CCT-K5, CCT-K5.1 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 961 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1000 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circles: participants in APMP.T-K5
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1064 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5, COOMET.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1084 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circle: participant in APMP.T-K5
Purple circle: participant in COOMET.T-K5.1
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1100 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circle: participant in APMP.T-K5
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1200 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circles: participants in APMP.T-K5
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1300 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circles: participants in APMP.T-K5
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1400 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circles: participants in APMP.T-K5
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1500 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circles: participants in APMP.T-K5
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1600 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circles: participants in APMP.T-K5
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1700 °C

Degrees of equivalence:
Di and expanded uncertainty Ui (k = 2), both expressed in K

Red diamonds: participants in CCT-K5
Green triangle: participant in CCT-K5.1
Blue circle: participant in APMP.T-K5
Orange squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey squares: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

CCT-K5, CCT-K5.1 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 961 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.0100.227
NPL-0.1240.310
NMIA-0.0950.123
KRISS -0.0500.255
NIM 0.0550.259
A*STAR-0.4800.319
NMIJ0.0100.263
VNIIM 0.5550.333
NIST --
CENAM 0.0910.433
LNE-INM-0.2940.350
INRIM0.1110.171
PTB 0.0510.169
NRC0.1310.511

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1000 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Blue: participants in APMP.T-K5
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.0020.221
NPL-0.1140.294
NMIA-0.0570.116
KRISS -0.0220.235
NIM 0.0880.279
A*STAR-0.5020.314
NMIJ0.0530.258
VNIIM 0.3330.345
NIST0.1510.487
CENAM 0.0510.449
LNE-INM-0.3040.366
INRIM0.1110.179
PTB -0.0540.183
NRC0.0810.559

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1064 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.0330.257
NPL-0.1810.329
NMIA-0.0380.155
KRISS 0.0070.279
NIM --
A*STAR-0.5380.350
NMIJ0.0620.291
VNIIM -0.0930.365
NIST0.1340.536
CENAM 0.1040.515
LNE-INM-0.3160.401
INRIM0.1290.206
PTB -0.0460.205
NRC-0.0610.597

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5, COOMET.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1084 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1

Blue: participant in APMP.T-K5
Purple: participant in COOMET.T-K5.1

Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1100 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Blue: participant in APMP.T-K5
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.0170.250
NPL-0.2950.342
NMIA-0.0120.145
KRISS 0.0080.255
NIM 0.1580.339
A*STAR-0.5770.345
NMIJ0.1030.287
VNIIM -0.1970.437
NIST0.2350.556
CENAM 0.0700.531
LNE-INM-0.3200.424
INRIM0.1000.216
PTB -0.0350.211
NRC0.0000.637

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1200 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Blue: participants in APMP.T-K5
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.0550.262
NPL-0.3770.411
NMIA-0.0250.122
KRISS 0.0000.278
NIM 0.0850.379
A*STAR-0.7050.384
NMIJ0.1450.320
VNIIM -0.4450.686
NIST0.1830.632
CENAM 0.0030.628
LNE-INM-0.4270.445
INRIM0.0480.222
PTB 0.0330.220
NRC0.0880.696

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1300 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Blue: participants in APMP.T-K5
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.0450.322
NPL-0.4970.486
NMIA0.0000.165
KRISS 0.0200.343
NIM 0.1400.437
A*STAR-0.7950.459
NMIJ0.2100.393
VNIIM -0.6600.956
NIST 0.1640.719
CENAM -0.0470.755
LNE-INM-0.4920.580
INRIM0.0640.284
PTB 0.0240.283
NRC0.1290.835

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1400 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Blue: participants in APMP.T-K5
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.0450.385
NPL-0.5100.600
NMIA0.0450.221
KRISS 0.0550.410
NIM 0.1950.522
A*STAR-0.8850.519
NMIJ0.3100.473
VNIIM -0.8801.228
NIST 0.2600.815
CENAM -0.0500.906
LNE-INM-0.5100.629
INRIM0.1150.352
PTB 0.0000.352
NRC0.0700.968

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1500 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Blue: participants in APMP.T-K5
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.0820.414
NPL-0.5500.683
NMIA0.0580.259
KRISS 0.0280.467
NIM 0.2130.593
A*STAR-1.0520.588
NMIJ0.3530.538
VNIIM -0.9121.520
NIST 0.3360.918
CENAM 0.0001.026
LNE-INM-0.5800.730
INRIM0.1460.396
PTB -0.0250.394
NRC0.2111.055

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1600 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Blue: participants in APMP.T-K5
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.1130.452
NPL-0.5400.781
NMIA0.0530.262
KRISS 0.0280.502
NIM 0.2380.660
A*STAR-1.1330.652
NMIJ0.4430.599
VNIIM -0.5781.797
NIST 0.5911.014
CENAM -0.0051.224
LNE-INM-0.6150.820
INRIM0.1260.436
PTB 0.0010.433
NRC 0.1211.146

Results are presented under A4 printable format in Summary Results (PDF file).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1700 °C

Degrees of equivalence relative to the CCT-K5 key comparison reference value:

Dark blue: participants in CCT-K5
Green: participant in CCT-K5.1
Blue: participant in APMP.T-K5
Orange: participants in EUROMET.T-K5 (measurements with Lamp S/N C564)
Grey: participants in EUROMET.T-K5 (measurements with Lamp S/N C681)

Pair-wise degrees of equivalence are computed for CCT-K5 and CCT-K5.1 only.

   
Lab iDiUi
 / K/ K
VSL-0.2350.480
NPL-0.5650.854
NMIA0.0300.259
KRISS 0.0000.571
NIM 0.2500.740
A*STAR-1.3900.734
NMIJ0.4550.667
VNIIM 0.0352.022
NIST 0.5001.129
CENAM -0.1701.321
LNE-INM-0.8050.892
INRIM0.0050.482
PTB -0.1000.480
NRC0.0151.313

Results are presented under A4 printable format in Summary Results (PDF file).

Comparison
Comparison close
CC comparison
Linked comparison
APMP.T-K5
Metrology area, Sub-field Thermometry, Pyrometry
Description Comparison of realization of the ITS-90 using radiation thermometry over the range 962 °C to 2800 °C
Time of measurements 1997 - 2000
Status Approved for equivalence
Final Reports of the comparisons
Measurand Temperature: 962 °C to 2800 °C
Transfer device One 650 nm radiation thermometer
Comparison type Key Comparison
Consultative Committee CCT (Consultative Committee for Thermometry)
Conducted by APMP (Asia Pacific Metrology Programme)
Comments

Results published on 13 May 2010

CCT-K5.1, APMP.T-K5, COOMET.T-K5 and EUROMET.T-K5 results are linked to those of CCT-K5

Realizations of the ITS-90 between 961 °C and 1700 °C

Pilot institute NMIJ AIST
National Metrology Institute of Japan
Japan
Contact person F. Sakuma

+81 29 861 4031
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

NIM

National Institute of Metrology, China, APMP

NMC, A*STAR

National Metrology Centre, Agency for Science, Technology and Research, Singapore, APMP

NMIA

National Measurement Institute, Australia, Australia, APMP

CCT-K5, CCT-K5.1 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 961 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

There are no APMP.T-K5 results to be linked to those of CCT-K5 for this nominal temperature.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1000 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1064 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5, COOMET.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1084 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report. N.B. Pair-wise degrees of equivalence are no longer issued.

  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking COOMET.T-K5 results to those of CCT-K5

The linkage is made through the common participation of VIIM in both key comparisons, and is detailed in the COOMET.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1100 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1200 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1300 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1400 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1500 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1600 °C

  • Key comparison CCT-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1700 °C

  • Key comparison CCT.T-K5

The key comparison reference value TR for each nominal temperature Tnom and each lamp k is calculated on the basis of the median of measured radiance temperatures Ti(k, Tnom). Its standard uncertainty, u(TR), is obtained as the standard uncertainty of the median.

For each temperature Tnom the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in K. The computation of Di and Ui is explained in the Addendum to the CCT-K5 Final Report.

For each temperature Tnom the pair-wise degree of equivalence between laboratories i and j is given by a pair of terms: Dij and its expanded uncertainty Uij (k = 2) both expressed in K. The computation of Dij and Uij is also explained in the Addendum to the CCT-K5 Final Report.
  • Linking CCT-K5.1 results to those of CCT-K5

The linkage is made through the common participation of PTB in both key comparisons, and is detailed in the CCT-K5 and CCT-K5.1 Linkage Report.

  • Linking APMP.T-K5 results to those of CCT-K5

The linkage is made through the common participation of NMIJ, NIM, KRISS and NMIA in both key comparisons, and is detailed in the Addendum to the APMP.T-K5 Final Report.

  • Linking key comparison EUROMET.T-K5 to CCT-K5

The measurements of the EUROMET.T-K5 participants are directly linked to the key comparison reference value obtained in CCT-K5 as the protocols of the two key comparisons are identical and the transfer standards are the same (see in Chapter VII of the EUROMET.T-K5 Final Report).

Degrees of equivalence relative to the CCT-K5 key comparison reference values are computed for each of the transfer standards. Pair-wise degrees of equivalence inside EUROMET.T-K5 are available in the EUROMET.T-K5 Final Report (Tables 15 to 36).

CCT-K5, CCT-K5.1 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 961 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL 964.050 0.10  NMIA 963.950 0.02  KRISS 963.910 0.09  NIM 964.070 0.11  A*STAR 963.690 0.15  NMIJ 964.070 0.12  VNIIM

964.660

0.16

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 963.550 0.10  NMIA 963.480 0.02  KRISS 963.610 0.09  NIM 963.660 0.11  A*STAR 962.970 0.15  NMIJ 963.570 0.12  VNIIM

964.070

0.16

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 962.000 0.14  NIST - -  CENAM 962.260 0.21  LNE-INM 961.710 0.13  INRIM 962.360 0.07  PTB 962.340 0.15 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 962.130 0.14  NIST - -  CENAM 962.300 0.21  LNE-INM 962.080 0.13  INRIM 962.240 0.07  PTB 962.140 0.15

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 962.22 °C, uNRC = 0.19 °C

Lamp S/N 644C, TNRC = 962.22 °C, uNRC = 0.19 °C

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1000 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL 1002.078 0.10  NMIA 1001.998 0.02  KRISS 1001.978 0.09  NIM 1002.098 0.12  A*STAR 1001.678 0.15  NMIJ 1002.128 0.12  VNIIM

1002.568

0.16

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1001.808 0.10  NMIA 1001.778 0.02  KRISS 1001.868 0.09  NIM 1001.968 0.12  A*STAR 1001.208 0.15  NMIJ 1001.868 0.12  VNIIM

1001.988

0.16

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1000.124 0.14  NIST 1000.464 0.24  CENAM 1000.294 0.22  LNE-INM 999.754 0.14  INRIM 1000.414 0.08  PTB 1000.284 0.16 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1000.118 0.14  NIST 1000.308 0.24  CENAM 1000.278 0.22  LNE-INM 1000.108 0.14  INRIM 1000.278 0.08  PTB 1000.078 0.16

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 999.95 °C, uNRC = 0.20 °C

Lamp S/N 644C, TNRC = 1000.14 °C, uNRC = 0.20 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom
= 1064 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL 1066.205 0.11  NMIA 1066.165 0.02  KRISS 1066.145 0.10  NIM - -  A*STAR 1065.825 0.16  NMIJ 1066.305 0.13  VNIIM

1066.325

0.16

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1066.129 0.11  NMIA 1066.159 0.02  KRISS 1066.269 0.10  NIM - -  A*STAR 1065.499 0.16  NMIJ 1066.219 0.13  VNIIM

1065.889

0.16

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1064.114 0.15  NIST 1064.434 0.26  CENAM 1064.444 0.25  LNE-INM 1063.824 0.16  INRIM 1064.524 0.08  PTB 1064.354 0.17 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1064.114 0.15  NIST 1064.424 0.26  CENAM 1064.354 0.25  LNE-INM 1064.134 0.16  INRIM 1064.324 0.08  PTB 1064.144 0.17

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1064.45 °C, uNRC = 0.22 °C

Lamp S/N 644C, TNRC = 1064.46 °C, uNRC = 0.22 °C

  • Key comparison APMP.T-K5

No results were obtained for this nominal temperature by the three laboratories (A*STAR, KIM-LIPI and CMS/ITRI) which participated in APMP.T-K5 only: see Appendix B of the APMP.T-K5 Final Report.

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5, COOMET.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1084 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti


Lamp S/N C564

Lab i Ti / °C ui / °C
VSL 1086.602 0.11
NMIA 1086.572 0.02
KRISS 1086.562 0.08
NIM - -
A*STAR 1086.182 0.16
NMIJ 1086.722 0.13
VNIIM

1086.662

0.19

Lamp S/N C681

Lab i Ti / °C ui / °C
VSL 1086.522 0.11
NMIA 1086.552 0.02
KRISS 1086.642 0.08
NIM - -
A*STAR 1085.842 0.16
NMIJ 1086.622 0.13
VNIIM

1086.282

0.19

Lamp S/N C860

Lab i Ti / °C ui / °C
NPL 1085.207 0.15
NIST - -
CENAM 1085.517 0.26
LNE-INM 1084.917 0.16
INRIM 1085.607 0.09
PTB 1085.497 0.17

Lamp S/N C864

Lab i Ti / °C ui / °C
NPL 1085.111 0.15
NIST - -
CENAM 1085.401 0.26
LNE-INM 1085.191 0.16
INRIM 1085.361 0.09
PTB 1085.221 0.17

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1084.85 °C, uNRC = 0.23 °C

Lamp S/N 644C, TNRC = 1084.98 °C, uNRC = 0.23 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

  • Key comparison COOMET.T-K5

This bilateral comparison was carried out between VNIIM and NSC IM. The measurements were carried out from August to October 2014 using a vacuum tungsten ribbon lamb (S/N SI 10-300 No. 5). The individual results may be consulted in the COOMET.T-K5 Final Report .

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1100 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL 1102.041 0.11  NMIA 1102.011 0.02  KRISS 1102.001 0.10  NIM 1102.111 0.14  A*STAR 1101.601 0.16  NMIJ 1102.161 0.13  VNIIM

1101.961

0.21

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1101.925 0.11  NMIA 1101.965 0.02  KRISS 1102.015 0.10  NIM 1102.205 0.14  A*STAR 1101.245 0.16  NMIJ 1102.045 0.13  VNIIM

1101.645

0.21

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1100.102 0.16  NIST 1100.762 0.27  CENAM 1100.502 0.26  LNE-INM 1099.902 0.17  INRIM 1100.602 0.09  PTB 1100.462 0.18 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1100.108 0.16  NIST 1100.508 0.27  CENAM 1100.438 0.26  LNE-INM 1100.258 0.17  INRIM 1100.398 0.09  PTB 1100.268 0.18

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1100.34°C, uNRC = 0.24 °C

Lamp S/N 644C, TNRC = 1100.40 °C, uNRC = 0.24 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1200 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL 1201.915 0.12  NMIA 1201.925 0.02  KRISS 1201.915 0.12  NIM 1201.925 0.16  A*STAR 1201.415 0.18  NMIJ 1202.105 0.15  VNIIM

1201.595

0.34

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1202.045 0.12  NMIA 1202.095 0.02  KRISS 1202.155 0.12  NIM 1202.315 0.16  A*STAR 1201.245 0.18  NMIJ 1202.255

0.15

 VNIIM

1201.585

0.34

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1200.191 0.20  NIST 1200.851 0.31  CENAM 1200.561 0.31  LNE-INM 1199.951 0.19  INRIM 1200.671 0.10  PTB 1200.621 0.19 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1200.105 0.20  NIST 1200.565 0.31  CENAM 1200.495 0.31  LNE-INM 1200.245 0.19  INRIM 1200.475 0.10  PTB 1200.495 0.19

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1200.19 °C, uNRC = 0.27 °C

Lamp S/N 644C, TNRC = 1200.33 °C, uNRC = 0.27 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1300 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL 1301.992 0.14  NMIA 1302.032 0.02  KRISS 1301.982 0.14  NIM 1302.032 0.18  A*STAR 1301.402 0.21  NMIJ 1302.252 0.18  VNIIM

1301.512

0.47

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1302.378 0.14  NMIA 1302.428 0.02  KRISS 1302.518 0.14  NIM 1302.708 0.18  A*STAR 1301.468 0.21  NMIJ 1302.628 0.18  VNIIM

1301.628

0.47

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1300.182 0.23  NIST 1300.952 0.35  CENAM 1300.672 0.37  LNE-INM 1299.882 0.22  INRIM 1300.812 0.12  PTB 1300.752 0.21 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1300.185 0.23  NIST 1300.735 0.35  CENAM 1300.595 0.37  LNE-INM 1300.495 0.22  INRIM 1300.675 0.12  PTB 1300.655 0.21

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1300.20 °C, uNRC = 0.33 °C

Lamp S/N 644C, TNRC = 1300.34 °C, uNRC = 0.33 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1400 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL 1402.315 0.16  NMIA 1402.395 0.02  KRISS 1402.315 0.16  NIM 1402.375 0.21  A*STAR 1401.645 0.23  NMIJ 1402.655 0.21  VNIIM

1401.685

0.60

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1402.715 0.16  NMIA 1402.815 0.02  KRISS 1402.915 0.16  NIM 1403.135 0.21  A*STAR 1401.705 0.23  NMIJ 1403.085 0.21  VNIIM

1401.675

0.60

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1400.245 0.28  NIST 1401.165 0.39  CENAM 1400.715 0.44  LNE-INM 1400.005 0.26  INRIM 1400.955 0.14  PTB 1400.795 0.24 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1400.165 0.28  NIST 1400.785 0.39  CENAM 1400.615 0.44  LNE-INM 1400.405 0.26  INRIM 1400.705 0.14  PTB 1400.635 0.24

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1400.31 °C, uNRC = 0.38 °C

Lamp S/N 644C, TNRC = 1400.31 °C, uNRC = 0.38 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1500 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL 1502.581 0.17  NMIA 1502.671 0.03  KRISS 1502.601 0.19  NIM 1502.671 0.24  A*STAR 1501.821 0.26  NMIJ 1502.991 0.24  VNIIM

1502.161

0.73

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1503.015 0.17  NMIA 1503.205 0.03  KRISS 1503.215 0.19  NIM 1503.515 0.24  A*STAR 1501.835 0.26  NMIJ 1503.475 0.24  VNIIM

1501.775

0.73

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1500.242 0.32  NIST 1501.312 0.44  CENAM 1500.872 0.50  LNE-INM 1499.922 0.30  INRIM 1501.042 0.16  PTB 1500.882 0.26 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1500.239 0.32  NIST 1500.939 0.44  CENAM 1500.709 0.50  LNE-INM 1500.499 0.30  INRIM 1500.829 0.16  PTB 1500.649 0.26

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1500.50 °C, uNRC = 0.40 °C

Lamp S/N 644C, TNRC = 1500.59 °C, uNRC = 0.40 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1600 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL

1602.854

0.19  NMIA 1602.984 0.03  KRISS 1602.914 0.21  NIM 1602.964 0.27  A*STAR 1601.974 0.30  NMIJ 1603.364 0.27  VNIIM

1602.904

0.86

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1603.051 0.19  NMIA 1603.251 0.03  KRISS 1603.271 0.21  NIM 1603.641 0.27  A*STAR 1601.891 0.30  NMIJ 1603.651 0.27  VNIIM

1602.071

0.86

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1600.304 0.37  NIST 1601.604 0.49  CENAM 1600.884 0.60  LNE-INM 1599.904 0.34  INRIM 1601.084 0.18  PTB 1600.984 0.29 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1600.227 0.37  NIST 1601.187 0.49  CENAM 1600.717 0.60  LNE-INM 1600.477 0.34  INRIM 1600.777 0.18  PTB 1600.627 0.29

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1600.61 °C, uNRC = 0.45 °C

Lamp S/N 644C, TNRC = 1600.69 °C, uNRC = 0.45 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

  • Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.

CCT-K5, CCT-K5.1, APMP.T-K5 and EUROMET.T-K5

MEASURAND : Temperature
NOMINAL TEMPERATURE : Tnom = 1700 °C

  • Key comparison CCT-K5

Four Tungsten-strip lamps were used as transfer standards for radiance temperature measurements at specific currents corresponding to each nominal temperature Tnom. To shorten the measurement time significantly the set of transfer standards was split in two sets of two lamps for simultaneous comparisons in two loops. The pilot of each loop measured both lamp sets in order to establish a linkage mechanism described on page 19 of the CCT-K5 Final Report.

Ti : temperature value measured by laboratory i
ui : standard uncertainty of Ti

Lamp S/N C564

 Lab i
Ti / °C ui / °C  VSL

1703.310

0.21  NMIA 1703.510 0.04  KRISS 1703.410 0.25  NIM 1703.480 0.31  A*STAR 1702.320 0.34  NMIJ 1703.930 0.31  VNIIM

1703.930

0.99

Lamp S/N C681

 Lab i
Ti / °C ui / °C  VSL 1703.010 0.21  NMIA 1703.340 0.04  KRISS 1703.380 0.25  NIM 1703.810 0.31  A*STAR 1701.690 0.34  NMIJ 1703.770 0.31  VNIIM

1702.930

0.99

Lamp S/N C860

 Lab i
Ti / °C ui / °C  NPL 1700.290 0.41  NIST 1701.530 0.55  CENAM 1700.730 0.65  LNE-INM 1699.710 0.38  INRIM 1700.970 0.21  PTB 1700.870 0.32 Lamp S/N C864

 Lab i
Ti / °C ui / °C  NPL 1700.300 0.41  NIST 1701.190 0.55  CENAM 1700.650 0.65  LNE-INM 1700.400 0.38  INRIM 1700.760 0.21  PTB 1700.650 0.32

  • Key comparison CCT-K5.1

This is a bilateral comparison between the PTB and the NRC.

TNRC : temperature value measured at the NRC
uNRC : standard uncertainty of TNRC

Lamp S/N C598, TNRC = 1700.66 °C, uNRC = 0.50 °C

Lamp S/N 644C, TNRC = 1700.72 °C, uNRC = 0.50 °C

  • Key comparison APMP.T-K5

Laboratory individual measurements of APMP.T-K5 participants are given in Appendix B of the APMP.T-K5 Final Report both in tabulated and in graphical forms. There were taken between 1997 and 2000.

Key comparison EUROMET.T-K5

This comparison involved eight participants and was carried out from October 1999 to February 2001.
The two transfer standards were Lamp S/N C564 and Lamp S/N C681 already used in CCT-K5.
The individual laboratory measurements and their uncertainties are given in Tables 5 to 11 of the EUROMET.T-K5 Final Report.