MEASURAND : Density of liquids
LIQUID : Water at 20 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID :Pentadecane at 20 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID :Pentadecane at 60 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 20 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 5 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID :Viscosity oil VO-2 at 20 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID : Water at 20 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

 

 

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 20 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

 

 

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 5 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Viscosity oil VO-2 at 20 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

Comparison
Comparison close
CC comparison
Linked comparison
EURAMET.M.D-K2
Metrology area, Sub-field Mass and related quantities, Density
Description Density of liquids
Time of measurements 2008 - 2009
Status Approved for equivalence
Final Reports of the comparisons
Measurand Density: 700 kg/m3 to 1650 kg/m3
Parameters Temperature: 5 °C to 60 °C
Atmospheric pressure
Transfer device Water, pentadecane, tetrachloroethylene and high-viscosity oil samples
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET Project No 1019
Comments

Results published on 30 October 2015

EURAMET.M.D-K2 results are linked to those of CCM.D-K2

Density of liquids

Pilot institute BEV
Bundesamt für Eich- und Vermessungswesen
Austria
Contact person Christian BUCHNER

+43 1 21 110 6361
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BEV

Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET

CEM

Centro Español de Metrologia, Spain, EURAMET

DZM

State Office for Metrology, Croatia, EURAMET

GUM

Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET

INM

National Institute of Metrology, Romania, EURAMET

INRIM

Istituto Nazionale di Ricerca Metrologica, Italy, EURAMET

IPQ

Instituto Portugues da Qualidade, Portugal, EURAMET

JV

Justervesenet, Norway, EURAMET

LNE

Laboratoire national de métrologie et d'essais, France, EURAMET

MIKES

VTT Technical Research Centre of Finland Ltd, Centre for Metrology/Mittatekniikan keskus, Finland, EURAMET

NPL

National Physical Laboratory, United Kingdom, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

UME

TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET

MEASURAND : Density of liquids
LIQUID : Water at 20 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure B explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report).
The key comparison reference value is computed by a Monte Carlo method using the median. The distribution is usually non-symmetric and the confidence interval may not be equally spaced around the reference value.

xR = 998.3220 kg/m3, UR lower = 0.0038 kg/m3 and UR upper = 0.0038 kg/m3

 

In Procedure B, the degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated by the Monte Carlo method described in Metrologia, 2012, 39, 589-595.

 

MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 768.5652 kg/m3, UR = 0.0027 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure B explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report).
The key comparison reference value is computed by a Monte Carlo method using the median. The distribution is usually non-symmetric and the confidence interval may not be equally spaced around the reference value.

xR = 772.0696kg/m3, UR lower = 0.0047 kg/m3 and UR upper = 0.0053 kg/m3

In Procedure B, the degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated by the Monte Carlo method described in Metrologia, 2012, 39, 589-595.

MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 754.5871 kg/m3, UR = 0.0040 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 740.6078 kg/m3, UR = 0.0048 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 20 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 1622.7252 kg/m3, UR = 0.0106 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 5 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 1647.5491 kg/m3, UR = 0.0119 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

MEASURAND : Density of liquids
LIQUID : Viscosity oil VO-2 at 20 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 845.6817 kg/m3, UR = 0.0052 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

MEASURAND : Density of liquids
LIQUID : Water at 20 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 9).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Tables 10 and 11).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 12).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 13).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 14).

MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 20 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 15).

MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 5 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 16).

MEASURAND : Density of liquids
LIQUID : Viscosity oil VO-2 at 20 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 17).