Comparison
Comparison close

BIPM.RI(I)-K5, APMP.RI(I)-K5, EURAMET.RI(I)-K5.1 and EURAMET.RI(I)-K5.2

 

MEASURAND       Air kerma

Radiation quality    Cs-137 beam

 

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2) expressed in mGy/Gy

Air Kerma
Di  (mGy/Gy)
Participating institutes
Comparison
Comparison close

BIPM.RI(I)-K5, APMP.RI(I)-K5, EURAMET.RI(I)-K5.1 and EURAMET.RI(I)-K5.2

 

MEASURAND       Air kerma

Radiation quality    Cs-137 beam

 

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2) expressed in mGy/Gy

LABi Di Uneg,i Upos,i
mGy/Gy
mGy/Gy
mGy/Gy
NIST -1.0 7.0
NMIJ -2.3 5.2
PTB 2.4 5.4
NIM -3.3 4.2
ININ 4.8 4.0
VSL -4.8 7.6
SMU 5.1 5.4
BEV 4.3 5.4
CIEMAT -0.8 4.4
GUM 0.5 5.6
LNE-LNHB -3.0 4.4
KRISS -0.8 4.0
BFKH 3.9 6.2
NRSL/INER -4.6 8.9
VINS -3.3 20
STUK 12.2 18
DSA 1.1 14
NPL 5.8 18
SSM 2.1 15
Participating institutes
BIPM.RI(I)-K5
APMP.RI(I)-K5
EURAMET.RI(I)-K5.1
EURAMET.RI(I)-K5.2
APMP.RI(I)-K5.2026
Metrology area, Sub-field Ionizing Radiation, Section I (x and gamma rays, electrons)
Description Measurement of air kerma for Caesium 137
Time of measurements 2026 -
Status Protocol complete
References

APMP.RI(I)-K5 Technical Protocol

Measurand Air kerma rate
Parameters Radiation type: Cs-137 gamma rays
Transfer device Ionization chambers
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by APMP (Asia Pacific Metrology Programme)
Pilot institute NRSL/INER
Institute of Nuclear Energy Research
Chinese Taipei
Contact person Yu-Tien Tsai

First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NRSL/INER

Institute of Nuclear Energy Research, Chinese Taipei, APMP

ARPANSA

Australian Radiation Protection and Nuclear Safety Agency, Australia, APMP

BARC

Bhabha Atomic Research Centre, India, APMP

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

NIM

National Institute of Metrology, China, APMP

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

Nuclear Malaysia

Malaysian Nuclear Agency, Malaysia, APMP

OAP

Office of Atoms for Peace, Thailand, APMP

UzNIM

Uzbek National Institute of Metrology, Uzbekistan, COOMET

Comparison
Comparison close

BIPM.RI(I)-K5, APMP.RI(I)-K5, EURAMET.RI(I)-K5.1 and EURAMET.RI(I)-K5.2

 

MEASURAND       Air kerma

Radiation quality    Cs-137 beam

 

The key comparison reference value is the BIPM evaluation of air kerma (see International framework of traceability for radiation dosimetry quantities) .

 

BIPM.RI(I)-K5

The degree of equivalence of each laboratory i with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy:
the relative difference Di and Ui, its expanded uncertainty (k = 2), with Ui = 2ui.

When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms both expressed in mGy/Gy:

Dij = Di - Dj  and Uij  , its expanded uncertainty (k = 2). The approximation for Uij is explained in section 5 of the Final Report.

 

Linking  EURAMET.RI(I)-K5.1 to BIPM.RI(I)-K5

The linking of EURAMET.RI(I)-K5.1 to BIPM.RI(I)-K5 is made via BEV (Austria) who participated in both comparisons.

 

Linking  APMP.RI(I)-K5 to BIPM.RI(I)-K5

The linking of APMP.RI(I)-K5 to BIPM.RI(I)-K5 is made via KRISS (Korea) and NMIJ AIST (Japan) who participated in both comparisons.

 

Linking  EURAMET.RI(I)-K5.1 to BIPM.RI(I)-K5

The linking of EURAMET.RI(I)-K5.2 to BIPM.RI(I)-K5 is made via CIEMAT (Spain) who participated in both comparisons.

Comparison
Comparison close

BIPM.RI(I)-K5, APMP.RI(I)-K5, EURAMET.RI(I)-K5.1 and EURAMET.RI(I)-K5.2

 

MEASURAND       Air kerma

Radiation quality    Cs-137 beam

 

xi     ratio of the measurement of laboratory i to that of the BIPM

ui     combined standard uncertainty of the ratio xi

 

BIPM.RI(I)-K5

  Labi xi ui year of measurement
  NIST 0.9990 0.0035 2011
  NMIJ 0.9977 0.0026 2012
  PTB 1.0024 0.0027 2014
  NIM 0.9967 0.0021 2014
  ININ 1.0048 0.0020 2015
  VSL 0.9952 0.0038 2016
  SMU 1.0051 0.0027 2017
  BEV 1.0043 0.0027 2023
  CIEMAT 0.9992 0.0022 2023
  GUM 1.0005 0.0028 2024
  LNE-LNHB 0.9970 0.0022 2024
  KRISS 0.9992 0.0020 2025
  BFKH 1.0039 0.0031 2025

 

APMP.RI(I)-K5

  Labi xi ui year of measurement
  KRISS 0.9931 0.0041 2014
  NMIJ 1.0033 0.0043 2014
  INER 0.9954 0.0045 2014
  CIEMAT 1.0042 0.0086 2014
  NIM 1.0013 0.0044 2014

 

EURAMET.RI(I)-K5.1

  Labi xi ui year of measurement
  VINS 0.9967 0.0101 2017

 

EURAMET.RI(I)-K5.2

  Labi xi ui year of measurement
  STUK 1.0122 0.0090 2025
  DSA 1.0011 0.0070 2025
  NPL 1.0058 0.0090 2025
  SSM 1.0021 0.0075 2025