| Metrology area, Sub-field | Mass and related quantities, Fluid Flow |
| Description | Comparison of standards for micro liquid flow rates |
| Time of measurements | 2025 - 2026 |
| Status | Protocol complete |
| References | |
| Measurand | Mass flow rate TS1 measurement flows range from 1 g/h to 200 g/h (1; 5; 10; 50; 100; 150; 200) g/h TS2 measurement flows range from 500 g/h to 30 000 g/h (200; 500; 1 000; 3 000; 6 000; 10 000; 15 000; 20 000; 30 000) g/h |
| Parameters | Working fluid temperature: 20 °C ± 5 °C Ambient temperature: 20 °C ± 5 °C Ambient humidity: from 30 % to 80 % Atmosphere pressure: from 86 kPa to 106 kPa Absence of free air in the measuring line of the standard |
| Transfer device | TS1: a Bronkhorst Low Flow Coriolis Mass Flow Meter, mini CORI-FLOW ML120 TS2: a Bronkhorst Low Flow Coriolis Mass Flow Meter, mini CORI-FLOW M14 |
| Comparison type | Supplementary Comparison |
| Consultative Committee | CCM (Consultative Committee for Mass and Related Quantities) |
| Conducted by | EURAMET (European Association of National Metrology Institutes) |
| Comments | A comparison is organized in order to determine the degree of equivalence of standards for micro liquid flow rates under static load changes in the range of 1 g/h to 30 000 g/h at ambient pressure and temperature. Two Coriolis Mass Flow Meters will be used as Transfer Standards (TS). If the participants are not able to calibrate in the full measurement range specified, they may calibrate the meter over a part of the flow range. Each laboratory will be responsible for transferring the meter to the next laboratory. |
| Pilot institute |
CMI
Czech Metrology Institute Czechia |
| Contact person | Miroslava Benkova +420 734 877 960 |
| BAM |
Bundesanstalt für Materialforschung und -prüfung, Germany, EURAMET |
| CHMU/CHMI |
Czech Hydrometeorological Institute, Czechia, EURAMET |
| DFM |
Danish Fundamental Metrology A/S, Denmark, EURAMET |
| IH-LQPM |
Instituto Hidrográfico - Laboratório de Química e Poluição do Meio Marinho, Portugal, EURAMET |
| IRA |
Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET |
| LNE |
Laboratoire national de métrologie et d'essais, France, EURAMET |
| RISE |
Research Institutes of Sweden AB, Sweden, EURAMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :