Comparison
Comparison close

BIPM.RI(I)-K5, EURAMET.RI(I)-K5.1 and APMP.RI(I)-K5

 

MEASURAND       Air kerma

Radiation quality    Cs-137 beam

 

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2) expressed in mGy/Gy

Air Kerma
Di  (mGy/Gy)
Participating institutes
Comparison
Comparison close

BIPM.RI(I)-K5, EURAMET.RI(I)-K5.1 and APMP.RI(I)-K5

 

MEASURAND       Air kerma

Radiation quality    Cs-137 beam

 

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2) expressed in mGy/Gy

LABi Di Uneg,i Upos,i
mGy/Gy
mGy/Gy
mGy/Gy
GUM -0.5 5.8
IST/ITN 1.3 4.2
LNE-LNHB -1.6 5.2
MKEH 5.3 5.0
VNIIM 1.3 5.4
KRISS -1.4 4.4
NIST -1.0 7.0
NMIJ -2.3 5.2
PTB 2.4 5.4
NIM -3.3 4.2
ININ 4.8 4.0
VSL -4.8 7.6
SMU 5.1 5.4
BEV 4.3 5.4
CIEMAT -0.8 4.4
INER -4.6 8.9
VINS -3.3 20.2
Participating institutes
BIPM.RI(I)-K5
APMP.RI(I)-K5
EURAMET.RI(I)-K5.1
Comparison
Comparison close
GULFMET.RI(I)-K5
Metrology area, Sub-field Ionizing Radiation, Section I (x and gamma rays, electrons)
Description Measurement of air kerma for Caesium 137
Time of measurements 2022
Status Measurements in progress
References

Protocol GULFMET.RI(I)-K5: in 137Cs  Ver 1  DOC

Protocol GULFMET.RI(I)-K5: in 137Cs  Ver 1  PDF

Protocol GULFMET.RI(I)-K5: in 137Cs  Ver 2  DOC

Protocol GULFMET.RI(I)-K5: in 137Cs  Ver 2  PDF

Protocol GULFMET.RI(I)-K5: in 137Cs  Ver 3 DOC

Protocol GULFMET.RI(I)-K5: in 137Cs  Ver 3 PDF

Measurand Air Kerma Calibration Coefficient
Transfer device 1 litre Spherical chamber PTW 32002 Serial Number 00765
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by GULFMET (Gulf Association for Metrology)
Comments NK corrected to T = 20ºC & P = 101.325 kPa,; Relative humidity between 30% and 70%; SCD = 300 cm; Minimum beam diameter at 20 cm; Beam homogeneity >= 5 %; Air kerma rates not less than 0.5 mGy/h.
Pilot institute SASO-NMCC
National Measurement and Calibration Center - Saudi Standards, Metrology and Quality Organization
Saudi Arabia
Contact person Mehenna Arib

+966509320509
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
SASO-NMCC

National Measurement and Calibration Center - Saudi Standards, Metrology and Quality Organization, Saudi Arabia, GULFMET

FANR

Federal Authority for Nuclear Regulation SSDL, United Arab Emirates, GULFMET

IMBIH

Institute of Metrology of Bosnia and Herzegovina, Bosnia and Herzegovina, EURAMET

PAI

Public Authority for industry, Kuwait, GULFMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

TENMAK-NUKEN

Nuclear Energy Research Institute, Türkiye, EURAMET

UzNIM

Uzbek National Institute of Metrology, Uzbekistan, COOMET

Comparison
Comparison close

BIPM.RI(I)-K5, EURAMET.RI(I)-K5.1 and APMP.RI(I)-K5

 

MEASURAND       Air kerma

Radiation quality    Cs-137 beam

 

The key comparison reference value is the BIPM evaluation of absrobed dose to water.

 

BIPM.RI(I)-K5

The degree of equivalence of each laboratory i with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy:
the relative difference Di and Ui, its expanded uncertainty (k = 2), with Ui = 2ui.

When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms both expressed in mGy/Gy:

Dij = Di - Dj  and Uij  , its expanded uncertainty (k = 2). The approximation for Uij is explained in section 5 of the Final Report.

 

Linking  EURAMET.RI(I)-K5.1 to BIPM.RI(I)-K5

The linking of EURAMET.RI(I)-K5.1 to BIPM.RI(I)-K5 is made via BEV (Austria) who participated in both comparisons.

 

Linking  APMP.RI(I)-K5 to BIPM.RI(I)-K5

The linking of APMP.RI(I)-K5 to BIPM.RI(I)-K5 is made via KRISS (Korea) and NMIJ AIST (Japan) who participated in both comparisons.

Comparison
Comparison close

BIPM.RI(I)-K5, EURAMET.RI(I)-K5.1 and APMP.RI(I)-K5

 

MEASURAND       Air kerma

Radiation quality    Cs-137 beam

 

xi     ratio of the measurement of laboratory i to that of the BIPM

ui     combined standard uncertainty of the ratio xi

 

BIPM.RI(I)-K5

  Labi xi ui year of measurement
  GUM 0.9995 0.0029 2006
  IST/ITN 1.0013 0.0021 2008
  LNE-LNHB 0.9984 0.0026 2008
  MKEH 1.0053 0.0025 2009
  VNIIM 1.0013 0.0027 2009
  KRISS 0.9986 0.0022 2010
  NIST 0.9990 0.0035 2011
  NMIJ 0.9977 0.0026 2012
  PTB 1.0024 0.0027 2014
  NIM 0.9967 0.0021 2014
  ININ 1.0048 0.0020 2015
  VSL 0.9952 0.0038 2016
  SMU 1.0051 0.0027 2017
  BEV 1.0043 0.0027 2023
  CIEMAT 0.9992 0.0022 2023

 

EURAMET.RI(I)-K5.1

  Labi xi ui year of measurement
  VINS 0.9967 0.0101 2017

 

APMP.RI(I)-K5

  Labi xi ui year of measurement
  KRISS 0.9931 0.0041 2014
  NMIJ 1.0033 0.0043 2014
  INER 0.9954 0.0045 2014
  CIEMAT 1.0042 0.0086 2014
  NIM 1.0013 0.0044 2014