CIPM MRA COMPARISON REPORTS
CCL-K3.n01
Abstract
At its meeting in 2014, the Consultative Committee for Length (CCL)
WG-MRA decided upon a key comparison on the measurement of angle
standards. Initially the comparison was to include both optical
polygons and angle encoders, however, the comparison was eventually
scaled back to only include optical polygons. The comparison was also
delayed due to a change in pilot laboratory, and artifact circulation
was started in April 2023 and completed in September 2025. Two optical
polygons by different manufacturers were circulated: a 12-sided
polygon similar to one used in the EURAMET.L-K3.n01 comparison, and a
10-sided polygon similar to the polygon used in the SIM.L-K3-2019
comparison. Unfortunately, the 12-sided polygon was not stable for the
duration of the comparison and an additional artifact stability term
was added in quadrature to the participants 12-sided polygon reported
uncertainties. The vast majority of participants successfully
supported their uncertainty claims.
Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.
The final report has been peer-reviewed and approved for publication by the CCL, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).