MEASURAND : Activity concentration in 3H
NOMINAL VALUE : 37 kBq g-1
Degrees of equivalence Di and expanded uncertainty Ui (k = 2) expressed in kBq/g
Note : the right-hand axis shows approximate relative values only
MEASURAND : Activity concentration in 3H
NOMINAL VALUE : 37 kBq g-1
Degrees of equivalence Di and expanded uncertainty Ui (k = 2) expressed in kBq/g
Results are presented under A4 printable format in Summary Results (PDF file).
Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
Description | Measurement of activity concentration of the same solution of tritiated water |
Time of measurements | 2008 - 2009 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
References | |
Measurand | Activity concentration of H-3 in kBq/ml |
Transfer device | Radioactive solution |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | CCRI (Consultative Committee for Ionizing Radiation) |
Comments | Results published on 26 June 2018 Activity concentration of the same H-3 solution
|
Pilot institute |
BIPM
Bureau International des Poids et Mesures BIPM - International Organization |
Contact person | Carine Michotte +33 (0) 1 45 07 70 85 |
Pilot laboratory | |
---|---|
BIPM |
Bureau International des Poids et Mesures, BIPM - International Organization, N/A |
ANSTO |
Australian Nuclear Science and Technology Organisation, Australia, APMP |
BARC |
Bhabha Atomic Research Centre, India, APMP |
CNEA |
National Commission of Atomic Energy, Argentina, SIM |
IFIN-HH |
"Horia Hulubei" National Institute of Research and Development for Physics and Nuclear Engineering, Romania, EURAMET |
IRA |
Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
LNE-LNHB |
Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET |
NIM |
National Institute of Metrology, China, APMP |
NIST |
National Institute of Standards and Technology, United States, SIM |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
NMISA |
National Metrology Institute of South Africa, South Africa, AFRIMETS |
NPL |
National Physical Laboratory, United Kingdom, EURAMET |
POLATOM |
Radioisotope Centre Polatom, Poland, EURAMET |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
MEASURAND : Activity concentration in 3H
NOMINAL VALUE : 37 kBq g-1
The key comparison reference value is obtained by applying the power moderated weighted mean of the results giving |
The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: |
When required, the degree of equivalence between two laboratories can be evaluated by a pair of terms: Dij = Di - Dj = (xi - xj) and Uij, its expanded uncertainty (k = 2), both expressed in kBq g-1. The approximation Uij ~ 2(ui2 + uj2)1/2 may be used for the computation of the pair-wise degrees of equivalence. |
MEASURAND : Activity concentration in 3H
NOMINAL VALUE : 37 kBq g-1
xi : result of measurement carried out by laboratory i
ui : combined standard uncertainty of xi