MEASURAND : Density of liquids
LIQUID : Water at 20 °C
Graph of equivalence
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
MEASURAND : Density of liquids
LIQUID :Pentadecane at 20 °C
Graph of equivalence
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C
Graph of equivalence
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C
Graph of equivalence
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
MEASURAND : Density of liquids
LIQUID :Pentadecane at 60 °C
Graph of equivalence
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 20 °C
Graph of equivalence
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 5 °C
Graph of equivalence
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
MEASURAND : Density of liquids
LIQUID :Viscosity oil VO-2 at 20 °C
Graph of equivalence
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
MEASURAND : Density of liquids
LIQUID : Water at 20 °C
Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
Results are presented under A4 printable format in Summary Results (PDF file).
MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C
Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
Results are presented under A4 printable format in Summary Results (PDF file).
MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C
Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
Results are presented under A4 printable format in Summary Results (PDF file).
MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C
Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
Results are presented under A4 printable format in Summary Results (PDF file).
MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C
Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
Results are presented under A4 printable format in Summary Results (PDF file).
MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 20 °C
Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
Results are presented under A4 printable format in Summary Results (PDF file).
MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 5 °C
Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
Results are presented under A4 printable format in Summary Results (PDF file).
MEASURAND : Density of liquids
LIQUID : Viscosity oil VO-2 at 20 °C
Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:
Red diamonds CCM.D-K2 participants
Blue diamonds EURAMET.M.D-K2 participants only
Results are presented under A4 printable format in Summary Results (PDF file).
Metrology area, Sub-field | Mass and related quantities, Density |
Description | Density of liquids |
Time of measurements | 2004 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
Measurand | Density: 700 kg/m3 to 1700 kg/m3 |
Parameters | Temperature: 5 °C to 60 °C Atmospheric pressure |
Transfer device | Pentadecane, water, tetrachloroethylene and high-viscosity oil samples |
Comparison type | Key Comparison |
Consultative Committee | CCM (Consultative Committee for Mass and Related Quantities) |
Conducted by | CCM (Consultative Committee for Mass and Related Quantities) |
Comments |
Results published on 27 March 2013 EURAMET.M.D-K2 results are linked to those of CCM.D-K2 Density of liquids
|
Pilot institute |
PTB
Physikalisch-Technische Bundesanstalt Germany |
Contact person | Horst BETTIN +49 (0) 531 592 3330 |
Pilot laboratory | |
---|---|
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
BFKH |
Government Office of the Capital City Budapest, Hungary, EURAMET |
CENAM |
Centro Nacional de Metrologia, Mexico, SIM |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
NRC |
National Research Council, Canada, SIM |
VNIIM |
D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
MEASURAND : Density of liquids
LIQUID : Water at 20 °C
The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure B explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report).
The key comparison reference value is computed by a Monte Carlo method using the median. The distribution is usually non-symmetric and the confidence interval may not be equally spaced around the reference value.
xR = 998.3220 kg/m3, UR lower = 0.0038 kg/m3 and UR upper = 0.0038 kg/m3
In Procedure B, the degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated by the Monte Carlo method described in Metrologia, 2012, 39, 589-595. |
MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C
The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.
xR = 768.5652 kg/m3, UR = 0.0027 kg/m3
The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2 – uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively. |
MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C
The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure B explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report).
The key comparison reference value is computed by a Monte Carlo method using the median. The distribution is usually non-symmetric and the confidence interval may not be equally spaced around the reference value.
xR = 772.0696kg/m3, UR lower = 0.0047 kg/m3 and UR upper = 0.0053 kg/m3
In Procedure B, the degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated by the Monte Carlo method described in Metrologia, 2012, 39, 589-595. |
MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C
The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.
xR = 754.5871 kg/m3, UR = 0.0040 kg/m3
The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2 – uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively. |
MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C
The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.
xR = 740.6078 kg/m3, UR = 0.0048 kg/m3
The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2 – uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively. |
MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 20 °C
The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.
xR = 1622.7252 kg/m3, UR = 0.0106 kg/m3
The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2 – uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively. |
MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 5 °C
The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.
xR = 1647.5491 kg/m3, UR = 0.0119 kg/m3
The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2 – uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively. |
MEASURAND : Density of liquids
LIQUID : Viscosity oil VO-2 at 20 °C
The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.
xR = 845.6817 kg/m3, UR = 0.0052 kg/m3
The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2 – uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively. |
MEASURAND : Density of liquids
LIQUID : Water at 20 °C
Measurements were carried from 27 April 2004 to 28 June 2004.
The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 9).
MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C
Measurements were carried from 27 April 2004 to 28 June 2004.
The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Tables 10 and 11).
MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C
Measurements were carried from 27 April 2004 to 28 June 2004.
The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 12).
MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C
Measurements were carried from 27 April 2004 to 28 June 2004.
The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 13).
MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C
Measurements were carried from 27 April 2004 to 28 June 2004.
The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 14).
MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 20 °C
Measurements were carried from 27 April 2004 to 28 June 2004.
The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 15).
MEASURAND : Density of liquids
LIQUID : Tetrachloroethylene at 5 °C
Measurements were carried from 27 April 2004 to 28 June 2004.
The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 16).
MEASURAND : Density of liquids
LIQUID : Viscosity oil VO-2 at 20 °C
Measurements were carried from 27 April 2004 to 28 June 2004.
The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 17).