CCQM-K170 and EURAMET.QM-K170

MEASURAND   Electrolytic conductivity of a KCl solution

NOMINAL VALUE 0.5 S.m-1

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in S.m-1.

Electrolytic conductivity at 0.5 S m-1
Di / S·m-1
Participating institutes

Enlarged graph

CCQM-K170_Graph 0.5

CCQM-K170 and EURAMET.QM-K170

MEASURAND   Electrolytic conductivity of a KCl solution

NOMINAL VALUE 20 S.m-1

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in S.m-1.

Electrolytic conductivity at 20 S m-1
Di / S·m-1
Participating institutes

Enlarged graph

CCQM-K170_Graph-20

CCQM-K170 and EURAMET.QM-K170

MEASURAND   Electrolytic conductivity of a KCl solution

NOMINAL VALUE 0.5 S.m-1

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in S.m-1.

LABi Di Uneg,i Upos,i
S·m-1
S·m-1
S·m-1
INTI -0.02100 0.01400
VNIIFTRI -0.00131 0.00050
CENAM -0.00100 0.01000
UME -0.00110 0.00200
NIMT -0.00050 0.00180
BFKH -0.00033 0.00070
NMIJ -0.00025 0.00091
NIM -0.00004 0.00042
RISE 0.00000 0.00130
INM (CO) 0.00001 0.00097
SMU 0.00004 0.00043
INACAL 0.00023 0.00097
CMI 0.00030 0.00100
LATU 0.00033 0.00084
KazStandard 0.00041 0.00094
IBMETRO 0.00050 0.00160
INMETRO 0.00075 0.00056
INTI -0.01200 0.01021
KEBS -0.00370 0.02716
KazStandard -0.00094 0.00139
PTB -0.00055 0.00053
SNSU-BSN -0.00034 0.00130
GUM -0.00010 0.00055
DFM 0.00015 0.00082
SE "Ukrmetrteststandard" -0.00084 0.00138
Participating institutes
CCQM-K170
EURAMET.QM-K170

CCQM-K170 and EURAMET.QM-K170

MEASURAND   Electrolytic conductivity of a KCl solution

NOMINAL VALUE 20 S.m-1

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in S.m-1.

LABi Di Uneg,i Upos,i
S·m-1
S·m-1
S·m-1
VNIIM -0.543 0.027
CENAM -0.332 0.065
INM(CO) -0.038 0.048
NIMT -0.038 0.050
LATU -0.017 0.038
NIM -0.009 0.028
RISE 0.008 0.057
NMIJ 0.009 0.048
INMETRO 0.018 0.044
IBMETRO 0.050 0.760
BFKH 0.116 0.029
KazStandard 0.137 0.033
SE "Ukrmetrteststandard" -0.170 0.059
DFM -0.020 0.032
PTB -0.017 0.031
GUM 0.006 0.045
KazStandard 0.021 0.040
CMI 0.180 0.060
KEBS 0.460 0.038
Participating institutes
CCQM-K170
EURAMET.QM-K170
Comparison
Comparison close
CC comparison
Linked comparison
EURAMET.QM-K170
Metrology area, Sub-field Chemistry and Biology, Electrochemistry
Description Measurement of electrolytic conductivity
Time of measurements 2024 - 2025
Status Approved for equivalence
Final Reports of the comparisons
Measurand Electrolytic conductivity
0.5 S/m and 20 S/m
Transfer device 2 samples of aqueous KCl solutions
Comparison type Key Comparison
Consultative Committee CCQM (Consultative Committee for Amount of Substance)
Conducted by EURAMET (European Association of National Metrology Institutes)
Pilot institute PTB
Physikalisch-Technische Bundesanstalt
Germany
Contact person Steffen SEITZ

+49 531 592 3130
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

CMI

Czech Metrology Institute, Czechia, EURAMET

DFM

Danish Fundamental Metrology A/S, Denmark, EURAMET

GUM

Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET

INTI

Instituto Nacional de Tecnologia Industrial, Argentina, SIM

KEBS

Kenya Bureau of Standards, Kenya, AFRIMETS

KazStandard

Kazakhstan Institute of Standardization and Metrology, Kazakhstan, COOMET

NIM

National Institute of Metrology, China, APMP

SE "Ukrmetrteststandard"

State Enterprise "Ukrmetrteststandard", Ukraine, EURAMET

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

SNSU-BSN

National Measurement Standard - National Standardization Agency of Indonesia, Indonesia, APMP

CCQM-K170 and EURAMET.QM-K170

 

MEASURAND   Electrolytic conductivity of a KCl solution

NOMINAL VALUE 0.5 S.m-1

 

CCQM-K170

The key comparison reference value xR and its associated standard uncertainty uR are determined from the median of eight results. Secondary measurement results traceable to CRMs from SMU (a participating NMI) and those using non-NMI CRMs as calibrants were excluded. Secondary measurement results traceable to DFM (a non-participating NMI) were first combined into a single result using the uncertainty-weighted mean without dispersion before being included in the xcalculation (see section 8 of the CCQM-K170 Final report).

 

        xR = 0.50150 S.m-1     uR = 0.00018 S.m-1

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in S.m-1:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence calculated as described in section 9 of the CCQM-K170 Final report.

 

Linking EURAMET.QM-K170 to CCQM-K170

NIM and SMU provide the link between the key comparison CCQM-K170 and URAMET.QM-K170, having participated in both comparisons. The linking process is described in Section 9 of the EURAMET.QM-K170 Final report.

CCQM-K170 and EURAMET.QM-K170

MEASURAND   Electrolytic conductivity of a KCl solution

NOMINAL VALUE 20 S.m-1

 

CCQM-K170

The key comparison reference value xR and its associated standard uncertainty uR are determined from the median of eight results. Secondary measurement results traceable to CRMs from SMU (a participating NMI) and those using non-NMI CRMs as calibrants were excluded. Secondary measurement results traceable to DFM (a non-participating NMI) were first combined into a single result using the uncertainty-weighted mean without dispersion before being included in the xcalculation (see section 8 of the CCQM-K170 Final report).

 

        xR = 20.148 S.m-1     uR = 0.014 S.m-1

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in S.m-1:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence calculated as described in section 9 of the CCQM-K170 Final report.

 

Linking EURAMET.QM-K170 to CCQM-K170

NIM provides the link between the key comparison CCQM-K170 and EURAMET.QM-K170, having participated in both comparisons. The linking process is described in Section 9 of the EURAMET.QM-K170 Final report.

CCQM-K170 and EURAMET.QM-K170

 

MEASURAND   Electrolytic conductivity of a KCl solution

NOMINAL VALUE 0.5 S.m-1

 

xi   measurement value submitted by institute iLabi

ui   combined standard uncertainty of xi

 

CCQM-K170

  Labi xi ui
  S·m-1 S·m-1
  INTI 0.4802 0.0052
  VNIIFTRI 0.500184 0.00020
  CENAM 0.5003 0.0051
  UME 0.5004 0.0010
  NIMT 0.5010 0.00086
  BFKH 0.50116 0.000345
  NMIJ 0.50125 0.00048
  NIM 0.50146 0.000125
  RISE 0.5015 0.0006
  INM (CO) 0.50151 0.00045
  SMU 0.50153 0.00013
  INACAL 0.50172 0.00045
  CMI 0.5018 0.00055
  LATU 0.50182 0.00038
  KazStandard 0.501901279 0.000436
  IBMETRO 0.5020 0.0008
  INMETRO 0.50225 0.00025

 

 

EURAMET.QM-K170

  Labi xi ui
  S m-1 S m-1
  INTI 0.4876 0.0051
  KEBS 0.49570 0.01358
  KazStandard 0.4984940 0.0006506
  PTB 0.49889 0.00010
  SNSU-BSN 0.4991 0.0006
  GUM 0.49934 0.00012
  DFM 0.499590 0.000175
  SE "Ukrmetrteststandard" 0.498600 0.000645

CCQM-K170 and EURAMET.QM-K170

MEASURAND   Electrolytic conductivity of a KCl solution

NOMINAL VALUE 20 S.m-1

 

xi   measurement value submitted by institute iLabi

ui   combined standard uncertainty of xi

 

CCQM-K170

  Labi xi ui
  S·m-1 S·m-1
  VNIIM 19.6049 0.0016
  CENAM 19.8159 0.0293
  INM (CO) 20.110 0.024
  NIMT 20.110 0.021
  LATU 20.131 0.013
  NIM 20.138 0.005
  RISE 20.156 0.025
  NMIJ 20.157 0.024
  INMETRO 20.166 0.021
  IBMETRO 20.1935 0.3717
  BFKH 20.2636 0.00627
  KazStandard 20.28405 0.00912

 

 

EURAMET.QM-K170

  Labi xi ui
  S m-1 S m-1
  SE "Ukrmetrteststandard" 19.733 0.0253
  DFM 19.8810 0.0065
  PTB 19.884 0.005
  GUM 19.907 0.017
  KazStandard 19.9224 0.0134
  CMI 20.077 0.026
  KEBS 20.36 0.012