Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

I-131
Di / MBq
Participating institutes

Results in absolute values

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6 and EURAMET.RI(II)-K2

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

I-131
Di / kBq/MBq
Participating institutes

Results in relative values

Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
CIEMAT -0.22 0.37
PTB -0.01 0.15
CMI 0.00 0.23
IFIN-HH 0.04 0.27
NMIJ 0.06 0.19
NIST -0.03 0.27
LNE-LNHB -0.19 0.18
BEV 0.34 0.35
POLATOM -0.10 0.31
NMISA 0.18 0.31
ANSTO 0.08 0.45
BARC -0.30 0.39
INER 0.82 0.35
KRISS 0.93 0.21
OAP 0.5 1.3
PTKMR -0.5 1.8
LNMRI/IRD -0.2 1.1
ENEA-INMRI -0.08 0.35
Participating institutes
BIPM.RI(II)-K1.I-131
APMP.RI(II)-K2.I-131
CCRI(II)-S6.I-131
EURAMET.RI(II)-K2.I-131

Results in absolute values

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
CIEMAT -5.5 9.3
PTB -0.3 3.7
CMI 0.0 5.8
IFIN-HH 1.0 6.8
NMIJ 1.5 4.8
NIST -0.8 6.8
LNE-LNHB -4.6 4.4
BEV 8.4 8.6
POLATOM -2.5 7.8
NMISA 4.4 7.8
ANSTO 2 11
BARC -7.5 10
INER 20 9
KRISS 23 5
OAP 14 32
PTKMR -12 45
LNMRI/IRD -5 27
ENEA-INMRI -2 8.6
Participating institutes
BIPM.RI(II)-K1.I-131
APMP.RI(II)-K2.I-131
CCRI(II)-S6.I-131
EURAMET.RI(II)-K2.I-131

Results in relative values

CCRI(II)-S6.I-131
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Measurement of radioactivity content of same solution
Time of measurements 2006
Status Approved
Final Reports of the comparisons
Measurand Activity per unit mass (Radionuclide: I-131)
Transfer device Aliquots of stock solution distributed to participants through third-party supplier
Comparison type Supplementary Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by CCRI (Consultative Committee for Ionizing Radiation)
Comments
 
Pilot institute IAEA
International Atomic Energy Agency
IAEA - International Organization
Contact person A. Meghzifene

+43 1 2600 21653
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
IAEA

International Atomic Energy Agency, IAEA - International Organization, SIM

BARC

Bhabha Atomic Research Centre, India, APMP

CMI

Czech Metrology Institute, Czechia, EURAMET

IFIN-HH

"Horia Hulubei" National Institute of Research and Development for Physics and Nuclear Engineering, Romania, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

 

BIPM.RI(II)-K1.I-131

MEASURAND : Equivalent activity of I-131

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is  xR = 40 371 kBq with a standard uncertainty uR = 34 kBq

(see Section 4.1 of the BIPM.RI(II)-K1.I-131Final Report).  The value xi is taken as the equivalent activity for a laboratory i.

 

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (x- xR) and Uits expanded uncertainty (k = 2), both expressed in MBq, and

Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

 

Linking APMP.RI(II)-K2.I-131 to BIPM.RI(II)-K1.I-131

The value xi is the equivalent activity for laboratory i participant in APMP.RI(II)-K2.I-131 having been normalized using the value of the NMIJ as the linking laboratory.

The degree of equivalence of laboratory i participant in APMP.RI(II)-K2.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value.

 

Linking CCRI(II)-S6.I-131 to BIPM.RI(II)-K1.I-131

The value xi is the equivalent activity for laboratory i participant in CCRI(II)-S6.I-131 having been normalized using the value of the CMI as the linking laboratory.

The degree of equivalence of laboratory i participant in CCRI(II)-S6.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value..

 

Linking EURAMET.RI(II)-K2.I-131 to BIPM.RI(II)-K1.I-131

The value xi is the equivalent activity for laboratory i participant in EURAMET.RI(II)-K2.I-131 having been normalized to the value of the POLATOM as the linking laboratory.

The degree of equivalence of laboratory i participant in EURAMET.RI(II)-K2.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value..

These statements make it possible to extend the BIPM.RI(II)-K1.I-131 matrices of equivalence to all participants in the APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131, and  EURAMET.RI(II)-K2.I-131 comparisons

 

Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

xi   measured mass fraction reported by laboratory iLabi

ui   combined standard uncertainty of xi

 

 

BIPM.RI(II)-K1.I-131

 

  Labi xi ui Date of measurement
    kBq kBq  
  CIEMAT 40 150 190 2002-06-26
  PTB 40 357 77 2004-04-30
  CMI 40 370 120 2006-08-22
  IFIN-HH 40 410 140 2008-01-03
  NMIJ 40 430 100 2009-03-09
  NIST 40 340 140 2012-03-28
  LNE-LNHB 40 184 92 2012-11-16
  BEV 40 710 170 2015-06-02
  POLATOM 40 270 160 2015-10-22
  NMISA 40 550 160 2017-09-13

 

APMP.RI(II)-K2.I-131

Labi xi ui Date of measurement
kBq kBq
ANSTO 40 450 220 March 2009
BARC 40 070 190 March 2009
INER 41 190 170 March 2009
KRISS 41 300 100 March 2009
OAP 40 920 640 March 2009
PTKMR 39 900 900 March 2009

 

CCRI(II)-S6.I-131

Labi xi ui Date of measurement
kBq kBq
LNMRI/IRD 40 160 540 July 2006
     

 

EURAMET.RI(II)-K2.I-131

Labi xi ui Date of measurement
kBq kBq
ENEA-INMRI 40 290 170 December 2015