CIPM MRA COMPARISON REPORTS

APMP.L-K3.n01

Angle gauge blocks
K. Vacharanukul , P. Cox , O. Novyanto , N. Alfiyati , R.M. Halim , R. Rosli , F. Alqahtani , Y.B. Talha , M.S.A.S. Al Senaidi , T.C. Dung ... , V.K. Phan , A. Fursa , Y. Glushko , G. Tang , H. Chiu , Y. Tilek , S. Gulaikhan
Published on 09/06/2026 - © 2026 K. Vacharanukul et al
Citation K. Vacharanukul et al 2026 CIPM MRA Comparison reports 04005

Abstract

The APMP.L-K3.n01 key comparison, led by NIMT, involved ten laboratories calibrating angle gauge blocks of 1′, 25′, 3°, and 30°. The study, focused on pitch angle deviation, utilized weighted means for KCRV determination and Birge ratios for consistency checks. Despite artifact stability throughout the 2021–2024 period, the agreement was not entirely satisfactory, as 20% of the results (8 out of 40) reported |En| values exceeding 1.