Comparison
Comparison close
Results

BIPM.RI(II)-K1.Tb-161

 

MEASURAND           Equivalent activity of Tb-161

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Tb-161
Di / MBq
Participating institutes

Results given in absolute values

To add comparison identifier and To add comparison identifier

MEASURAND           Equivalent activity of Tb-161

 

Degrees of equivalence represented by Di = (xI - xR) / xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in KBq/MBq.

Tb-161
Di / xR / (kBq/MBq)
Participating institutes

Results given in relative values

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Tb-161

 

MEASURAND           Equivalent activity of Tb-161

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
IRA 5 13
NPL -3.1 7.4
Participating institutes
BIPM.RI(II)-K1.Tb-161

Results given in absolute values

MEASURAND           Equivalent activity of Tb-161

 

Degrees of equivalence represented by Di = (xI - xR) / xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in KBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
IRA 3.1 7.7
NPL -1.8 4.4
Participating institutes
BIPM.RI(II)-K1.Tb-161

Results given in relative values

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K1.Tb-161
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Tb-161
Time of measurements 2019
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
First results published on 23 April 2021.
Last updated results on 12 February 2024.
 
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

IRA

Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET

NPL

National Physical Laboratory, United Kingdom, EURAMET


Comparison
Comparison close
Results

BIPM.RI(II)-K1.Tb-161

 

MEASURAND           Equivalent activity of Tb-161

 

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is  xR = 1704.7 MBq with a standard uncertainty uR = 4.1 MBq

(see Section 4.1 of the Final report).  The value xi is taken as the equivalent activity for a laboratory i.

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (x- xR) and Uits expanded uncertainty (k = 2), both expressed in MBq, and

Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

 

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Tb-161

 

MEASURAND           Equivalent activity of Tb-161

 

xi   result of measurement carried out in the SIR for the sample submitted by laboratory i, Labi

ui   combined standard uncertainty of xi

 

  Labi xi ui Date of measurement
    kBq kBq  
  IRA 1710 10 2019-08-29
  NPL 1701.6 3.4 2022-03-17