CIPM MRA COMPARISON REPORTS

CCQM-K182

Lithium isotope delta measurements
L. Flierl , O. Rienitz , A. Pramann , A. Roethke , J. Eberhardt , D. Arnold , J. Vogl , M. Oelze , P. Seena Prem , T. Ren ... , P. Song , L. Yang , K.S. Lee , S.W. Heo , Y.G. Kim
Published on 30/09/2026 - © 2026 L. Flierl et al
Citation L. Flierl et al 2026 CIPM MRA Comparison reports 08020

Abstract

Five national metrology institutes and designated institutes participated in the key comparison CCQM-K182 "Lithium isotope delta measurements". Participants were requested to evaluate the δLSVEC 7Li, expressed in ‰, of two different Li2CO3 samples dissolved in dilute nitric acid. Due to the overall good agreement between the reported results for both measurands the arithmetic mean was selected as the key comparison reference value. Participants were encouraged to use MCICP- MS or the most accurate method available. Moreover, participants were required to apply appropriate measures to avoid the severe Li memory effect in their mass spectrometers. Successful participation in CCQM-K182 demonstrates measurement capabilities in determining δLSVEC7Li,ranging from -17 ‰ to 17 ‰ in dilute nitric acid.

Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.

The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).