BIPM.RI(II)-K1.Cr-51 and APMP.RI(II)-K2.Cr-51
MEASURAND : SIR equivalent activity of 51Cr
Degrees of equivalence Di and expanded uncertainty Ui (k = 2) expressed in MBq
Red diamonds and black squares: participants in BIPM.RI(II)-K1.Cr-51
Green triangle: participant in APMP.RI(II)-K2.Cr-51 (2004)
The black colour identifies participant's results prior to 1983
Note: the right-hand axis shows approximate relative values only
BIPM.RI(II)-K1.Cr-51 and APMP.RI(II)-K2.Cr-51
MEASURAND : SIR equivalent activity of 51Cr
Lab i | Di | Ui |
/ MBq | / MBq | |
IRMM | -4.1 | 2.5 |
CMI-IIR | 0.7 | 2.7 |
MKEH | -0.7 | 2.6 |
PTB | -0.4 | 2.0 |
NIST | 1.4 | 3.0 |
NMIJ | -0.9 | 3.1 |
NPL | -1.0 | 6.2 |
LNE-LNHB | 1.2 | 2.3 |
VNIIM | 0.3 | 2.6 |
In green: participant in APMP.RI(II)-K2.Cr-51
Results are presented under A4 printable format in Summary Results (.PDF file).
Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
Description | Activity of radionuclide Cr-51 |
Time of measurements | 1977 - |
Status | Continuous, approved for equivalence |
Final Reports of the comparisons | |
Measurand | SIR equivalent activity |
Transfer device | Radioactive solution |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | BIPM (Bureau International des Poids et Mesures) |
Comments | SIR: Système International de Référence, International Reference System APMP.RI(II)-K2.Cr-51 results are linked to those of BIPM.RI(II)-K1.Cr-51 First results published on 7 March 2003 Update and APMP data published on 10 October 2005 Update published on 07 February 2009 SIR equivalent activity of the radionuclide 51Cr: |
Pilot institute |
BIPM
Bureau International des Poids et Mesures BIPM - International Organization |
Contact person | Carine Michotte +33 (0) 1 45 07 70 85 |
Pilot laboratory | |
---|---|
BIPM |
Bureau International des Poids et Mesures, BIPM - International Organization, N/A |
ANSTO |
Australian Nuclear Science and Technology Organisation, Australia, APMP |
1994 |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
2012 |
Report in progress, Draft B |
LNE-LNHB |
Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET |
2006 |
NIST |
National Institute of Standards and Technology, United States, SIM |
1999 |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
2004 |
NPL |
National Physical Laboratory, United Kingdom, EURAMET |
2004 |
POLATOM |
Radioisotope Centre Polatom, Poland, EURAMET |
2022 |
report in progress draft B |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
1998 |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
BIPM.RI(II)-K1.Cr-51 and APMP.RI(II)-K2.Cr-51
MEASURAND : SIR equivalent activity of 51Cr
Key comparison reference value: The SIR reference value for this radionuclide is xR = 488.0 MBq, with a standard uncertainty uR = 0.3 MBq, see section 4.1 of the Final Report dated January 2009. |
The degree of equivalence of each laboratory with respect to the key comparison reference value is given by a pair of terms: Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. With n the number of laboratories, ![]() |
The degree of equivalence between two laboratories is given by a pair of terms: Dij = Di - Dj = (xi - xj) and Uij, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Uij ~ 2(ui2 + uj2)1/2 is used for the computation of the pair-wise degrees of equivalence. |
The value xi is the equivalent activity for laboratory i participant in APMP.RI(II)-K2.Cr-51 having been normalized to the value of NMIJ as the linking laboratory (see Final Report dated October 2005). |
The degree of equivalence of laboratory i participant in APMP.RI(II)-K2.Cr-51 with respect to the key comparison reference value is given by a pair of terms: Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the Matrix of equivalence. |
The degree of equivalence between two laboratories i and j is given by a pair of terms: Dij = Di - Dj and Uij, its expanded uncertainty (k = 2), both expressed in MBq, where the approximation Uij = 2(ui2 + uj2 - 2ful2)1/2 is used (l is the NMIJ and f the correlation coefficient). |
BIPM.RI(II)-K1.Cr-51 and APMP.RI(II)-K2.Cr-51
MEASURAND : SIR equivalent activity of 51Cr
xi : result of measurement carried out in the SIR for the sample submitted by laboratory i
ui : combined standard uncertainty of xi
Lab i | xi / MBq |
ui / MBq |
Date of measurement |
IRMM | 483.9 | 1.1 | 1981-06-17 |
CMI-IIR | 488.7 | 1.4 | 1982-06-30 |
MKEH | 487.3 | 1.3 | 1989-10-09 |
PTB | 487.6 | 0.9 | 1998-04-24 |
NIST | 489.4 | 1.4 | 1999-05-03 |
NMIJ | 487.1 | 1.6 | 2004-03-15 |
NPL | 487.0 | 3.4 | 2004-05-28 |
LNE-LNHB | 489.2 | 1.1 | 2006-07-17 |
xi : result of measurement carried out at laboratory i, converted to the equivalent activity through the NMJ SIR measurement (see Final Report dated October 2005)
ui : combined standard uncertainty of xi
Lab i | xi / MBq |
ui / MBq |
Date of measurement |
VNIIM | 488.3 | 1.2 | 2004 |