CCM.P-K15 and CCM.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

1.0 × 10-4 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

3.0 × 10-4 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

9.0 × 10-4 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

3.0 × 10-3 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

9.0 × 10-3 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-2 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

3.0 × 10-2 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-2 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

9.0 × 10-2 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       0.3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

0.3 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

1.0 Pa
Di (relative)
Participating institutes

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 9.0E-03 3.9E-02
CENAM -9.3E-03 7.7E-02
NIST -2.0E-04 2.0E-02
INRIM -4.0E-04 2.9E-02
PTB -6.5E-03 2.7E-02
METAS 2.0E-03 1.7E-02
NIM 9.0E-04 1.5E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.5E-03 1.7E-02
CENAM 5.4E-03 2.5E-02
NIST -1.9E-03 7.0E-03
INRIM 6.0E-04 1.9E-02
PTB -3.3E-03 1.3E-02
METAS 1.1E-03 9.2E-03
NIM 1.2E-03 7.1E-03
NIMT -9.8E-03 3.2E-02
UME -6.4E-03 1.7E-02
IMT -4.4E-03 1.6E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-4 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 5.8E-03 1.1E-02
KRISS -1.9E-03 2.0E-02
CENAM 2.8E-03 1.3E-02
NIST -4.0E-04 3.5E-03
INRIM -7.0E-04 1.5E-02
PTB -3.2E-03 1.0E-02
METAS -9.0E-04 9.7E-03
NIM 1.2E-03 4.7E-03
NIMT -9.0E-03 1.9E-02
UME -4.1E-03 1.5E-02
IMT -3.6E-03 1.4E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.5E-03 1.1E-02
KRISS -1.2E-03 2.0E-02
CENAM 2.0E-04 1.1E-02
NIST -4.0E-04 3.9E-03
INRIM 0.0E+00 1.2E-02
PTB -2.9E-03 9.9E-03
METAS -3.0E-04 5.7E-03
NIM 1.0E-03 3.9E-03
NIMT -8.6E-03 1.8E-02
UME -4.8E-03 1.4E-02
IMT -2.8E-03 1.4E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.0E-03 1.1E-02
KRISS -2.9E-03 2.0E-02
CENAM 1.0E-04 5.4E-03
NIST -2.0E-04 3.4E-03
INRIM -5.0E-04 1.0E-02
PTB -1.6E-03 9.5E-03
METAS -5.0E-04 5.9E-03
NIM 6.0E-04 3.6E-03
NIMT -7.0E-03 1.6E-02
UME -4.6E-03 1.4E-02
IMT -2.5E-03 1.3E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-2 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.0E-03 1.0E-02
KRISS -6.2E-03 2.1E-02
CENAM -4.0E-04 4.9E-03
NIST -2.0E-04 3.5E-03
INRIM -4.0E-04 1.0E-02
PTB -2.0E-03 9.8E-03
METAS 3.0E-04 5.9E-03
NIM 1.3E-03 3.4E-03
NIMT -7.2E-03 1.8E-02
UME -3.7E-03 1.4E-02
IMT -1.4E-03 1.3E-02
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-2 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 6.5E-03 1.1E-02
KRISS -1.3E-02 2.4E-02
CENAM -1.0E-04 5.0E-03
NIST 6.0E-04 3.8E-03
INRIM -3.0E-04 7.5E-03
PTB -2.6E-03 6.1E-03
METAS 3.0E-04 6.1E-03
NIM 1.3E-03 2.8E-03
NIMT -7.5E-03 1.3E-02
UME -4.1E-03 1.0E-02
IMT -2.4E-03 9.3E-03
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       0.3 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 5.3E-03 1.1E-02
CENAM -1.1E-03 5.0E-03
NIST 1.0E-03 4.1E-03
INRIM -8.0E-04 6.7E-03
PTB -2.7E-03 5.5E-03
METAS 9.0E-04 4.3E-03
NIM 3.0E-04 3.1E-03
NIMT -6.3E-03 1.3E-02
UME -4.1E-03 9.3E-03
IMT -1.7E-03 8.8E-03
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 Pa

 

Degrees of equivalence represented by Di = (xI - xR)/xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in non-dimensional units.

LABi Di Uneg,i Upos,i
Pa/Pa
Pa/Pa
Pa/Pa
NMIJ 4.6E-03 1.1E-02
CENAM 2.0E-04 4.1E-03
NIST 1.3E-03 3.1E-03
INRIM -3.2E-03 7.5E-03
PTB -3.6E-03 5.9E-03
METAS 1.0E-04 3.9E-03
NIM 4.0E-04 3.0E-03
NIMT -5.6E-03 1.3E-02
UME -4.0E-03 9.6E-03
IMT -1.7E-03 9.1E-03
Participating institutes
CCM.P-K15
CCM.P-K15.1
EURAMET.M.P-K15.1
Comparison
Comparison close
CC comparison
Linked comparison
EURAMET.M.P-K15.1
Metrology area, Sub-field Mass and related quantities, Pressure
Description Comparison of absolute pressure in the range from 0.0003 Pa to 1 Pa
Time of measurements 2017
Status Approved for equivalence
Final Reports of the comparisons
References

EURAMET.M.P-K15.1 Technical Protocol

Measurand Pressure
0.0003 Pa, 0.0009 Pa, 0.003 Pa, 0.009 Pa, 0.03 Pa, 0.09 Pa, 0.3 Pa and 1 Pa
Parameters Nitrogen gas with 99.99% purity used for the calibration
Temperature at [23 ± 2] °C
Transfer device Two spinning rotor gauges (SRG)
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET.M.1405
Comments Results published on 2 December 2020

Results at 1 Pa are also linked to CCM.P-K4
Pilot institute PTB
Physikalisch-Technische Bundesanstalt
Germany
Contact person K. Jousten

+49 (0) 30 34 81 7262
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

CEM

Centro Español de Metrologia, Spain, EURAMET

NIMT

National Institute of Metrology (Thailand), Thailand, APMP

UME

TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET

CCM.P-K15 and CCM.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 x 10-4 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-4 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-4 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-3 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-3 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-2 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-2 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       0.3 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 Pa

 

CCM.P-K15

The key comparison reference value, xR, is adopted as the nominal value of the pressure. The standard uncertainty of the key comparison reference value, uR, is calculated according to Eq. (40) of the CCM.P-K15 Final Report, at each pressure.

For each pressure, the degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) / xR, the relative deviation from the reference value, and Ui , its relative expanded uncertainty (k = 2) calculated according to Eq. (41) of the CCM.P-K15 Final Report.

 

Linking CCM.P-K15.1 to CCM.P-K15

The results of CCM.P-K15.1 were linked to CCM.P-K15 via METAS who participated in both comparisons.

 

Linking EURAMET.M.P-K15.1 to CCM.P-K15

The results of EURAMET.M.P-K15.1 were linked to CCM.P-K15 via PTB who participated in both comparisons.

CCM.P-K15 and CCM.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 x 10-4 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

 

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-4 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-4 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-3 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-3 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       3.0 x 10-2 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       9.0 x 10-2 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       0.3 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.

CCM.P-K15, CCM.P-K15.1 and EURAMET.M.P-K15.1

 

MEASURAND         Pressure

NOMINAL VALUE       1.0 Pa

 

CCM.P-K15

The laboratory measurement obtained in CCM.P-K15 are listed in Sec. 5 of the CCM.P-K15 Final report.

 

CCM.P-K15.1

The laboratory measurement obtained in CCM.P-K15.1 are listed in Sec. 5 of the CCM.P-K15.1 Final report.

 

EURAMET.M.P-K15.1

The laboratory measurement obtained in ERURAMET.M.P-K15.1 are listed in Sec.  of the EURAMET.M.P-K15.1 Final report.