.43 | -0.01 | 0.43 | -0.03 | 0.43 | -0.02 | 0.43 | -0.07 | 0.43 | -0.04 | 0.43 | -0.06 | 0.43 | |||||||
KRISS | 0.07 | 0.57 | 0.14 | 0.57 | 0.19 | 0.57 | 0.18 | 0.57 | 0.13 | 0.43 | -0.03 | 0.43 | -0.01 | 0.43 | -0.01 | 0.43 | |||||||
NMISA | -0.02 | 0.70 | 0.02 | 0.70 | 0.03 | 0.43 | 0.04 | 0.43 | 0.05 | 0.43 | 0.00 | 0.43 | 0.03 | 0.43 | 0.02 | 0.43 | |||||||
NPLI | 0.15 | 1.0 | 0.18 | 1.0 | 0.12 | 1.0 | 0.19 | 1.0 | 0.19 | 1.0 | 0.09 | 1.0 | 0.17 | 1.0 | 0.17 | 1.0 |
Degrees of equivalence: Di and Ui (continued)
Lab i | 4 Hz | 5 Hz | 6.3 Hz | 8 Hz | 10 Hz | 12.5 Hz | 16 Hz | 20 Hz | |||||||||||||||
Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | ||||||||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | ||||||||||||||||
NIM | 0.01 | 0.26 | -0.03 | 0.26 | -0.02 | 0.26 | 0.02 | 0.26 | 0.06 | 0.26 | 0.07 | 0.26 | 0.16 | 0.26 | -0.15 | 0.26 | |||||||
LNE | 0.03 | 0.40 | 0.01 | 0.40 | 0.00 | 0.40 | 0.03 | 0.40 | 0.06 | 0.40 | 0.05 | 0.40 | 0.08 | 0.40 | 0.04 | 0.40 | |||||||
PTB | -0.09 | 0.26 | -0.12 | 0.26 | -0.13 | 0.26 | -0.08 | 0.26 | -0.07 | 0.12 | -0.07 | 0.12 | -0.02 | 0.12 | -0.02 | 0.11 | |||||||
BKSV-DPLA | -0.02 | 0.54 | -0.02 | 0.54 | -0.04 | 0.54 | 0.01 | 0.54 | 0.04 | 0.54 | 0.04 | 0.54 | 0.07 | 0.54 | 0.04 | 0.54 | |||||||
GUM | 0.10 | 0.54 | 0.11 | 0.40 | 0.10 | 0.40 | 0.15 | 0.40 | 0.20 | 0.40 | 0.14 | 0.40 | 0.18 | 0.40 | 0.14 | 0.40 | |||||||
METAS | 0.09 | 0.39 | 0.07 | 0.32 | 0.04 | 0.32 | 0.10 | 0.32 | 0.14 | 0.50 | 0.08 | 0.50 | 0.12 | 0.50 | 0.06 | 0.50 | |||||||
NMISA | -0.03 | 0.40 | -0.07 | 0.40 | -0.08 | 0.40 | -0.02 | 0.40 | 0.02 | 0.40 | 0.04 | 0.40 | 0.05 | 0.40 | 0.07 | 0.40 | |||||||
INMETRO | -0.07 | 0.33 | -0.09 | 0.33 | -0.07 | 0.33 | -0.09 | 0.33 | -0.07 | 0.33 | -0.06 | 0.33 | -0.12 | 0.33 | -0.09 | 0.33 | |||||||
CENAM | -0.09 | 0.40 | -0.11 | 0.40 | -0.10 | 0.40 | -0.06 | 0.40 | -0.02 | 0.40 | -0.07 | 0.40 | 0.05 | 0.40 | -0.03 | 0.40 | |||||||
NMIA | 0.00 | 0.26 | -0.01 | 0.26 | -0.02 | 0.26 | 0.03 | 0.26 | 0.06 | 0.26 | 0.05 | 0.26 | 0.08 | 0.26 | 0.03 | 0.26 | |||||||
NMIJ | -0.01 | 0.19 | -0.01 | 0.19 | -0.03 | 0.19 | 0.01 | 0.19 | 0.04 | 0.19 | 0.04 | 0.19 | 0.07 | 0.19 | 0.03 | 0.19 | |||||||
KRISS | 0.02 | 0.11 | 0.04 | 0.11 | 0.06 | 0.11 | -0.03 | 0.11 | -0.03 | 0.12 | -0.01 | 0.12 | -0.12 | 0.12 | -0.35 | 0.16 | |||||||
A*STAR | 0.07 | 0.40 | 0.09 | 0.42 | 0.08 | 0.42 | 0.12 | 0.42 | 0.13 | 0.45 | 0.13 | 0.43 | 0.14 | 0.47 | 0.09 | 0.47 | |||||||
NMISA | -0.02 | 0.21 | -0.02 | 0.21 | -0.01 | 0.21 | 0.01 | 0.21 | 0.02 | 0.21 | 0.02 | 0.21 | 0.03 | 0.21 | 0.05 | 0.21 | |||||||
INMETRO | 0.02 | 0.21 | 0.02 | 0.21 | 0.01 | 0.21 | -0.01 | 0.21 | -0.02 | 0.21 | -0.02 | 0.21 | -0.03 | 0.21 | -0.05 | 0.21 | |||||||
NIM | 0.09 | 0.22 | 0.09 | 0.22 | 0.10 | 0.22 | 0.10 | 0.22 | 0.06 | 0.22 | 0.10 | 0.22 | 0.13 | 0.22 | 0.12 | 0.21 | |||||||
NMIA | -0.04 | 0.22 | -0.05 | 0.22 | -0.03 | 0.22 | -0.04 | 0.22 | 0.03 | 0.22 | -0.05 | 0.22 | -0.04 | 0.22 | -0.07 | 0.21 | |||||||
NMISA | -0.05 | 0.38 | -0.07 | 0.38 | -0.06 | 0.38 | -0.08 | 0.37 | -0.08 | 0.37 | -0.06 | 0.37 | -0.15 | 0.38 | -0.10 | 0.37 | |||||||
KRISS | -0.04 | 0.38 | -0.01 | 0.38 | -0.09 | 0.38 | -0.10 | 0.52 | -0.18 | 0.52 | -0.12 | 0.52 | -0.04 | 0.38 | -0.14 | 0.45 | |||||||
CMS | 0.83 | 0.71 | 0.37 | 0.70 | 0.60 | 0.70 | 0.59 | 0.71 | 0.14 | 0.70 | 0.71 | 0.57 | 0.84 | 0.71 | 0.56 | 0.71 | |||||||
NMIJ | - | - | - | - | - | - | - | - | - | - | - | - | - | - | - | - | |||||||
CMI | 0.16 | 0.58 | 0.15 | 0.57 | 0.17 | 0.57 | 0.22 | 0.57 | 0.25 | 0.56 | 0.25 | 0.56 | 0.24 | 0.56 | 0.26 | 0.56 | |||||||
RISE | 0.02 | 0.45 | 0.02 | 0.45 | 0.04 | 0.45 | 0.03 | 0.45 | 0.11 | 0.43 | 0.14 | 0.43 | 0.10 | 0.43 | 0.10 | 0.43 | |||||||
INRiM | -0.28 | 0.57 | -0.26 | 0.57 | -0.26 | 0.57 | -0.24 | 0.57 | -0.19 | 0.56 | -0.19 | 0.56 | -0.14 | 0.56 | -0.12 | 0.56 | |||||||
CEM | -0.03 | 0.71 | -0.01 | 0.70 | -0.02 | 0.70 | 0.01 | 0.70 | 0.06 | 0.69 | 0.03 | 0.69 | 0.05 | 0.69 | 0.04 | 0.70 | |||||||
MIKES | -0.03 | 0.71 | -0.01 | 0.70 | 0.01 | 0.70 | 0.01 | 0.70 | 0.06 | 0.69 | 0.03 | 0.69 | 0.02 | 0.69 | -0.01 | 0.70 | |||||||
CMS | -0.03 | 0.43 | -0.08 | 0.43 | -0.06 | 0.43 | -0.02 | 0.43 | 0.01 | 0.43 | 0.03 | 0.43 | 0.13 | 0.43 | -0.21 | 0.43 | |||||||
KRISS | -0.03 | 0.43 | -0.06 | 0.43 | -0.01 | 0.43 | 0.03 | 0.43 | 0.07 | 0.43 | 0.08 | 0.43 | 0.22 | 0.43 | -0.10 | 0.43 | |||||||
NMISA | 0.04 | 0.43 | 0.01 | 0.43 | -0.01 | 0.43 | -0.01 | 0.43 | 0.02 | 0.43 | 0.01 | 0.43 | 0.08 | 0.43 | -0.32 | 0.43 | |||||||
NPLI | 0.12 | 1.0 | 0.09 | 1.0 | 0.16 | 1.0 | 0.12 | 1.0 | 0.14 | 1.0 | 0.23 | 1.0 | 0.30 | 1.0 | -0.04 | 1.2 |
Degrees of equivalence: Di and Ui (continued)
Lab i | 25 Hz | 31.5 Hz | 40 Hz | |||||
Di | Ui | Di | Ui | Di | Ui | |||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | ||||||
NIM | -0.10 | 0.26 | 0.05 | 0.27 | 0.09 | 0.27 | ||
LNE | 0.01 | 0.40 | 0.06 | 0.41 | 0.02 | 0.41 | ||
PTB | 0.01 | 0.11 | 0.11 | 0.12 | 0.16 | 0.12 | ||
BKSV-DPLA | -0.04 | 0.68 | 0.03 | 0.69 | 0.46 | 0.69 | ||
GUM | 0.13 | 0.40 | 0.16 | 0.41 | 0.12 | 0.41 | ||
METAS | 0.03 | 0.49 | 0.10 | 0.49 | 0.02 | 0.49 | ||
NMISA | 0.07 | 0.40 | 0.09 | 0.41 | 0.12 | 0.41 | ||
INMETRO | -0.07 | 0.33 | -0.08 | 0.34 | -0.16 | 0.34 | ||
CENAM | -0.08 | 0.40 | -0.06 | 0.41 | -0.11 | 0.41 | ||
NMIA | 0.02 | 0.36 | 0.06 | 0.27 | 0.02 | 0.27 | ||
NMIJ | 0.02 | 0.19 | 0.00 | 0.27 | 0.00 | 0.27 | ||
KRISS | -0.38 | 0.16 | -0.18 | 0.12 | -0.22 | 0.12 | ||
A*STAR | 0.07 | 0.42 | 0.11 | 0.42 | 0.12 | 0.40 | ||
NMISA | 0.04 | 0.21 | 0.09 | 0.21 | 0.12 | 0.22 | ||
INMETRO | -0.04 | 0.21 | -0.09 | 0.21 | -0.12 | 0.21 | ||
CMI | 0.31 | 0.56 | 0.34 | 0.57 | 0.25 | 0.57 | ||
RISE | 0.09 | 0.43 | 0.10 | 0.43 | 0.03 | 0.43 | ||
INRiM | -0.15 | 0.56 | -0.12 | 0.56 | -0.11 | 0.57 | ||
CEM | 0.04 | 0.70 | 0.04 | 0.70 | 0.03 | 0.70 | ||
MIKES | 0.01 | 0.70 | -0.01 | 0.70 | -0.08 | 0.70 | ||
CMS | -0.19 | 0.43 | -0.03 | 0.43 | 0.07 | 0.43 | ||
KRISS | -0.07 | 0.43 | 0.06 | 0.43 | -0.09 | 0.43 | ||
NMISA | -0.36 | 0.43 | -0.29 | 0.43 | -0.40 | 0.43 | ||
NPLI | -0.05 | 1.2 | 0.13 | 1.3 | -0.02 | 1.3 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.1 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | -0.01 | 0.19 | ||
PTB | -0.06 | 0.19 | ||
BKSV-DPLA | -0.04 | 0.30 | ||
NMISA | -0.04 | 0.40 | ||
CENAM | 0.06 | 0.19 | ||
NMIA | -0.03 | 0.30 | ||
NMIJ | 0.00 | 0.02 | ||
A*STAR | 0.10 | 0.87 | ||
CMS | -0.02 | 0.31 | ||
KRISS | -0.01 | 0.21 | ||
NMISA | -0.03 | 0.21 | ||
NPLI | -0.04 | 1.01 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.2 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | -0.01 | 0.20 | ||
PTB | -0.12 | 0.20 | ||
BKSV-DPLA | -0.01 | 0.30 | ||
GUM | -0.10 | 0.90 | ||
METAS | 0.09 | 0.78 | ||
NMISA | -0.01 | 0.40 | ||
INMETRO | -0.01 | 0.25 | ||
CENAM | 0.25 | 0.20 | ||
NMIA | -0.01 | 0.30 | ||
NMIJ | -0.01 | 0.02 | ||
A*STAR | -0.02 | 0.51 | ||
CMS | -0.01 | 0.31 | ||
KRISS | 0.00 | 0.21 | ||
NMISA | -0.01 | 0.21 | ||
NPLI | 0.03 | 1.01 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.4 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | 0.00 | 20.00 | ||
LNE | 1.34 | 0.50 | ||
PTB | -0.12 | 0.20 | ||
BKSV-DPLA | 0.00 | 0.30 | ||
GUM | -0.06 | 0.50 | ||
METAS | 0.06 | 0.78 | ||
NMISA | 0.02 | 0.40 | ||
INMETRO | 0.01 | 0.25 | ||
CENAM | 0.30 | 0.50 | ||
NMIA | 0.01 | 0.30 | ||
NMIJ | 0.01 | 0.02 | ||
A*STAR | 0.01 | 0.31 | ||
CMS | 0.01 | 0.31 | ||
KRISS | 0.03 | 0.21 | ||
NMISA | 0.01 | 0.21 | ||
NPLI | 0.03 | 1.01 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.8 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | 0.05 | 0.20 | ||
LNE | 0.01 | 0.50 | ||
PTB | -0.05 | 0.20 | ||
BKSV-DPLA | 0.01 | 0.30 | ||
GUM | -0.03 | 0.50 | ||
METAS | 0.04 | 0.78 | ||
NMISA | 0.01 | 0.40 | ||
INMETRO | 0.01 | 0.25 | ||
CENAM | -0.52 | 0.50 | ||
NMIA | 0.01 | 0.30 | ||
NMIJ | 0.01 | 0.03 | ||
KRISS | -0.05 | 0.09 | ||
A*STAR | 0.01 | 0.31 | ||
CMS | 0.06 | 0.31 | ||
KRISS | 0.08 | 0.21 | ||
NMISA | 0.07 | 0.21 | ||
NPLI | -0.15 | 0.71 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 1.6 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | 0.02 | 0.20 | ||
PTB | -0.04 | 0.20 | ||
BKSV-DPLA | 0.02 | 0.30 | ||
GUM | 0.00 | 0.50 | ||
METAS | 0.04 | 0.46 | ||
NMISA | 0.02 | 0.40 | ||
INMETRO | 0.02 | 0.25 | ||
CENAM | -1.09 | 0.50 | ||
NMIA | 0.03 | 0.30 | ||
NMIJ | 0.02 | 0.03 | ||
KRISS | -0.09 | 0.09 | ||
A*STAR | 0.03 | 0.35 | ||
CMS | 0.03 | 0.31 | ||
KRISS | 0.04 | 0.21 | ||
NMISA | 0.06 | 0.21 | ||
NPLI | -0.08 | 0.71 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 3.15 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | 0.03 | 0.20 | ||
LNE | 0.01 | 0.50 | ||
PTB | -0.02 | 0.20 | ||
BKSV-DPLA | 0.03 | 0.30 | ||
GUM | 0.03 | 0.50 | ||
METAS | 0.05 | 0.46 | ||
NMISA | 0.04 | 0.40 | ||
INMETRO | 0.04 | 0.25 | ||
CENAM | -1.98 | 0.50 | ||
NMIA | 0.05 | 0.30 | ||
NMIJ | 0.03 | 0.03 | ||
KRISS | -0.16 | 0.09 | ||
A*STAR | 0.01 | 0.35 | ||
CMS | 0.04 | 0.33 | ||
KRISS | 0.04 | 0.21 | ||
NMISA | 0.11 | 0.21 | ||
NPLI | -0.06 | 0.71 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 6.3 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | 0.02 | 0.19 | ||
LNE | 0.00 | 0.50 | ||
PTB | -0.03 | 0.19 | ||
BKSV-DPLA | 0.02 | 0.29 | ||
GUM | 0.02 | 0.50 | ||
METAS | 0.02 | 0.45 | ||
NMISA | 0.02 | 0.39 | ||
INMETRO | 0.02 | 0.24 | ||
CENAM | -4.11 | 0.50 | ||
NMIA | 0.05 | 0.29 | ||
NMIJ | 0.00 | 0.07 | ||
KRISS | -0.35 | 0.12 | ||
A*STAR | -0.03 | 0.34 | ||
CMS | 0.05 | 0.33 | ||
KRISS | 0.03 | 0.22 | ||
NMISA | 0.16 | 0.22 | ||
NPLI | -0.13 | 0.71 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 12.5 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | 0.01 | 0.19 | ||
LNE | 0.01 | 0.50 | ||
PTB | -0.06 | 0.19 | ||
BKSV-DPLA | 0.01 | 0.29 | ||
GUM | 0.01 | 0.50 | ||
METAS | 0.01 | 0.39 | ||
NMISA | 0.01 | 0.39 | ||
INMETRO | 0.00 | 0.24 | ||
CENAM | -7.89 | 0.50 | ||
NMIA | 0.06 | 0.29 | ||
NMIJ | 0.00 | 0.07 | ||
KRISS | -0.70 | 0.12 | ||
A*STAR | -0.02 | 0.34 | ||
CMS | 0.04 | 0.33 | ||
KRISS | 0.03 | 0.22 | ||
NMISA | 0.35 | 0.22 | ||
NPLI | -0.01 | 0.71 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 25 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | 0.07 | 0.19 | ||
LNE | 0.03 | 0.50 | ||
PTB | -0.12 | 0.19 | ||
BKSV-DPLA | 0.02 | 0.29 | ||
GUM | 0.03 | 0.50 | ||
METAS | 0.04 | 0.40 | ||
NMISA | -0.01 | 0.50 | ||
INMETRO | -0.03 | 0.24 | ||
CENAM | -15.66 | 0.50 | ||
NMIA | 0.13 | 0.29 | ||
NMIJ | 0.01 | 0.07 | ||
KRISS | -1.41 | 0.12 | ||
A*STAR | -0.04 | 0.34 | ||
CMS | 0.19 | 0.33 | ||
KRISS | 0.10 | 0.22 | ||
NMISA | 0.79 | 0.22 | ||
NPLI | -0.16 | 0.81 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 40 Hz
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in °.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
(°) |
(°) |
(°) |
||
NIM | 0.03 | 0.39 | ||
LNE | 0.05 | 0.49 | ||
PTB | -0.19 | 0.19 | ||
BKSV-DPLA | 0.11 | 0.49 | ||
GUM | 0.06 | 0.49 | ||
METAS | 0.09 | 0.40 | ||
NMISA | -0.03 | 0.49 | ||
INMETRO | -0.04 | 0.24 | ||
CENAM | -24.68 | 0.50 | ||
NMIA | 0.19 | 0.29 | ||
NMIJ | 0.02 | 0.07 | ||
KRISS | -2.23 | 0.12 | ||
A*STAR | 0.01 | 0.34 | ||
CMS | 0.05 | 0.33 | ||
KRISS | 0.27 | 0.22 | ||
NMISA | 1.40 | 0.22 | ||
NPLI | 0.11 | 0.91 |
CCAUV.V-K3 and APMP.AUV.V-K3.1
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
Degrees of equivalence: Di and Ui
Lab i | 0.1 Hz | 0.125 Hz | 0.16 Hz | 0.2 Hz | 0.25 Hz | 0.315 Hz | 0.4 Hz | 0.5 Hz | |||||||||||||||
Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | ||||||||
in ° | in ° | in ° | in ° | in ° | in ° | in ° | in ° | ||||||||||||||||
NIM | -0.01 | 0.19 | -0.01 | 0.19 | -0.01 | 0.19 | -0.01 | 0.20 | 0.00 | 0.19 | 0.00 | 0.19 | 0.00 | 20.00 | 0.02 | 20.00 | |||||||
LNE | - | - | - | - | - | - | - | - | - | - | - | - | 1.34 | 0.50 | 0.01 | 0.50 | |||||||
PTB | -0.06 | 0.19 | -0.08 | 0.19 | -0.10 | 0.19 | -0.12 | 0.20 | -0.08 | 0.19 | -0.10 | 0.19 | -0.12 | 0.20 | -0.05 | 0.20 | |||||||
BKSV-DPLA | -0.04 | 0.30 | -0.04 | 0.30 | -0.01 | 0.30 | -0.01 | 0.30 | 0.00 | 0.30 | -0.02 | 0.30 | 0.00 | 0.30 | 0.01 | 0.30 | |||||||
GUM | - | - | - | - | - | - | -0.10 | 0.90 | -0.09 | 0.90 | -0.08 | 0.90 | -0.06 | 0.50 | -0.05 | 0.50 | |||||||
METAS | - | - | - | - | - | - | 0.09 | 0.78 | 0.09 | 0.78 | 0.07 | 0.78 | 0.06 | 0.78 | 0.05 | 0.78 | |||||||
NMISA | -0.04 | 0.40 | -0.02 | 0.40 | -0.01 | 0.40 | -0.01 | 0.40 | 0.00 | 0.40 | 0.00 | 0.40 | 0.02 | 0.40 | 0.01 | 0.40 | |||||||
INMETRO | - | - | - | - | - | - | -0.01 | 0.25 | 0.01 | 0.25 | 0.01 | 0.25 | 0.01 | 0.25 | 0.01 | 0.25 | |||||||
CENAM | 0.06 | 0.19 | 0.17 | 0.19 | 0.19 | 0.19 | 0.25 | 0.20 | 0.17 | 0.50 | 0.24 | 0.50 | 0.30 | 0.50 | -0.31 | 0.50 | |||||||
NMIA | -0.03 | 0.30 | -0.01 | 0.30 | 0.00 | 0.30 | -0.01 | 0.30 | 0.00 | 0.30 | 0.00 | 0.30 | 0.01 | 0.30 | 0.01 | 0.30 | |||||||
NMIJ | 0.00 | 0.02 | 0.00 | 0.02 | 0.00 | 0.02 | -0.01 | 0.02 | 0.00 | 0.02 | 0.00 | 0.02 | 0.01 | 0.02 | 0.01 | 0.02 | |||||||
KRISS | - | - | - | - | - | - | - | - | - | - | - | - | - | - | -0.02 | 0.09 | |||||||
A*STAR | 0.10 | 0.87 | 0.11 | 0.86 | -0.03 | 0.86 | -0.02 | 0.51 | 0.01 | 0.51 | -0.01 | 0.51 | 0.01 | 0.31 | 0.02 | 0.31 | |||||||
CMS | -0.02 | 0.31 | -0.01 | 0.31 | -0.02 | 0.31 | -0.01 | 0.31 | 0.00 | 0.31 | 0.00 | 0.31 | 0.01 | 0.31 | 0.02 | 0.31 | |||||||
KRISS | -0.01 | 0.21 | -0.01 | 0.21 | 0.00 | 0.21 | 0.00 | 0.21 | 0.02 | 0.21 | 0.03 | 0.21 | 0.03 | 0.21 | 0.03 | 0.21 | |||||||
NMISA | -0.03 | 0.21 | -0.02 | 0.21 | -0.01 | 0.21 | -0.01 | 0.21 | 0.00 | 0.21 | 0.00 | 0.21 | 0.01 | 0.21 | 0.03 | 0.21 | |||||||
NPLI | -0.04 | 1.01 | -0.03 | 1.01 | 0.00 | 1.01 | 0.03 | 1.01 | 0.02 | 1.01 | -0.03 | 1.01 | 0.03 | 1.01 | -0.28 | 0.71 |
Degrees of equivalence: Di and Ui (Continued)
Lab i | 0.63 Hz | 0.8 Hz | 1 Hz | 1.25 Hz | 1.6 Hz | 2 Hz | 2.50 Hz | 3.15 Hz | |||||||||||||||
Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | ||||||||
in ° | in ° | in ° | in ° | in ° | in ° | in ° | in ° | ||||||||||||||||
NIM | -0.01 | 0.20 | 0.05 | 0.20 | 0.02 | 0.20 | 0.02 | 0.20 | 0.02 | 0.20 | 0.01 | 0.20 | 0.02 | 0.20 | 0.03 | 0.20 | |||||||
LNE | 0.01 | 0.50 | 0.01 | 0.50 | - | - | - | - | - | - | 0.00 | 0.50 | 0.01 | 0.50 | 0.01 | 0.50 | |||||||
PTB | -0.05 | 0.20 | -0.05 | 0.20 | -0.03 | 0.20 | -0.03 | 0.20 | -0.04 | 0.20 | -0.06 | 0.20 | -0.02 | 0.20 | -0.02 | 0.20 | |||||||
BKSV-DPLA | 0.01 | 0.30 | 0.01 | 0.30 | 0.01 | 0.30 | 0.02 | 0.30 | 0.02 | 0.30 | 0.02 | 0.30 | 0.03 | 0.30 | 0.03 | 0.30 | |||||||
GUM | -0.04 | 0.50 | -0.03 | 0.50 | -0.02 | 0.50 | 0.00 | 0.50 | 0.00 | 0.50 | 0.00 | 0.50 | 0.02 | 0.50 | 0.03 | 0.50 | |||||||
METAS | 0.04 | 0.78 | 0.04 | 0.78 | 0.03 | 0.46 | 0.04 | 0.46 | 0.04 | 0.46 | 0.03 | 0.46 | 0.04 | 0.46 | 0.05 | 0.46 | |||||||
NMISA | 0.01 | 0.40 | 0.01 | 0.40 | 0.01 | 0.40 | 0.03 | 0.40 | 0.02 | 0.40 | 0.02 | 0.40 | 0.03 | 0.40 | 0.04 | 0.40 | |||||||
INMETRO | 0.01 | 0.25 | 0.01 | 0.25 | 0.01 | 0.25 | 0.03 | 0.25 | 0.02 | 0.25 | 0.02 | 0.25 | 0.03 | 0.25 | 0.04 | 0.25 | |||||||
CENAM | -0.39 | 0.50 | -0.52 | 0.50 | -0.63 | 0.50 | -0.86 | 0.50 | -1.09 | 0.50 | -1.26 | 0.50 | -1.47 | 0.50 | -1.98 | 0.50 | |||||||
NMIA | 0.01 | 0.30 | 0.01 | 0.30 | 0.02 | 0.30 | 0.03 | 0.30 | 0.03 | 0.30 | 0.03 | 0.30 | 0.04 | 0.30 | 0.05 | 0.30 | |||||||
NMIJ | 0.01 | 0.03 | 0.01 | 0.03 | 0.01 | 0.03 | 0.02 | 0.03 | 0.02 | 0.03 | 0.02 | 0.03 | 0.03 | 0.03 | 0.03 | 0.03 | |||||||
KRISS | -0.04 | 0.09 | -0.05 | 0.09 | -0.05 | 0.09 | -0.06 | 0.09 | -0.09 | 0.09 | -0.11 | 0.09 | -0.13 | 0.09 | -0.16 | 0.09 | |||||||
A*STAR | 0.02 | 0.31 | 0.01 | 0.31 | 0.02 | 0.35 | 0.03 | 0.35 | 0.03 | 0.35 | 0.02 | 0.35 | 0.02 | 0.35 | 0.01 | 0.35 | |||||||
CMS | -0.01 | 0.31 | 0.06 | 0.31 | 0.03 | 0.31 | 0.03 | 0.31 | 0.03 | 0.31 | 0.03 | 0.31 | 0.05 | 0.33 | 0.04 | 0.33 | |||||||
KRISS | 0.00 | 0.21 | 0.08 | 0.21 | 0.05 | 0.21 | 0.05 | 0.21 | 0.04 | 0.21 | 0.04 | 0.21 | 0.05 | 0.21 | 0.04 | 0.21 | |||||||
NMISA | 0.00 | 0.21 | 0.07 | 0.21 | 0.05 | 0.21 | 0.06 | 0.21 | 0.06 | 0.21 | 0.06 | 0.21 | 0.10 | 0.21 | 0.11 | 0.21 | |||||||
NPLI | -0.26 | 0.71 | -0.15 | 0.71 | -0.13 | 0.71 | -0.10 | 0.71 | -0.08 | 0.71 | -0.09 | 0.71 | -0.06 | 0.71 | -0.06 | 0.71 |
Degrees of equivalence: Di and Ui (Continued)
Lab i | 4 Hz | 5 Hz | 6.3 Hz | 8 Hz | 10 Hz | 12.5 Hz | 16 Hz | 20 Hz | |||||||||||||||
Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | Di | Ui | ||||||||
in ° | in ° | in ° | in ° | in ° | in ° | in ° | in ° | ||||||||||||||||
NIM | 0.03 | 0.20 | 0.03 | 0.20 | 0.02 | 0.19 | 0.00 | 0.19 | 0.01 | 0.19 | 0.01 | 0.19 | 0.01 | 0.19 | 0.04 | 0.19 | |||||||
LNE | 0.04 | 0.50 | 0.00 | 0.50 | 0.00 | 0.50 | 0.01 | 0.50 | 0.01 | 0.50 | 0.01 | 0.50 | 0.02 | 0.50 | 0.02 | 0.50 | |||||||
PTB | -0.03 | 0.20 | -0.03 | 0.20 | -0.03 | 0.19 | -0.06 | 0.19 | -0.05 | 0.19 | -0.06 | 0.19 | -0.08 | 0.19 | -0.09 | 0.19 | |||||||
BKSV-DPLA | 0.05 | 0.30 | 0.01 | 0.30 | 0.02 | 0.29 | 0.02 | 0.29 | 0.01 | 0.29 | 0.01 | 0.29 | 0.02 | 0.29 | 0.00 | 0.29 | |||||||
GUM | 0.04 | 0.50 | 0.01 | 0.50 | 0.02 | 0.50 | 0.02 | 0.50 | 0.01 | 0.50 | 0.01 | 0.50 | 0.03 | 0.50 | 0.03 | 0.50 | |||||||
METAS | 0.05 | 0.46 | 0.01 | 0.46 | 0.02 | 0.45 | 0.02 | 0.45 | 0.02 | 0.39 | 0.01 | 0.39 | 0.03 | 0.39 | 0.04 | 0.40 | |||||||
NMISA | 0.05 | 0.40 | 0.01 | 0.40 | 0.02 | 0.39 | 0.02 | 0.39 | 0.01 | 0.80 | 0.01 | 0.39 | 0.03 | 0.50 | 0.01 | 0.50 | |||||||
INMETRO | 0.05 | 0.25 | 0.01 | 0.25 | 0.02 | 0.24 | 0.01 | 0.24 | 0.00 | 0.24 | 0.00 | 0.24 | 0.01 | 0.24 | -0.04 | 0.24 | |||||||
CENAM | -2.51 | 0.50 | -3.18 | 0.50 | -4.11 | 0.50 | -5.06 | 0.50 | -6.33 | 0.50 | -7.89 | 0.50 | #### | 0.50 | #### | 0.50 | |||||||
NMIA | 0.06 | 0.30 | 0.03 | 0.30 | 0.05 | 0.29 | 0.03 | 0.29 | 0.04 | 0.29 | 0.06 | 0.29 | 0.08 | 0.29 | 0.11 | 0.29 | |||||||
NMIJ | 0.04 | 0.03 | 0.00 | 0.02 | 0.00 | 0.07 | 0.01 | 0.07 | 0.00 | 0.07 | 0.00 | 0.07 | 0.01 | 0.07 | 0.00 | 0.07 | |||||||
KRISS | -0.20 | 0.09 | -0.29 | 0.09 | -0.35 | 0.12 | -0.45 | 0.12 | -0.57 | 0.12 | -0.70 | 0.12 | -0.92 | 0.12 | -1.12 | 0.12 | |||||||
A*STAR | 0.02 | 0.35 | -0.04 | 0.35 | -0.03 | 0.34 | -0.03 | 0.34 | -0.03 | 0.34 | -0.02 | 0.34 | -0.12 | 0.34 | -0.04 | 0.34 | |||||||
CMS | 0.06 | 0.33 | 0.03 | 0.33 | 0.05 | 0.33 | 0.03 | 0.33 | 0.08 | 0.33 | 0.04 | 0.33 | 0.04 | 0.33 | 0.09 | 0.33 | |||||||
KRISS | 0.05 | 0.21 | 0.04 | 0.21 | 0.03 | 0.22 | 0.02 | 0.22 | 0.03 | 0.22 | 0.03 | 0.22 | 0.04 | 0.22 | 0.06 | 0.22 | |||||||
NMISA | 0.14 | 0.21 | 0.16 | 0.21 | 0.16 | 0.22 | 0.21 | 0.22 | 0.29 | 0.22 | 0.35 | 0.22 | 0.45 | 0.22 | 0.59 | 0.22 | |||||||
NPLI | -0.02 | 0.71 | -0.02 | 0.71 | -0.13 | 0.71 | -0.04 | 0.71 | -0.02 | 0.71 | -0.01 | 0.71 | -0.03 | 0.71 | -0.09 | 0.81 |
Degrees of equivalence: Di and Ui (Continued)
Lab i | 25 Hz | 31.5 Hz | 40 Hz | |||||
Di | Ui | Di | Ui | Di | Ui | |||
in ° | in ° | in ° | ||||||
NIM | 0.07 | 0.19 | 0.09 | 0.19 | 0.03 | 0.39 | ||
LNE | 0.03 | 0.50 | 0.03 | 0.50 | 0.05 | 0.49 | ||
PTB | -0.12 | 0.19 | -0.14 | 0.19 | -0.19 | 0.19 | ||
BKSV-DPLA | 0.02 | 0.29 | -0.01 | 0.50 | 0.11 | 0.49 | ||
GUM | 0.03 | 0.50 | 0.05 | 0.50 | 0.06 | 0.49 | ||
METAS | 0.04 | 0.40 | 0.07 | 0.40 | 0.09 | 0.40 | ||
NMISA | -0.01 | 0.50 | 0.00 | 0.50 | -0.03 | 0.49 | ||
INMETRO | -0.03 | 0.24 | -0.01 | 0.24 | -0.04 | 0.24 | ||
CENAM | -15.66 | 0.50 | -19.55 | 0.50 | -24.68 | 0.50 | ||
NMIA | 0.13 | 0.29 | 0.15 | 0.29 | 0.19 | 0.29 | ||
NMIJ | 0.01 | 0.07 | -0.01 | 0.07 | 0.02 | 0.07 | ||
KRISS | -1.41 | 0.12 | -1.73 | 0.12 | -2.23 | 0.12 | ||
A*STAR | -0.04 | 0.34 | 0.00 | 0.34 | 0.01 | 0.34 | ||
CMS | 0.19 | 0.33 | 0.21 | 0.21 | 0.05 | 0.33 | ||
KRISS | 0.10 | 0.22 | 0.14 | 0.14 | 0.27 | 0.22 | ||
NMISA | 0.79 | 0.22 | 1.05 | 1.05 | 1.40 | 0.22 | ||
NPLI | -0.16 | 0.81 | -0.28 | -0.28 | 0.11 | 0.91 |
Metrology area, Sub-field | Acoustics, Ultrasound, Vibration, Vibration |
Description | Acceleration complex sensitivity |
Time of measurements | 2025 - 2026 |
Status | Protocol complete |
References | |
Measurand | Acceleration complex sensitivity Nominal values for the 3 artifacts are: 1) Sensitivity magnitude: 80 mV/(m/s2); Phase: 0 degree 2) Sensitivity magnitude: 200 mV/(m/s2); Phase: 0 degree 3) Sensitivity magnitude: 1 V/(m/s2); Phase: 0 degree |
Parameters | Sensitivity magnitude range: 50 mV/(m/s2) to 1.5 V/(m/s2) Phase range: 0 degree to 360 degree |
Transfer device | Three accelerometers |
Comparison type | Key Comparison |
Consultative Committee | CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration) |
Conducted by | SIM (Inter-American Metrology System) |
Comments | This regional KC is intended to extend the results of CCAUV.V-K3 to SIM NMIs who were unable to participate in CCAUV.V-K3. This regional KC will compare measurements of the complex sensitivity of accelerometers exposed to sinusoidal accelerations in the frequency range from 0.1 Hz to 40 Hz; the mandatory range is 0.4 Hz to 40 Hz. In addition, there is an extended range of 0.05 Hz to 0.1 Hz and 40 Hz to 200 Hz. |
Pilot institute |
NIST
National Institute of Standards and Technology United States |
Contact person | Akobuije Chijioke |
Additional |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCAUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.1 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 136.75 mV / (m/s2) UR = 0.15 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.2 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 136.72 mV / (m/s2) UR = 0.12 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3..
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.4 Hz
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 136.64 mV / (m/s2) UR = 0.10 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3..
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.8 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 136.55 mV / (m/s2) UR = 0.08 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3 to CCAUV.V-K3
The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3..
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 1.6 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 136.54 mV / (m/s2) UR = 0.08 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3 to CCAUV.V-K3
The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 3.15 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 136.55 mV / (m/s2) UR = 0.08 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3 to CCAUV.V-K3
The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 6.3 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 136.62 mV / (m/s2) UR = 0.08 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3 to CCAUV.V-K3
The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 12.5 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 136.71 mV / (m/s2) UR = 0.07 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3 to CCAUV.V-K3
The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 25 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 137.36 mV / (m/s2) UR = 0.08 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 40 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = 138.47 mV / (m/s2) UR = 0.07 mV / (m/s2)
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the Final Report and is based on a weighted mean.
0.1 Hz | 0.125 Hz | 0.16 Hz | 0.2 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | |||||||
136.75 | 0.15 | 136.71 | 0.15 | 136.69 | 0.15 | 136.72 | 0.12 | |||
0.25 Hz | 0.315 Hz | 0.4 Hz | 0.5 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | |||||||
136.68 | 0.12 | 136.66 | 0.12 | 136.64 | 0.10 | 136.58 | 0.08 | |||
0.63 Hz | 0.8 Hz | 1 Hz | 1.25 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | |||||||
136.56 | 0.08 | 136.55 | 0.08 | 136.54 | 0.08 | 136.54 | 0.08 | |||
1.6 Hz | 2 Hz | 2.5 Hz | 3.15 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | |||||||
136.54 | 0.08 | 136.54 | 0.08 | 136.55 | 0.08 | 136.55 | 0.08 |
4 Hz | 5 Hz | 6.3 Hz | 8 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | |||||||
136.57 | 0.08 | 136.59 | 0.08 | 136.62 | 0.08 | 136.61 | 0.08 | |||
10 Hz | 12.5 Hz | 16 Hz | 20 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | |||||||
136.63 | 0.07 | 136.71 | 0.07 | 136.82 | 0.07 | 137.06 | 0.08 | |||
25 Hz | 31.5 Hz | 40 Hz | ||||||||
xR | UR | xR | UR | xR | UR | |||||
mV / (m/s2) | mV / (m/s2) | mV / (m/s2) | ||||||||
137.36 | 0.08 | 137.77 | 0.07 | 138.47 | 0.07 |
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR,M and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3 to CCAUV.V-K3
The linking was made via NIM, NMIA, NMISA, KRISS and NMIJ/AIST, having participated in both comparisons.
Linking of AFRIMETS.AUV.V-K3 to CCAUV.V-K3
The linking was made via NMISA and INMETRO, having participated in both comparisons. The linking process is described in the Appendix B of the AFRIMETS.AUV.V-K3 Final Report.
Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3
The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3..
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.1 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -0.34 ° UR = 0.04 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.2 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -0.67 ° uR = 0.04 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.4 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -1.36 ° UR = 0.04 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.8 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -0.04 ° UR = 0.04 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 1.6 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -0.08 ° UR = 0.04 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 3.15 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -0.15 ° UR = 0.04 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 6.3 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -0.25 ° UR = 0.07 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 12.5 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -0.48 ° UR = 0.07 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 25 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -1.03 ° UR = 0.07 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 40 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.
xR = -1.87 ° UR = 0.07 °
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
CCAUV.V-K3
The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the Final Report and is based on a weighted mean.
0.1 Hz | 0.125 Hz | 0.16 Hz | 0.2 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
in ° | in ° | in ° | in ° | |||||||
-0.34 | 0.04 | -0.42 | 0.04 | -0.54 | 0.04 | -0.67 | 0.04 | |||
0.25 Hz | 0.315 Hz | 0.4 Hz | 0.5 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
in ° | in ° | in ° | in ° | |||||||
-0.85 | 0.04 | -1.07 | 0.04 | -1.36 | 0.04 | -0.03 | 0.04 | |||
0.63 Hz | 0.8 Hz | 1 Hz | 1.25 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
in ° | in ° | in ° | in ° | |||||||
-0.03 | 0.04 | -0.04 | 0.04 | -0.05 | 0.04 | -0.07 | 0.04 | |||
1.6 Hz | 2 Hz | 2.5 Hz | 3.15 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
in ° | in ° | in ° | in ° | |||||||
-0.08 | 0.04 | -0.09 | 0.04 | -0.12 | 0.04 | -0.15 | 0.04 |
4 Hz | 5 Hz | 6.3 Hz | 8 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
in ° | in ° | in ° | in ° | |||||||
-0.19 | 0.04 | -0.19 | 0.04 | -0.25 | 0.07 | -0.31 | 0.07 | |||
10 Hz | 12.5 Hz | 16 Hz | 20 Hz | |||||||
xR | UR | xR | UR | xR | UR | xR | UR | |||
in ° | in ° | in ° | in ° | |||||||
-0.38 | 0.07 | -0.48 | 0.07 | -0.63 | 0.07 | -0.80 | 0.07 | |||
25 Hz | 31.5 Hz | 40 Hz | ||||||||
xR | UR | xR | UR | xR | UR | |||||
in ° | in ° | in ° | ||||||||
-1.03 | 0.07 | -1.37 | 0.07 | -1.87 | 0.07 |
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.
Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3
The linking was made via NIM, having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.1 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.2 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.4 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 0.8 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
APMP.AUV.V-K3
The laboratories' individual measurement may be found in Section 7.2 of the APMP.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 1.6 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Sections Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
APMP.AUV.V-K3
The laboratories' individual measurement may be found in Section 7.2 of the APMP.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 3.15 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Sections Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
APMP.AUV.V-K3
The laboratories' individual measurement may be found in Section 7.2 of the APMP.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 6.3 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
APMP.AUV.V-K3
The laboratories' individual measurement may be found in Section 7.2 of the APMP.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 12.5 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
APMP.AUV.V-K3
The laboratories' individual measurement may be found in Section 7.2 of the APMP.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 25 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Magnitude
FREQUENCY : 40 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3, AFRIMETS.AUV.V-K3, APMP.AUV.V-K3, EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Magnitude
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
AFRIMETS.AUV.V-K3
The laboratories' individual measurement may be found in Section 9.2 of the AFRIMETS.AUV.V-K3 Final Report.
APMP.AUV.V-K3
The laboratories' individual measurement may be found in Section 7.2 of the APMP.AUV.V-K3 Final Report.
EURAMET.AUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.1 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.2 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.4 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 0.8 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 1.6 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report;
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 3.15 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 6.3 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report;
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 12.5 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report;
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 25 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report
CCAUV.V-K3 and APMP.AUV.V-K3.1
MEASURAND: Phase
FREQUENCY : 40 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report
CCAUV.V-K3 and APMP.AUV.V-K3.1
DEVICE : Low-frequency vibration transfer standard
MEASURAND : Phase
FREQUENCIES : 0.1 Hz, 0.125 Hz, 0.16 Hz, 0.2 Hz, 0.25 Hz, 0.315 Hz, 0.4 Hz, 0.5 Hz, 0.63 Hz, 0.8 Hz, 1 Hz, 1.25 Hz, 1.60 Hz, 2 Hz, 2.5 Hz, 3.15 Hz, 4 Hz, 5 Hz, 6.3 Hz, 8 Hz, 10 Hz, 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz
CCAUV.V-K3
The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.
APMP.AUV.V-K3.1
The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report