Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Equivalent activity of I-131
Di / MBq
Participating institutes

Results in absolute values.

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

Degrees of equivalence represented by Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Equivalent activity of I-131
Di / kBq/MBq
Participating institutes

Results in relative values

Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
CMI 0.00 0.23
IFIN-HH 0.04 0.27
NMIJ 0.06 0.19
NIST -0.03 0.27
LNE-LNHB -0.19 0.18
BEV 0.34 0.35
POLATOM -0.10 0.31
NMISA 0.18 0.31
ANSTO 0.08 0.45
BARC -0.30 0.39
INER 0.82 0.35
KRISS 0.93 0.21
OAP 0.50 1.3
PTKMR -0.50 1.8
LNMRI/IRD -0.2 1.1
ENEA-INMRI -0.08 0.35
Participating institutes
BIPM.RI(II)-K1.I-131
APMP.RI(II)-K2.I-131
CCRI(II)-S6.I-131
EURAMET.RI(II)-K2.I-131

Results in absolute values

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

Degrees of equivalence represented by Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
CMI 0.0 5.8
IFIN-HH 1.0 6.8
NMIJ 1.5 4.8
NIST -0.8 6.8
LNE-LNHB -4.6 4.4
BEV 8.4 8.6
POLATOM -2.5 7.8
NMISA 4.4 7.8
ANSTO 2.0 11
BARC -7.5 10
INER 20 9.0
KRISS 23 5.0
OAP 14 32
PTKMR -12 45
LNMRI/IRD -5.0 27
ENEA-INMRI -2.0 8.6
Participating institutes
BIPM.RI(II)-K1.I-131
APMP.RI(II)-K2
CCRI(II)-S6
EURAMET.RI(II)-K2

Results in relative values

APMP.RI(II)-K2.I-131
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide I-131
Time of measurements 2009
Status Approved for equivalence
Final Reports of the comparisons
Measurand Activity concentration in kBq/g
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by APMP (Asia Pacific Metrology Programme)
Comments Results published on 06 February 2014
APMP.RI(II)-K2.I-131 results are linked to those of BIPM.RI(II)-K1.I-131
SIR equivalent activity of the radionuclide 131I:
Pilot institute NMIJ AIST
National Metrology Institute of Japan
Japan
Contact person A. Yunoki

+81 29 861 3470
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

ANSTO

Australian Nuclear Science and Technology Organisation, Australia, APMP

BARC

Bhabha Atomic Research Centre, India, APMP

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

NRSL/INER

Institute of Nuclear Energy Research, Chinese Taipei, APMP

OAP

Office of Atoms for Peace, Thailand, APMP

Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

 

BIPM.RI(II)-K1.I-131

MEASURAND : Equivalent activity of I-131

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is  xR = 40 371 kBq with a standard uncertainty uR = 34 kBq

(see Section 4.1 of the BIPM.RI(II)-K1.I-131Final Report).  The value xi is taken as the equivalent activity for a laboratory i.

 

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (x- xR) and Uits expanded uncertainty (k = 2), both expressed in MBq, and

Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xi - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

 

Linking APMP.RI(II)-K2.I-131 to BIPM.RI(II)-K1.I-131

The value xi is the equivalent activity for laboratory i participant in APMP.RI(II)-K2.I-131 having been normalized using the value of the NMIJ as the linking laboratory.

The degree of equivalence of laboratory i participant in APMP.RI(II)-K2.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value.

 

Linking CCRI(II)-S6.I-131 to BIPM.RI(II)-K1.I-131

The value xi is the equivalent activity for laboratory i participant in CCRI(II)-S6.I-131 having been normalized using the value of the CMI as the linking laboratory.

The degree of equivalence of laboratory i participant in CCRI(II)-S6.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value..

 

Linking EURAMET.RI(II)-K2.I-131 to BIPM.RI(II)-K1.I-131

The value xi is the equivalent activity for laboratory i participant in EURAMET.RI(II)-K2.I-131 having been normalized to the value of the POLATOM as the linking laboratory.

The degree of equivalence of laboratory i participant in EURAMET.RI(II)-K2.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value..

These statements make it possible to extend the BIPM.RI(II)-K1.I-131 matrices of equivalence to all participants in the APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131, and  EURAMET.RI(II)-K2.I-131 comparisons

Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131

MEASURAND : Equivalent activity of I-131

Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value

 

xi   result of measurement carried out in the SIR for the sample submitted by laboratory iLabi

ui   combined standard uncertainty of xi

 

BIPM.RI(II)-K1.I-131

  Labi xi ui Date of measurement
    kBq kBq  
  CMI 40 370 120 2006-08-22
  IFIN-HH 40 410 140 2008-01-03
  NMIJ 40 430 100 2009-03-09
  NIST 40 340 140 2012-03-28
  LNE-LNHB 40 184 92 2012-11-16
  BEV 40 710 170 2015-06-02
  POLATOM 40 270 160 2015-10-22
  NMISA 40 550 160 2017-09-13

 

 

APMP.RI(II)-K2.I-131

  Labi xi ui Date of measurement
  kBq kBq
  ANSTO 40 450 220 March 2009
  BARC 40 070 190 March 2009
  INER 41 190 170 March 2009
  KRISS 41 300 100 March 2009
  OAP 40 920 640 March 2009
  PTKMR 39 900 900 March 2009

 

 

CCRI(II)-S6.I-131

  Labi xi ui Date of measurement
  kBq kBq
  LNMRI/IRD 40160 540 July 2006

 

 

 

EURAMET.RI(II)-K2.I-131

  Labi xi ui Date of measurement
  kBq kBq
  ENEA-INMRI 40290 170 December 2015