CIPM MRA COMPARISON REPORTS
APMP.M.FF-S2.2016
Abstract
The supplementary comparison, APMP.M.FF-S2.2016 in the field of water
flow was organized and conducted between two National Metrology
Institutes (NMIs), National Metrology Institute of Japan (NMIJ/AIST)
and Research Center for Metrology (RCM) in Indonesian Institute of
Sciences (Lembaga Ilmu Pengetahuan Indonesia or LIPI) in Indonesia.
This bilateral comparison was undertaken by the Technical Committee
for Fluid Flow (TCFF) of Asia Pacific Metrology Program (APMP), and
was piloted by RCM-LIPI. The travelling standards are two turbine
meters provided by RCM-LIPI covering the range of 0.5 m3/h
to 30 m3/h (about 8 L/min to 500 L/min). Each of turbine
meters was tested by each participating NMIs in the measurement ranges
of 0.5 m3/h to 4.5 m3/h and 3.5 m3/h
to 30 m3/h. The degree of equivalence between RCM-LIPI and
NMIJ/AIST was obtained at every measurement point.
The travelling
standards showed good stability for check of the capability in
RCM-LIPI since the uncertainty due to the transfer standard was less
than quoted uncertainty in RCM-LIPI. The En values were less than 1 in
the flow rate range between 1 m3/h and 30 m3/h,
resulting the RCM-LIPI results match those of NMIJ within the claimed
uncertainties. However, En value at the 0.5 m3/h was from 1
to 1.2, indicating “warningâ€.