CCM.F-K3.a, CCM.F-K3.b, APMP.M.F-K3.a and APMP.M.F-K3.b
MEASURAND Force
NOMINAL VALUE 500 kN
Degrees of equivalence relative to the CCM.F-K3 reference value and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-6.
Enlarged graph

CCM.F-K3.a and APMP.M.F-K3.a
MEASURAND Force
NOMINAL VALUE 1 MN
Degrees of equivalence relative to the CCM.F-K3 reference value and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-6.
Enlarged graph

CCM.F-K3.a, CCM.F-K3.b, APMP.M.F-K3.a and APMP.M.F-K3.b
MEASURAND Force
NOMINAL VALUE 500 kN
Degrees of equivalence relative to the CCM.F-K3 reference value and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-6.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
10-6 |
10-6 |
10-6 |
||
| NPL | -12.0 | 11.3 | ||
| NIST | -1.9 | 18.9 | ||
| INRIM | 10.4 | 20.4 | ||
| VNIIM | 8.8 | 20.7 | ||
| NIM | 7.8 | 14.0 | ||
| PTB | 1.5 | 20.2 | ||
| LNE | -5.8 | 20.7 | ||
| CEM | 5.1 | 20.6 | ||
| GUM | 36.6 | 120.1 | ||
| NMIA | -15.6 | 20.7 | ||
| NMIJ | -4.4 | 12.5 | ||
| KRISS | 61.3 | 21.0 | ||
| NMIA | 190 | 217 | ||
| NMC, A*STAR | -63.6 | 209 | ||
| NPLI | 24.5 | 110 | ||
| SCL | -124 | 270 | ||
| CMS | 445 | 504 | ||
| SNSU-BSN | 10.9 | 203 | ||
| VMI-STAMEQ | 125 | 113 |
CCM.F-K3.a and APMP.M.F-K3.a
MEASURAND Force
NOMINAL VALUE 1 MN
Degrees of equivalence relative to the CCM.F-K3 reference value and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-6.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
10-6 |
10-6 |
10-6 |
||
| NPL | -9.5 | 12.7 | ||
| NIST | -9.2 | 19.4 | ||
| INRiM | 8.9 | 22.5 | ||
| VNIIM | 3.9 | 22.4 | ||
| NIM | 10.3 | 16.1 | ||
| PTB | 4.2 | 22.2 | ||
| NMIA | 249 | 216 | ||
| NMC, A*STAR | -80.6 | 207 | ||
| NPLI | 38.1 | 109 | ||
| SCL | -1129 | 503 | ||
| CMS | 343 | 506 | ||
| SNSU-BSN | -17.8 | 203 | ||
| VMI-STAMEQ | 50.9 | 118 |
| Metrology area, Sub-field | Mass and related quantities, Force |
| Description | High force measurements |
| Time of measurements | 2023 - 2024 |
| Status | Measurements in progress |
| References | |
| Measurand | Force 0 MN, 0.5 MN and 1 MN |
| Transfer device | Force measurement device |
| Comparison type | Key Comparison |
| Consultative Committee | CCM (Consultative Committee for Mass and Related Quantities) |
| Conducted by | SIM (Inter-American Metrology System) |
| Comments | The technical comparison protocol has been updated to version 4. |
| Pilot institute |
INM (CO)
Instituto Nacional de Metrología de Colombia Colombia |
| Contact person | Juan Alberto Arias Prieto +571-2 54 22 22 |
| Additional |
| Pilot laboratory | |
|---|---|
| INM (CO) |
Instituto Nacional de Metrología de Colombia, Colombia, SIM |
| CENAM |
Centro Nacional de Metrologia, Mexico, SIM |
| IDIC |
Instituto de Investigaciones y Control, Chile, SIM |
| INEN |
Servicio Ecuatoriano de Normalización, Ecuador, SIM |
| INMETRO |
Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM |
| INTI |
Instituto Nacional de Tecnologia Industrial, Argentina, SIM |
| LACOMET |
Laboratorio Costarricense de Metrologia, Costa Rica, SIM |
| NIST |
National Institute of Standards and Technology, United States, SIM |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCM.F-K3.a, CCM.F-K3.b, APMP.M.F-K3.a and APMP.M.F-K3.b
MEASURAND Force
NOMINAL VALUE 500 kN
The key comparison reference value, xR, is calculated as the weighted mean of the corrected measurement results from CCM.F-K3.a and CCM.F-K3.b. Correlations were considered when evaluating the associated standard uncertainty, uR.
xR = 0.5 × 10-6 uR = 1.8 × 10-6
The degree of equivalence of each laboratory i with respect to the key comparison reference reference value is given by a pair of terms, both expressed in 10-6:
Di , the relative deviation of the final result from xR, and Ui, its expanded uncertainty (k = 2).
Linking APMP.M.F-K3.a and APMP.M.F-K3.b to CCM.F-K3.a and CCM.F-K3.b
The institutesNIM, PTB and NMIJ provide the link between the key comparison CCM.F-K3.a and CCM.F-K3.b and APMP.M.F-K3.a and APMP.M.F-K3.b, having participated in both comparisons. The linking process is described in Section 8 of the APMP.M.F-K3.a and b Final report.
CCM.F-K3.a and APMP.M.F-K3.a
MEASURAND Force
NOMINAL VALUE 1 MN
The key comparison reference value, xR, is calculated as the weighted mean of the corrected measurement results from CCM.F-K3.a. Correlations were considered when evaluating the associated standard uncertainty, uR.
xR = 1.0 × 10-6 uR = 4.4 × 10-6
The degree of equivalence of each laboratory i with respect to the key comparison reference reference value is given by a pair of terms, both expressed in 10-6:
Di , the relative deviation of the final result from xR, and Ui, its expanded uncertainty (k = 2).
Linking APMP.M.F-K3.a and APMP.M.F-K3.b to CCM.F-K3.a and CCM.F-K3.b
The institutesNIM, PTB and NMIJ provide the link between the key comparison CCM.F-K3.a and CCM.F-K3.b and APMP.M.F-K3.a and APMP.M.F-K3.b, having participated in both comparisons. The linking process is described in Section 8 of the APMP.M.F-K3.a and b Final report.
CCM.F-K3.a, CCM.F-K3.b, APMP.M.F-K3.a and APMP.M.F-K3.b
MEASURAND Force
NOMINAL VALUE 500 kN
CCM.F-K3.a and CCM.F-K3.b
CCM.F-K3.a: Two force steps of 500 kN were employed.
CCM.F-K3.b: Two compression force transducers of 500 kN and one force step of 500 kN were employed.
Measurements were carried out from 2005 to 2011.
The measurement results for CCM.F-K3.a and CCM.F-K3.b are presented in Section 7 (pp. 6 to 17) of the CCM.F-K3 Final report.
APMP.M.F-K3.a and APMP.M.F-K3.b
The measurement results for APMP.M.F-K3.a and APMP.M.F-K3.b are presented in Section 6 of the APMP.M.F-K3 a and b Final report.
CCM.F-K3.a and APMP.M.F-K3.a
MEASURAND Force
NOMINAL VALUE 1 MN
CCM.F-K3.a
CCM.F-K3.a: Two compression force transducers of 1 MN were employed.
Measurements were carried out from 2005 to 2011.
The measurement results for CCM.F-K3.a are presented in Section 7 (pp. 6 to 17) of the CCM.F-K3 Final report.
APMP.M.F-K3.a
The measurement results for APMP.M.F-K3.a are presented in Section 6 of the APMP.M.F-K3 a and b Final report.