BIPM.RI(II)-K1.Cr-51 and APMP.RI(II)-K2.Cr-51

MEASURAND : SIR equivalent activity of 51Cr

Degrees of equivalence Di and expanded uncertainty Ui (k = 2) expressed in MBq

Red diamonds and black squares: participants in BIPM.RI(II)-K1.Cr-51
Green triangle: participant in APMP.RI(II)-K2.Cr-51 (2004)
The black colour identifies participant's results prior to 1983

Note: the right-hand axis shows approximate relative values only

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cr-51 and APMP.RI(II)-K2.Cr-51

MEASURAND : SIR equivalent activity of 51Cr

   
Lab iDiUi
 / MBq/ MBq
IRMM-4.12.5
CMI-IIR0.72.7
MKEH-0.72.6
PTB-0.42.0
NIST1.43.0
NMIJ-0.93.1
NPL-1.06.2
LNE-LNHB1.22.3
VNIIM0.32.6

In green: participant in APMP.RI(II)-K2.Cr-51

Results are presented under A4 printable format in Summary Results (.PDF file).

EURAMET.RI(II)-K2.Cr-51
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description ACTIVITY OF RADIONUCLIDE Cr-51
Time of measurements 2022
Status Approved for equivalence
Final Reports of the comparisons
Measurand Activity per mass
kBq/g
Transfer device Radioactive 51Cr solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET Project 1554
Comments The comparison is organized in a frame of EURAMET Mentoring Scheme Award MSA-002-2022_POLATOM-LNHB
Pilot institute POLATOM
Radioisotope Centre Polatom
Poland
Contact person Ryszard Broda

+48 22 273 19 53
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
POLATOM

Radioisotope Centre Polatom, Poland, EURAMET

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cr-51 and APMP.RI(II)-K2.Cr-51

MEASURAND : SIR equivalent activity of 51Cr

  • Key comparison BIPM.RI(II)-K1.Cr-51

Key comparison reference value:
The SIR reference value for this radionuclide is xR = 488.0 MBq, with a standard uncertainty uR = 0.3 MBq, see section 4.1 of the Final Report dated January 2009.

The degree of equivalence of each laboratory with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. With n the number of laboratories, when each laboratory has contributed to the computation of xR, see Metrologia, 42, 140-144.

The degree of equivalence between two laboratories is given by a pair of terms:
Dij = Di - Dj = (xi - xj) and Uij, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Uij ~ 2(ui2 + uj2)1/2 is used for the computation of the pair-wise degrees of equivalence.
  • Linking APMP.RI(II)-K2.Cr-51 to BIPM.RI(II)-K1.Cr-51

The value xi is the equivalent activity for laboratory i participant in APMP.RI(II)-K2.Cr-51 having been normalized to the value of NMIJ as the linking laboratory (see Final Report dated October 2005).

The degree of equivalence of laboratory i participant in APMP.RI(II)-K2.Cr-51 with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq.
The approximation Ui = 2(ui2 + uR2)1/2 is used in the Matrix of equivalence.

The degree of equivalence between two laboratories i and j is given by a pair of terms: Dij = Di - Dj and Uij, its expanded uncertainty (k = 2), both expressed in MBq, where the approximation Uij = 2(ui2 + uj2 - 2ful2)1/2 is used (l is the NMIJ and f the correlation coefficient).
These statements make it possible to extend the BIPM.RI(II)-K1.Cr-51 Matrix of equivalence to the participant in the APMP.RI(II)-K2.Cr-51 comparison.
Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cr-51 and APMP.RI(II)-K2.Cr-51

MEASURAND : SIR equivalent activity of 51Cr

  • Key comparison BIPM.RI(II)-K1.Cr-51

xi : result of measurement carried out in the SIR for the sample submitted by laboratory i
ui
: combined standard uncertainty of xi

 Lab i xi
/ MBq
ui
/ MBq
Date of measurement
 IRMM 483.9 1.1 1981-06-17
 CMI-IIR 488.7 1.4 1982-06-30
 MKEH 487.3 1.3 1989-10-09
 PTB 487.6 0.9 1998-04-24
 NIST 489.4 1.4 1999-05-03
 NMIJ 487.1 1.6 2004-03-15
 NPL 487.0 3.4 2004-05-28
 LNE-LNHB 489.2 1.1 2006-07-17
  • Key comparison APMP.RI(II)-K2.Cr-51

xi : result of measurement carried out at laboratory i, converted to the equivalent activity through the NMJ SIR measurement (see Final Report dated October 2005)
ui
: combined standard uncertainty of xi

 Lab i xi
/ MBq
ui
/ MBq
Date of measurement
 VNIIM 488.3 1.2 2004