CCRI(III)-K9.AmBe, CCRI(III)-K9.AmBe.2 and CCRI(III)-K9.AmBe.1

 

MEASURAND   Neutron source emission rate

SOURCE                   241Am-Be(α,n)

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in s-1.

Emission rate
Di / s-1
Participating institutes

CCRI(III)-K9.AmBe, CCRI(III)-K9.AmBe.2 and CCRI(III)-K9.AmBe.1

 

MEASURAND   Neutron source emission rate

SOURCE                   241Am-Be(α,n)

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in s-1.

LABi Di Uneg,i Upos,i
s-1
s-1
s-1
CIAE -6.47E+04 3.34E+04
CMI -1.27E+04 3.68E+04
KRISS -2.71E+03 2.65E+04
LNE-LNHB -1.46E+05 6.52E+04
LNMRI -4.71E+03 2.50E+04
NIST -3.57E+04 5.25E+04
NPL -1.27E+04 2.72E+04
VNIIM 2.53E+04 2.04E+04
ENEA-INMRI 3.41E+05 8.00E+04
LNE-LNHB -7.19E+04 7.06E+04
NIM -6.12E+03 5.87E+04
Participating institutes
CCRI(III)-K9.AmBe
CCRI(III)-K9.AmBe.2
CCRI(III)-K9.AmBe.1
Comparison
Comparison close
CCRI(III)-K9.AmBe
Metrology area, Sub-field Ionizing Radiation, Section III (Neutron Measurements)
Description Comparison of emission rate
Time of measurements 1999 - 2005
Status Approved for equivalence
Final Reports of the comparisons
Measurand Emission rate
Transfer device AmBe source
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by CCRI (Consultative Committee for Ionizing Radiation)
Comments

Results published on 06 October 2011

Comparison of emission rate

Pilot institute NPL
National Physical Laboratory
United Kingdom
Contact person Neil Roberts

+44 (0) 20 8943 6869
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NPL

National Physical Laboratory, United Kingdom, EURAMET

CMI

Czech Metrology Institute, Czechia, EURAMET

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

NIST

National Institute of Standards and Technology, United States, SIM

VNIIM

D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET

CCRI(III)-K9.AmBe, CCRI(III)-K9.AmBe.2 and CCRI(III)-K9.AmBe.1

 

MEASURAND   Neutron source emission rate

SOURCE                   241Am-Be(α,n)

 

CCRI(III)-K9.AmBe

The key comparison reference value xR and its standard uncertainty uR have been calculated from the participants' results, excluding CIAE and LNE-LNHB data (see details in Section 5.1.1 and Appendix A of the Final Report).

 

xR = 2.445 x 106 s-1

uR = 0.006 x 106 s-1

 

The degree of equivalence of laboratory i with respect to xR is given by a pair of terms: Di and Ui its expanded uncertainty (k = 2), both expressed in s-1.

 

Linking CCRI(III)-K9.AmBe.2 to CCRI(III)-K9.AmBe

The results obtained in CCRI(III)-K9.AmBe.2 were linked to CCRI(III)-K9.AmBe via NPL (United Kingdom) who participated in both comparisons.

 

Linking CCRI(III)-K9.AmBe.2 to CCRI(III)-K9.AmBe

The results obtained in CCRI(III)-K9.AmBe.1 were linked to CCRI(III)-K9.AmBe via NPL (United Kingdom) who participated in both comparisons.

CCRI(III)-K9.AmBe, CCRI(III)-K9.AmBe.2 and CCRI(III)-K9.AmBe.1

 

MEASURAND   Neutron source emission rate

SOURCE                   241Am-Be(α,n)

 

xi   emission rate submitted by each institute i at the reference date of 1 January 2000

ui   combined standard uncertainty of xi

 

CCRI(III)-K9.AmBe

(a)   The NPL value is an average of the measurements made in 1999 and 2004.

(b)   The VNIIM value is an average of results obtained for the manganese bath and associated particle measurements.

 

CCRI(III)-K9.AmBe.2

 

CCRI(III)-K9.AmBe.1