Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cr-51, APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51

MEASURAND : Equivalent activity of Cr-51

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Cr-51
Di / MBq
Participating institutes

Results in absolute values

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Cr-51
Di / kBq/MBq
Participating institutes

Results in relative values

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cr-51, APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51

MEASURAND : Equivalent activity of Cr-51

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
NMIJ -0.9 3.1
NPL -1.0 6.7
KRISS 0.4 4.7
POLATOM -4.4 5.9
VNIIM 0.8 2.5
LNE-LNHB -1.6 3.2
Participating institutes
BIPM.RI(II)-K1.Cr-51
APMP.RI(II)-K2.Cr-51
EURAMET.RI(II)-K2.Cr-51

Results in absolute values

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
NMIJ -1.9 6.3
NPL -2 14
KRISS 0.7 9.7
POLATOM -9 12
VNIIM 1.6 5.1
LNE-LNHB -3.2 6.6
Participating institutes
BIPM.RI(II)-K1.Cr-51
APMP.RI(II)-K2.Cr-51
EURAMET.RI(II)-K2.Cr-51

Results in relative values

BIPM.RI(II)-K1.Cr-51
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Cr-51
Time of measurements 1977 -
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
First results published on 7 March 2003
Most recent results published on 16 February 2024
SIR equivalent activity of the radionuclide 51Cr:
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

ANSTO

Australian Nuclear Science and Technology Organisation, Australia, APMP

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

Report in progress, Draft B

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

NIST

National Institute of Standards and Technology, United States, SIM

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NPL

National Physical Laboratory, United Kingdom, EURAMET

POLATOM

Radioisotope Centre Polatom, Poland, EURAMET

report in progress draft B

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cr-51, APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51

MEASURAND : Equivalent activity of Cr-51

 

BIPM.RI(II)-K1.Cr-51

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is  xR = 488.05 MBq with a standard uncertainty uR = 0.41 MBq

(see Section 4.1 of the BIPM.RI(II)-K1.I-Cr-51 Final Report).  The value xi is taken as the equivalent activity for a laboratory i.

 

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (x- xR) and Uits expanded uncertainty (k = 2), both expressed in MBq, and

Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

 

Linking APMP.RI(II)-K2.Cr-51 to BIPM.RI(II)-K1.Cr-51

The value xi is the equivalent activity for laboratory i participant in APMP.RI(II)-K2.Cr-51 having been normalized using the value of the NMIJ as the linking laboratory.

The degree of equivalence of laboratory i participant in APMP.RI(II)-K2.Cr-51 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value.

 

Linking EURAMET.RI(II)-K2.Cr-51 to BIPM.RI(II)-K1.Cr-51

The value xi is the equivalent activity for laboratory i participant in EURAMET.RI(II)-K2.Cr-51 having been normalized to the value of the POLATOM as the linking laboratory.

The degree of equivalence of laboratory i participant in EURAMET.RI(II)-K2.Cr-511 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value..

These statements make it possible to extend the BIPM.RI(II)-K1.Cr-51 matrices of equivalence to all participants in the APMP.RI(II)-K2.Cr-51 and  EURAMET.RI(II)-K2.Cr-51 comparisons

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cr-51, APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51

MEASURAND : Equivalent activity of Cr-51

 

xi   measured mass fraction reported by laboratory iLabi

ui   combined standard uncertainty of xi

 

 

BIPM.RI(II)-K1.Cr-51

  Labi xi ui Date of measurement
    kBq kBq  
  NMIJ 487.1 1.6 2004-03-15
  NPL 487.0 3.4 2004-05-28
  KRISS 488.4 2.4 2012-01-27
  POLATOM 483.6 3.0 2022-06-03

 

APMP.RI(II)-K2.Cr-51

  Labi xi ui Date of measurement
    kBq kBq
  VNIIM 488.8 1.2 2004

 

EURAMET.RI(II)-K2.Cr-51

  Labi xi ui Date of measurement
    kBq kBq
  LNE-LNHB 486.5 1.6 2022