CCM.FF-K3.2011 and CCM.FF-K3.2011.1

MEASURAND                   Air speed
NOMINAL VALUE           0.5 m/s, 1 m/s, 2 m/s, 5 m/s, 10 m/s, 15 m/s, 20 m/s, 30 m/s and 40 m/s


TRANSFER STANDARD  Ultrasonic anemometer

Degrees of equivalence: offset Di and expanded uncertainty (k = 2) Ui

NOMINAL VALUE           0.5 m/s

0.5 m/s
Di
Participating institutes

NOMINAL VALUE           1 m/s

1 m/s
Di
Participating institutes

NOMINAL VALUE           2 m/s

2 m/s
Di
Participating institutes

NOMINAL VALUE           5 m/s

5 m/s
Di
Participating institutes

NOMINAL VALUE           10 m/s

10 m/s
Di
Participating institutes

NOMINAL VALUE           15 m/s

15 m/s
Di
Participating institutes

NOMINAL VALUE           20 m/s

20 m/s
Di
Participating institutes

NOMINAL VALUE           30 m/s

30 m/s
Di
Participating institutes

NOMINAL VALUE           40 m/s

40 m/s
Di
Participating institutes

CCM.FF-K3.2011 and CCM.FF-K3.2011.1

MEASURAND                    Air speed
NOMINAL VALUE            0.5 m/s, 1 m/s, 2 m/s, 5 m/s, 10 m/s, 15 m/s, 20 m/s, 30 m/s and 40 m/s


TRANSFER STANDARD   Laser Doppler anemometer

Degrees of equivalence: offset Di and expanded uncertainty (k = 2) Ui

NOMINAL VALUE           0.5 m/s

0.5 m/s
Di
Participating institutes

NOMINAL VALUE           1 m/s

1 m/s
Di
Participating institutes

NOMINAL VALUE           2 m/s

2 m/s
Di
Participating institutes

NOMINAL VALUE           5 m/s

5 m/s
Di
Participating institutes

NOMINAL VALUE           10 m/s

10 m/s
Di
Participating institutes

NOMINAL VALUE           15 m/s

15 m/s
Di
Participating institutes

NOMINAL VALUE           20 m/s

20 m/s
Di
Participating institutes

NOMINAL VALUE           30 m/s

30 m/s
Di
Participating institutes

NOMINAL VALUE           40 m/s

40 m/s
Di
Participating institutes

CCM.FF-K3.2011

 

OPTIONAL MEASURAND       Laser Doppler anemometer calibration using a primary standard

TRANSFER STANDARD           Laser Doppler anemometer

Degrees of equivalence: offset Di and expanded uncertainty (k = 2) Ui

 

Calibration
Di
Participating institutes

CCM.FF-K3.2011 and CCM.FF-K3.2011.1

MEASURAND                   Air speed
NOMINAL VALUE           0.5 m/s, 1 m/s, 2 m/s, 5 m/s, 10 m/s, 15 m/s, 20 m/s, 30 m/s and 40 m/s


TRANSFER STANDARD  Ultrasonic anemometer

Degrees of equivalence: offset Di and expanded uncertainty (k = 2) Ui

NOMINAL VALUE           0.5 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0051 0.0204
LNE-CETIAT -0.0008 0.0272
VSL -0.0311 0.0492
E+E -0.0221 0.0198
NMIJ -0.0092 0.0433
NIM -0.0003 0.0376
CMS -0.0014 0.0179
NIST 0.0131 0.0178
VNIIM -0.0109 0.0309
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           1 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0023 0.012
LNE-CETIAT 0.0066 0.0163
VSL -0.0147 0.013
E+E -0.01 0.0122
NMIJ -0.0048 0.0185
NIM -0.0009 0.0188
CMS 0.0022 0.0106
NIST 0.0109 0.0117
VNIIM -0.005 0.0186
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           2 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0088 0.0106
LNE-CETIAT 0.003 0.0132
VSL -0.0004 0.0134
E+E -0.0095 0.0115
NMIJ -0.0033 0.0115
NIM 0.0062 0.0171
CMS 0.0047 0.0109
NIST 0.0027 0.0101
INRIM -0.0001 0.0136
VNIIM -0.0036 0.0164
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           5 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0034 0.0046
LNE-CETIAT -0.0011 0.0072
VSL -0.0028 0.0101
E+E -0.0053 0.0066
NMIJ 0.0001 0.0059
NIM -0.001 0.0084
CMS 0.0005 0.0054
NIST 0.0047 0.0046
INRIM -0.0004 0.0092
VNIIM -0.0035 0.008
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           10 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0018 0.0042
LNE-CETIAT 0.0014 0.0066
VSL 0.0026 0.0102
E+E -0.0021 0.0064
NMIJ 0.0012 0.0048
NIM 0.0029 0.0066
CMS 0.0026 0.0053
NIST -0.0017 0.0043
INRIM -0.0074 0.0093
VNIIM -0.0018 0.0074
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           15 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0008 0.0041
LNE-CETIAT 0.0013 0.0063
VSL -0.0003 0.0101
E+E -0.0005 0.0064
NMIJ 0.001 0.0052
NIM -0.0021 0.0064
CMS 0.0071 0.0056
NIST -0.0019 0.0043
INRIM -0.0082 0.0093
VNIIM 0.0003 0.0072
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           20 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0004 0.004
LNE-CETIAT 0.0016 0.0062
VSL -0.0048 0.0103
E+E -0.0005 0.0063
NMIJ -0.0005 0.0052
NIM -0.002 0.0065
CMS 0.0087 0.0055
NIST -0.0019 0.0043
INRIM -0.0079 0.0092
VNIIM -0.0026 0.0071
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           30 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.0009 0.0039
LNE-CETIAT 0.0039 0.006
VSL -0.0069 0.013
E+E 0.0001 0.0063
NMIJ -0.0003 0.0044
NIM -0.0008 0.007
NIST -0.0002 0.0043
INRIM -0.0061 0.0093
VNIIM -0.0027 0.0069
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           40 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.0004 0.0038
LNE-CETIAT 0.0047 0.006
VSL -0.0111 0.015
E+E -0.0002 0.0062
NMIJ -0.0013 0.0041
NIST 0.001 0.0042
VNIIM -0.002 0.0068
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

CCM.FF-K3.2011 and CCM.FF-K3.2011.1

MEASURAND                    Air speed
NOMINAL VALUE            0.5 m/s, 1 m/s, 2 m/s, 5 m/s, 10 m/s, 15 m/s, 20 m/s, 30 m/s and 40 m/s


TRANSFER STANDARD   Laser Doppler anemometer

Degrees of equivalence: offset Di and expanded uncertainty (k = 2) Ui

NOMINAL VALUE           0.5 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0003 0.0132
LNE-CETIAT 0.0005 0.0214
VSL -0.0411 0.0449
E+E 0.0012 0.0125
NIM 0.0026 0.007
CMS -0.0028 0.0043
NIST 0.0024 0.0048
VNIIM -0.0103 0.02
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           1 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.003 0.0079
LNE-CETIAT 0.0065 0.0131
VSL -0.0152 0.0101
E+E 0.0002 0.0082
NIM -0.0011 0.007
CMS -0.0007 0.0055
NIST 0.0032 0.0049
VNIIM -0.0041 0.0128
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           2 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0002 0.0055
LNE-CETIAT 0.0032 0.009
VSL -0.0013 0.0103
E+E -0.0009 0.0063
NMIJ -0.0026 0.0062
NIM 0.0004 0.0036
CMS -0.0004 0.0066
NIST -0.0003 0.0072
INRIM 0.0034 0.0101
VNIIM -0.0019 0.0093
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           5 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.0021 0.0041
LNE-CETIAT -0.0005 0.0066
VSL -0.0043 0.0099
E+E -0.0012 0.0052
NMIJ 0.0006 0.0049
NIM -0.0006 0.0027
CMS -0.0018 0.005
NIST 0.0007 0.0038
INRIM 0.0004 0.0089
VNIIM -0.0029 0.0073
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           10 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.0017 0.0036
LNE-CETIAT -0.0005 0.0058
VSL 0.0063 0.01
E+E -0.0022 0.0048
NMIJ -0.0005 0.0044
NIM 0.0002 0.0025
CMS 0.0003 0.0054
NIST -0.0009 0.0038
INRIM -0.0088 0.0089
VNIIM -0.0013 0.0066
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           15 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.0018 0.0034
LNE-CETIAT 0.0001 0.0055
VSL 0.002 0.0102
E+E -0.0019 0.0046
NMIJ 0.0001 0.0047
NIM -0.0007 0.0037
CMS 0.0006 0.0051
NIST -0.0006 0.0037
INRIM -0.0077 0.009
VNIIM -0.0024 0.0063
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           20 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.0017 0.0034
LNE-CETIAT -0.0003 0.0054
VSL 0.0024 0.0117
E+E -0.002 0.0046
NMIJ -0.0007 0.0047
NIM 0.0006 0.0027
CMS -0.0008 0.005
NIST -0.0009 0.0038
INRIM -0.0066 0.0091
VNIIM -0.0006 0.0062
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           30 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.0014 0.0032
LNE-CETIAT 0.0001 0.0052
VSL 0.0025 0.0139
E+E -0.0015 0.0045
NMIJ -0.0012 0.0034
NIM 0.0014 0.0049
NIST -0.0002 0.0037
INRIM -0.0066 0.009
VNIIM -0.0032 0.0061
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

NOMINAL VALUE           40 m/s

LABi Di Uneg,i Upos,i
-
-
-
PTB 0.0015 0.0031
LNE-CETIAT 0.0016 0.0052
VSL -0.0024 0.0149
E+E -0.0007 0.0044
NMIJ -0.0011 0.0025
NIST 0.0005 0.0037
VNIIM -0.0027 0.006
Participating institutes
CCM.FF-K3.2011
CCM.FF-K3.2011.1

CCM.FF-K3.2011

 

OPTIONAL MEASURAND       Laser Doppler anemometer calibration using a primary standard

TRANSFER STANDARD           Laser Doppler anemometer

Degrees of equivalence: offset Di and expanded uncertainty (k = 2) Ui

 

LABi Di Uneg,i Upos,i
-
-
-
PTB -0.0006 0.0015
LNE-CETIAT -0.0002 0.0009
NMIJ 0.0003 0.0018
NIM -0.0005 0.0023
CMS 0.0024 0.0025
Participating institutes
CCM.FF-K3.2011
SIM.M.FF-K3.2023
Metrology area, Sub-field Mass and related quantities, Fluid Flow
Description Bilateral key comparison of airspeed standards.
Time of measurements 2023 - 2025
Status Report in progress, draft A
Measurand Air
Transfer device Ultrasonic Anemometer, Laser Doppler Anemometer, Pitot Tube
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by SIM (Inter-American Metrology System)
Comments Bilateral key comparison between NIST (pilot laboratory) and INMETRO, focusing on determining the degree of equivalence of national airspeed standards. Three transfer standards were employed: a Laser Doppler Anemometer (LDA), an ultrasonic anemometer, and a Pitot tube.
Pilot institute NIST
National Institute of Standards and Technology
United States
Contact person Joey Boyd

3019752943
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NIST

National Institute of Standards and Technology, United States, SIM

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

CCM.FF-K3.2011 and CCM.FF-K3.2011.1

MEASURAND                   Air speed
NOMINAL VALUE           0.5 m/s, 1 m/s, 2 m/s, 5 m/s, 10 m/s, 15 m/s, 20 m/s, 30 m/s and 40 m/s


TRANSFER STANDARD  Ultrasonic anemometer

 

CCM.FF-K3.2011

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms, both dimensionless: Di = (xi - xR) and Ui (k=2) its expanded uncertainty, with Ui = 2 [ (ULab / 2)2 - (U/ 2)]1/2.

 

For each nominal air speed value, the key comparison reference value, xR, is computed as the weighted mean of the participants' results. Its expanded uncertainty (k = 2), UR, is computed as explained in section 7.2.1 of the Final Report, noting an additional contribution due to the instability of the transfer standard was included when calculating the uncertainty.

 


Linking CCM.FF-K3.2011.1 to CCM.FF-K3.2011

The results obtained by VNIIM in the bilateral comparison CCM.FF-K3.2011.1 were linked to CCM.FF-K3.2011 via PTB who participated in both comparisons.

CCM.FF-K3.2011 and CCM.FF-K3.2011.1

MEASURAND                    Air speed
NOMINAL VALUE            0.5 m/s, 1 m/s, 2 m/s, 5 m/s, 10 m/s, 15 m/s, 20 m/s, 30 m/s and 40 m/s


TRANSFER STANDARD   Laser Doppler anemometer

 

 

CCM.FF-K3.2011

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms, both dimensionless: Di = (xi - xR) and Ui (k=2) its expanded uncertainty, with Ui = 2 [ (ULab i / 2)2 - (UR / 2)2 ]1/2.

 

For each nominal air speed value, the key comparison reference value, xR, is computed as the weighted mean of the participants' results. Its expanded uncertainty (k = 2), UR, is computed as explained in section 7.2.2 of the Final Report, noting an additional contribution due to the instability of the transfer standard was included when calculating the uncertainty.

 

 

Linking CCM.FF-K3.2011.1 to CCM.FF-K3.2011

The results obtained by VNIIM in the bilateral comparison CCM.FF-K3.2011.1 were linked to CCM.FF-K3.2011 via PTB who participated in both comparisons.

 

CCM.FF-K3.2011

 

OPTIONAL MEASURAND       Laser Doppler anemometer calibration using a primary standard

TRANSFER STANDARD           Laser Doppler anemometer

 

The key comparison reference value, xR, and iIts expanded uncertainty (k = 2), UR, are computed for the overall dataset as explained in Section 8 of the CCM.FF-K3.2011 Final Report.

                                                     xR = 0.9988         UR = 0.0008

 

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms, both dimensionless: Di = (xi - xR) and Ui (k=2) its expanded uncertainty, with Ui = 2[ (ULab / 2)2 - (UR / 2)2 ]1/2.

CCM.FF-K3.2011 and CCM.FF-K3.2011.1

MEASURAND                   Air speed
NOMINAL VALUE           0.5 m/s, 1 m/s, 2 m/s, 5 m/s, 10 m/s, 15 m/s, 20 m/s, 30 m/s and 40 m/s


TRANSFER STANDARD  Ultrasonic anemometer

 

CCM.FF-K3.2011

 

xi                measurement result reported by laboratory i

ULab i         expanded uncertainty (k = 2) of xi

The measurement result, xi, reported by laboratory i is the ratio of the reference air speed measured by the laboratory which was obtained from five repeated measurements, to the air speed reading of the transfer standard. xi is a dimensionless quantity.

 

 

 

CCM.FF-K3.2011.1

 

The results obtained in CCM.FF-K3.2011.1 are listed in Section 7 of the CCM.FF-K3.2011.1 Final Report.

 

 

CCM.FF-K3.2011 and CCM.FF-K3.2011.1

MEASURAND                    Air speed
NOMINAL VALUE            0.5 m/s, 1 m/s, 2 m/s, 5 m/s, 10 m/s, 15 m/s, 20 m/s, 30 m/s and 40 m/s


TRANSFER STANDARD   Laser Doppler anemometer

 

CCM.FF-K3.2011

 

xi               measurement result reported by laboratory i
ULab i        expanded uncertainty (k = 2) of xi

The measurement result, xi, reported by laboratory i is the ratio of the reference air speed measured by the laboratory which was obtained from five repeated measurements, to the air speed reading of the transfer standard. xi is a dimensionless quantity.

 

 

 

CCM.FF-K3.2011.1

 

The results obtained in CCM.FF-K3.2011.1 are listed in Section 7 of the CCM.FF-K3.2011.1 Final Report.

CCM.FF-K3.2011

 

OPTIONAL MEASURAND       Laser Doppler anemometer calibration using a primary standard

TRANSFER STANDARD           Laser Doppler anemometer

 

The measurement results obtained by the PTB, LNE-CETIAT, NMIJ, NIM and CMS are listed in Appendix A of the CCM.FF-K3.2011 Final Report.