Comparison
Comparison close
Results

CCPR-K4, EURAMET.PR-K4, SIM.PR-K4, COOMET.PR-K4.1, APMP.PR-K4, 
EURAMET.PR-K4.2, EURAMET.PR-K4.3, GULFMET.PR-K4.2021.1 and GULFMET.PR-K4.2021.2

MEASURAND: Luminous flux

 

Degrees of equivalence relative to the CCPR-K4 key comparison value, Di  and its expanded uncertainty Ui at a 95 % expressed in relative terms.

Luminous flux
Di / 10-2
Participating institutes
Comparison
Comparison close
Results

CCPR-K4, EURAMET.PR-K4, SIM.PR-K4, COOMET.PR-K4.1, APMP.PR-K4, 
EURAMET.PR-K4.2, EURAMET.PR-K4.3, GULFMET.PR-K4.2021.1 and GULFMET.PR-K4.2021.2

MEASURAND: Luminous flux

 

Degrees of equivalence relative to the CCPR-K4 key comparison value, Di  and its expanded uncertainty Ui at a 95 % expressed in relative terms.

LABi Di Uneg,i Upos,i
10-2
10-2
10-2
LNE-INM/Cnam 0.69 0.58
NMISA -0.01 1.06
NMIA 0.13 0.58
NMIJ 0.18 0.68
INRIM -0.06 0.96
IO-CSIC -0.43 1.70
NIM -0.22 0.52
NIST -0.21 0.62
NPL 0.37 0.40
NRC 0.99 2.00
METAS -0.57 1.38
MKEH 0.43 1.30
PTB -0.42 0.56
SMU -0.88 2.20
VNIIOFI -0.51 0.66
INTI -0.43 1.00
BIPM 0.32 1.02
MIKES 0.03 0.76
CMI 0.35 1.00
DMDM 0.43 1.20
INM(RO) -0.71 1.30
VSL 1.20 1.10
GUM -0.73 1.00
BIM -1.00 2.60
BEV -0.80 1.40
SP 0.02 1.40
CENAM -0.41 1.09
INMETRO 0.48 2.08
INTI -0.45 1.01
NSC "Institute of Metrology" -0.94 1.05
CMS 0.29 1.63
KRISS -1.00 1.32
NIMT 0.41 1.22
NML-SIRIM 0.40 1.63
NPLI 0.51 1.37
NMC-A*STAR -0.49 1.11
RMC-LIPI 0.07 1.30
UME -0.25 1.60
SASO-NMCC 1.36 2.20
NIS 1.43 1.60
Participating institutes
CCPR-K4
EURAMET.PR-K4
SIM.PR-K4
COOMET.PR-K4.1
APMP.PR-K4
EURAMET.PR-K4.2
EURAMET.PR-K4.3
GULFMET.PR-K4.2021.1
GULFMET.PR-K4.2021.2
CCPR-K4
Metrology area, Sub-field Photometry and Radiometry, Photometry
Description Luminous flux
Time of measurements 1997 - 1998
Status Approved for equivalence
Final Reports of the comparisons
Measurand Luminous flux
Transfer device Lamps
Comparison type Key Comparison
Consultative Committee CCPR (Consultative Committee for Photometry and Radiometry)
Conducted by CCPR (Consultative Committee for Photometry and Radiometry)
Comments

Results published on 5 July 2001

Luminous flux of lamps

EURAMET.PR-K4, SIM.PR-K4, COOMET.PR-K4.1, APMP.PR-K4, EURAMET.PR-K4.2 and EURAMET.PR-K4.3 results are linked to those of CCPR-K4

Pilot institute PTB
Physikalisch-Technische Bundesanstalt
Germany
Contact person G. Sauter

+49 (0) 531 592 4120
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

BNM-INM

Bureau National de Métrologie - Institut National de Métrologie, France, EURAMET

CSIR-NML

Council for Scientific and Industrial Research, National Metrology Laboratory (became NMISA in 2007), South Africa, AFRIMETS

CSIRO-NML (now NMIA)

Commonwealth Scientific and Industrial Research Organization - National Measurement Laboratory; became National Measurement Institute, Australia, NMIA in 2004, Australia, APMP

IEN

Istituto Elettrotecnico Nazionale Galileo Ferraris; from 1st January 2006: Istituto Nazionale di Ricerca Metrologica, INRIM, Italy, EURAMET

IFA-CSIC

Instituto de Fisica Aplicada - Consejo Superior de Investigaciones Cientificas, now IO-CSIC, Instituto de Optica Daza de Valdés, Spain, EURAMET

INTI

Instituto Nacional de Tecnologia Industrial, Argentina, SIM

METAS

Federal Institute of Metrology, Switzerland, EURAMET

NIM

National Institute of Metrology, China, APMP

NIST

National Institute of Standards and Technology, United States, SIM

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NPL

National Physical Laboratory, United Kingdom, EURAMET

NRC

National Research Council, Canada, SIM

OMH

Országos Mérésügyi Hivatal; Subsequently MKEH, now BKFH, Hungary, EURAMET

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

VNIIOFI

All-Russian Scientific Institute for Optical and Physical Measurements, Rosstandart, Russian Federation, COOMET

Comparison
Comparison close
Results

CCPR-K4, EURAMET.PR-K4, SIM.PR-K4, COOMET.PR-K4.1, APMP.PR-K4, 
EURAMET.PR-K4.2, EURAMET.PR-K4.3, GULFMET.PR-K4.2021.1 and GULFMET.PR-K4.2021.2

 

MEASURAND: Luminous flux

 

The key comparison reference value, xR, is calculated as the weighted average of the invidual results xi, weighted by the inverse square of the individual standard uncertainties, ui, with the application of a minimum cutoff of 0.30% to the measurements of LNE-INM/Cnam, NMIA, NIM, NPL and PTB.
INTI and the BIPM are excluded from the calculation of xR.
The standard uncertainty of xR is uR = 0.1%. It is negligible compared to the ui values.

 

The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: Di = (xi - xR) / xR and Ui, its expanded uncertainty (k = 2), both expressed in relative units. Ui = 2ui.

 

Linking EURAMET.PR-K4 to CCPR-K4

PTB, INRIM and LNE-INM/Cnam acted as linking laboratories for EURAMET.PR-K4. The luminous flux maintained at the linking laboratories was used to evaluate the EURAMET reference value as a weighted luminous flux average.

By definition, the CCPR-K4 key comparison reference value and the EURAMET.PR-K4 reference value have identical values (see Section 4.2 of the EURAMET.PR-K4 Final Report, especially Table 4 and Figure 4), but the expanded uncertainty associated with the EURAMET reference value is about triply the expanded uncertainty associated with the CCPR-K4 key comparison reference value.

 

Linking SIM.PR-K4 to CCPR-K4

NRC and NIST provide the link between key comparisons CCPR-K4 and SIM.PR-K4 and, the linkage process is described in Section 9.2 of the SIM.PR-K4 Final Report.

 

Linking COOMET.PR-K4.1 to CCPR-K4

VNIIOFI provides the link between key comparisons CCPR-K4 and COOMET.PR-K4.1, and the linkage process is described in Section 5 of the COOMET.PR-K4.1 Final Report.

 

Linking APMP.PR-K4 to CCPR-K4

NIM, NMISA and NMIJ-AIST provide the link between key comparisons CCPR-K4 and APMP.PR-K4, and the linkage process is described in Section 8 of the APMP.PR-K4 Final Report.

 

Linking EURAMET.PR-K4.2 to CCPR-K4

LNE/INM-Cnam provides the link between key comparisons CCPR-K4 and EURAMET.PR-K4.2, and the linkage process is described in Section 4 of the EURAMET.PR-K4.2 Final Report.

 

Linking EURAMET.PR-K4.3 to CCPR-K4

IO-CSIC provides the link between key comparisons CCPR-K4 and EURAMET.PR-K4.3, and the linkage process is described in Section 5 of the EURAMET.PR-K4.3 Final Report.

 

Linking GULFMET.PR-K4.2021.1 to CCPR-K4

UME provides the link between key comparisons CCPR-K4 and GULFMET.PR-K4.2021.1, through its participation in EURAMET.PR-K4.3. The linking process is described in Section 5.2 of the GULFMET.PR-K4.2021.1 Final Report.

 

Linking GULFMET.PR-K4.2021.2 to CCPR-K4

UME provides the link between key comparisons CCPR-K4 and GULFMET.PR-K4.2021.1, through its participation in EURAMET.PR-K4.3. The linking process is described in Section 5.2 of the GULFMET.PR-K4.2021.2 Final Report.

Comparison
Comparison close
Results

CCPR-K4, EURAMET.PR-K4, SIM.PR-K4, COOMET.PR-K4.1, APMP.PR-K4, 
EURAMET.PR-K4.2, EURAMET.PR-K4.3, GULFMET.PR-K4.2021.1 and GULFMET.PR-K4.2021.2

 

MEASURAND: Luminous flux

 

CCPR-K4

The individual measurements, xi, of the participating laboratories take the form of ratios which depend on the reference used at the pilot laboratory, the PTB. This is cancelled out by normalization using the key comparison reference value, xR. This procedure leads to normalized individual measurements: xi / xR.

 

EURAMET.PR-K4

The participants' measurements are given in Annex B of the EURAMET.PR-K4 Final Report.

 

SIM.PR-K4

The ratio of the measurement results of the laboratories participating in SIM.PR-K4 to the measurement results obtained at the pilot laboratory, the CENAM, are given in table 15 of the SIM.PR-K4 Final Report.

 

COOMET.PR-K4.1

The measurement results of the laboratories participating in COOMET.PR-K4.1 are given in table 7 of the COOMET.PR-K4.1 Final Report.

 

APMP.PR-K4

The measurement results of the participants in APMP.PR-K4 are given in Section 6  of the APMP.PR-K4 Final Report.

 

EURAMET.PR-K4.2

The measurement results of the participants in EURAMET.PR-K4.2 are given in Section 4  of the EURAMET.PR-K4.2 Final Report.

 

EURAMET.PR-K4.3

The measurement results of the participants in EURAMET.PR-K4.3 are given in Section 3 and 4  of the EURAMET.PR-K4.3 Final Report.

 

GULFMET.PR-K4.2021.1

The measurement results of the participants in GULFMET.PR-K4.2021.1 are given in Section 5.1  of the GULFMET.PR-K4.2021.1 Final Report.

 

GULFMET.PR-K4.2021.2

The measurement results of the participants in GULFMET.PR-K4.2021.1 are given in Section 5.1  of the GULFMET.PR-K4.2021.2 Final Report.