CCM.P-K15 and CCM.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

1.0 × 10-4 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

3.0 × 10-4 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

9.0 × 10-4 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

3.0 × 10-3 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

9.0 × 10-3 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-2 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

3.0 × 10-2 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-2 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

9.0 × 10-2 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       0.3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

0.3 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

1.0 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 9.0E-03 3.9E-02
CENAM -9.3E-03 7.7E-02
NIST -2.0E-04 2.0E-02
INRIM -4.0E-04 2.9E-02
PTB -6.5E-03 2.7E-02
METAS 2.0E-03 1.7E-02
NIM 9.0E-04 1.5E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.5E-03 1.7E-02
CENAM 5.4E-03 2.5E-02
NIST -1.9E-03 7.0E-03
INRIM 6.0E-04 1.9E-02
PTB -3.3E-03 1.3E-02
METAS 1.1E-03 9.2E-03
NIM 1.2E-03 7.1E-03
NIMT -9.8E-03 3.2E-02
UME -6.4E-03 1.7E-02
IMT -4.4E-03 1.6E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 5.8E-03 1.1E-02
KRISS -1.9E-03 2.0E-02
CENAM 2.8E-03 1.3E-02
NIST -4.0E-04 3.5E-03
INRIM -7.0E-04 1.5E-02
PTB -3.2E-03 1.0E-02
METAS -9.0E-04 9.7E-03
NIM 1.2E-03 4.7E-03
NIMT -9.0E-03 1.9E-02
UME -4.1E-03 1.5E-02
IMT -3.6E-03 1.4E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.5E-03 1.1E-02
KRISS -1.2E-03 2.0E-02
CENAM 2.0E-04 1.1E-02
NIST -4.0E-04 3.9E-03
INRIM 0.0E+00 1.2E-02
PTB -2.9E-03 9.9E-03
METAS -3.0E-04 5.7E-03
NIM 1.0E-03 3.9E-03
NIMT -8.6E-03 1.8E-02
UME -4.8E-03 1.4E-02
IMT -2.8E-03 1.4E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.0E-03 1.1E-02
KRISS -2.9E-03 2.0E-02
CENAM 1.0E-04 5.4E-03
NIST -2.0E-04 3.4E-03
INRIM -5.0E-04 1.0E-02
PTB -1.6E-03 9.5E-03
METAS -5.0E-04 5.9E-03
NIM 6.0E-04 3.6E-03
NIMT -7.0E-03 1.6E-02
UME -4.6E-03 1.4E-02
IMT -2.5E-03 1.3E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-2 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.0E-03 1.0E-02
KRISS -6.2E-03 2.1E-02
CENAM -4.0E-04 4.9E-03
NIST -2.0E-04 3.5E-03
INRIM -4.0E-04 1.0E-02
PTB -2.0E-03 9.8E-03
METAS 3.0E-04 5.9E-03
NIM 1.3E-03 3.4E-03
NIMT -7.2E-03 1.8E-02
UME -3.7E-03 1.4E-02
IMT -1.4E-03 1.3E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-2 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.5E-03 1.1E-02
KRISS -1.3E-02 2.4E-02
CENAM -1.0E-04 5.0E-03
NIST 6.0E-04 3.8E-03
INRIM -3.0E-04 7.5E-03
PTB -2.6E-03 6.1E-03
METAS 3.0E-04 6.1E-03
NIM 1.3E-03 2.8E-03
NIMT -7.5E-03 1.3E-02
UME -4.1E-03 1.0E-02
IMT -2.4E-03 9.3E-03
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       0.3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 5.3E-03 1.1E-02
CENAM -1.1E-03 5.0E-03
NIST 1.0E-03 4.1E-03
INRIM -8.0E-04 6.7E-03
PTB -2.7E-03 5.5E-03
METAS 9.0E-04 4.3E-03
NIM 3.0E-04 3.1E-03
NIMT -6.3E-03 1.3E-02
UME -4.1E-03 9.3E-03
IMT -1.7E-03 8.8E-03
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 4.6E-03 1.1E-02
CENAM 2.0E-04 4.1E-03
NIST 1.3E-03 3.1E-03
INRIM -3.2E-03 7.5E-03
PTB -3.6E-03 5.9E-03
METAS 1.0E-04 3.9E-03
NIM 4.0E-04 3.0E-03
NIMT -5.6E-03 1.3E-02
UME -4.0E-03 9.6E-03
IMT -1.7E-03 9.1E-03
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1
Comparison
Comparison close
CC comparison
Linked comparison
CCM.P-K15.1
Metrology area, Sub-field Mass and related quantities, Pressure
Description Comparison of absolute pressure in the range 0.1 mPa to 1 Pa
Time of measurements 2011 - 2012
Status Approved for equivalence
Final Reports of the comparisons
References

CCM.P-K15.1 Technical Protocol

Measurand Pressure: 0.1 mPa to 1 Pa
Transfer device Two spinning rotor gauges
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by CCM (Consultative Committee for Mass and Related Quantities)
Comments Comparison renamed from CCM.P-K14.1 to CCM.P-K15.1 on 28 November 2016
Results published on 23 January 2017
Absolute pressure
Pilot institute METAS
Federal Institute of Metrology
Switzerland
Contact person C. Wüthrich

+41 (0) 31 32 33 423
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
METAS

Federal Institute of Metrology, Switzerland, EURAMET

NIM

National Institute of Metrology, China, APMP

CCM.P-K15 and CCM.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 x 10-4 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-4 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-4 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-3 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-3 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-2 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-2 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       0.3 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15 and CCM.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 x 10-4 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

 

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-4 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-4 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-3 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-3 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-2 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-2 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       0.3 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.