CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Water at 20 °C

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kg m-3.

Deuterated water at 20 °C
Di / kg m-3
Participating institutes

Enlarged graph

H2O@20°C

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Tetrachloroethylene at 20 °C

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kg m-3.

Tetrachloroethylene at 20 °C
Di / kg m-3
Participating institutes

Enlarged graph

TCE@20°C

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Tetrachloroethylene at 5 °C

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kg m-3.

Tetrachloroethylene at 5 °C
Di / kg m-3
Participating institutes

Enlarged graph

TCE@5°C

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Viscosity oil at 20 °C

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kg m-3.

Viscosity oil at 20 °C
Di / kg m-3
Participating institutes

Enlarged graph

VO@20°C

MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID :Pentadecane at 20 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

MEASURAND : Density of liquids
LIQUID :Pentadecane at 60 °C

Graph of equivalence

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Water at 20 °C

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kg m-3.

LABi Di Uneg,i Upos,i
kg m-3
kg m-3
kg m-3
NRC -0.0105 0.0149 0.0126
PTB 0.0045 0.0045 0.0040
MKEH -0.0012 0.0058 0.0045
NMIJ 0.0020 0.0137 0.0140
KRISS -0.0050 0.0050 0.0050
CENAM 0.000 0.0051 0.0050
VNIIM 0.0125 0.0069 0.0072
BEV 0.0025 0.0053 0.0053
CEM -0.1040 0.0540 0.0540
GUM 0.0016 0.0046 0.0046
INM 0.0349 0.0119 0.0119
INRiM 0.0132 0.0058 0.0058
IPQ 1 -0.8770 0.0690 0.0690
IPQ 2 -0.0298 0.0345 0.0345
LNE -0.0096 0.0093 0.0093
MIKES 0.0030 0.0076 0.0076
NPL 0.0034 0.0069 0.0069
SMU 0.0110 0.0127 0.0127
UME 0.0168 0.0111 0.0111
INM 0.0131 0.0096 0.0096
CMI 0.008 0.023 0.023
DMDM -0.002 0.011 0.011
GUM -0.011 0.016 0.016
IMBIH 0.009 0.013 0.013
IPQ 2 0.0104 0.0087 0.0087
JV -0.029 0.081 0.081
UME 0.007 0.011 0.011
Participating institutes
CCM.D-K2
EURAMET.M.D-K2
EURAMET.M.D-K2.1

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Tetrachloroethylene at 20 °C

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kg m-3.

LABi Di Uneg,i Upos,i
kg m-3
kg m-3
kg m-3
NRC 0.0258 0.0935
PTB -0.0054 0.0204
MKEH 0.0008 0.0197
NMIJ -0.0211 0.0269
KRISS 0.0091 0.0212
CENAM 0.0162 0.0371
VNIIM 0.0028 0.0255
BEV -0.0021 0.0151
GUM -0.0185 0.0175
INM 0.0278 0.0236
INRiM -0.0038 0.0240
IPQ 1 -2.3137 0.1314
IPQ 2 0.0472 0.0664
LNE -0.0482 0.0302
NPL -0.1660 0.0285
SMU -0.0042 0.0245
UME 0.0077 0.0301
INM -0.334 0.053
CMI 0.056 0.096
DMDM -0.077 0.060
GUM -0.011 0.054
IPQ 2 -0.109 0.054
JV -0.404 0.078
UME 0.011 0.056
Participating institutes
CCM.D-K2
EURAMET.M.D-K2
EURAMET.M.D-K2.1

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Tetrachloroethylene at 5 °C

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kg m-3.

LABi Di Uneg,i Upos,i
kg m-3
kg m-3
kg m-3
NRC 0.0249 0.1200
PTB 0.0052 0.0197
MKEH -0.0068 0.0195
NMIJ -0.0143 0.0279
KRISS 0.0070 0.0205
CENAM 0.0149 0.0543
BEV -0.0005 0.0200
GUM -0.0151 0.0287
LNE -0.0107 0.0391
SMU 0.0113 0.0276
UME 0.0410 0.0353
INM -0.350 0.054
CMI 0.299 0.096
DMDM -0.135 0.060
GUM 0.003 0.054
IPQ 2 -0.047 0.055
UME 0.020 0.057
Participating institutes
CCM.D-K2
EURAMET.M.D-K2
EURAMET.M.D-K2.1

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Viscosity oil at 20 °C

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kg m-3.

LABi Di Uneg,i Upos,i
kg m-3
kg m-3
kg m-3
NRC -0.0187 0.0585
PTB 0.0079 0.0087
MKEH -0.0030 -0.0106
KRISS -0.0078 -0.0089
CENAM 0.0085 0.0143
VNIIM -0.0025 -0.0119
BEV 0.0049 0.0079
CEM -0.2047 0.0401
GUM 0.0091 0.0083
INM 0.0163 0.0126
INRiM 0.0099 0.0116
IPQ 1 -1.0698 0.0741
IPQ 2 -0.2159 0.0376
LNE 0.0206 0.0136
NPL 0.0034 0.0095
SMU 0.0383 0.0157
UME 0.0243 0.0137
INM 0.055 0.046
CMI 0.011 0.054
DMDM 0.197 0.046
GUM -0.009 0.046
IMBIH -0.059 0.060
IPQ 2 0.021 0.046
JV 0.010 0.055
UME 0.001 0.046
Participating institutes
CCM.D-K2
EURAMET.M.D-K2
EURAMET.M.D-K2.1

MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

 

 

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C

Degrees of equivalence relative to the CCM.D-K2 key comparison reference value:

Red diamonds CCM.D-K2 participants

Blue diamonds EURAMET.M.D-K2 participants only

Results are presented under A4 printable format in Summary Results (PDF file).

Comparison
Comparison close
CC comparison
Linked comparison
CCM.D-K2
Metrology area, Sub-field Mass and related quantities, Density
Description Density of liquids
Time of measurements 2004
Status Approved for equivalence
Final Reports of the comparisons
Measurand Density: 700 kg/m3 to 1700 kg/m3
Parameters Temperature: 5 °C to 60 °C
Atmospheric pressure
Transfer device Pentadecane, water, tetrachloroethylene and high-viscosity oil samples
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by CCM (Consultative Committee for Mass and Related Quantities)
Comments

 

 

Results published on 27 March 2013

EURAMET.M.D-K2 results are linked to those of CCM.D-K2

Density of liquids

 

 

Pilot institute PTB
Physikalisch-Technische Bundesanstalt
Germany
Contact person Horst BETTIN

+49 (0) 531 592 3330
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

BFKH

Government Office of the Capital City Budapest, Hungary, EURAMET

CENAM

Centro Nacional de Metrologia, Mexico, SIM

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NRC

National Research Council, Canada, SIM

VNIIM

D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Water at 20 °C

 

CCM.D-K2

The key comparison reference value xR and its associated expanded uncertainty at a 95% level of confidence, UR are calculated according to section 5 of the CCM.D-K2 Final Report, where the key comparison reference value is computed by a Monte Carlo method using the median. The distribution is usually non-symmetric and the confidence interval may not be equally spaced around the reference value.

        xR = 998.3220 kg/m3     UR lower = 0.0038 kg/m3  and  UR upper = 0.0038 kg/m3

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in kg/m3:

Di and its expanded uncertainty Ui at a 95 % level of confidence are calculated by the Monte Carlo method according to section 8 of the CCM.D-K2 Final Report.

 

Linking EURAMET.M.D-K2 to CCM.D-K2

The PTB and MKEH provides the link between the key comparison CCM.D-K2 and EURAMET.M.D-K2, having participated in both comparisons. The linking process is described in Section 4 of the EURAMET.M.D-K2 Final Report.

 

Linking EURAMET.M.D-K2.1 to CCM.D-K2

The BEV and PTB provide the link between the key comparison CCM.D-K2 and EURAMET.M.D-K2.1. BEV's results are linked to CCM.D-K2 through its participation in EURAMET.M.D-K2, while PTB participated in CCM.D-K2 and provided the link for EURAMET.M.D-K2. The linking process is described in Section 5 of the EURAMET.M.D-K2.1 Final Report.

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Tetrachloroethylene at 20 °C

 

CCM.D-K2

The key comparison reference value xR and its associated expanded uncertainty at a 95% level of confidence, UR are calculated according to Procedure A explained in section 5 of the CCM.D-K2 Final Report. It is the weighted mean of the results.

        xR = 1622.7252 kg/m3     UR = 0.0106 kg/m3 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in kg/m3:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 (ui2 – uR2)1/2  where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

 

Linking EURAMET.M.D-K2 to CCM.D-K2

The PTB and MKEH provides the link between the key comparison CCM.D-K2 and EURAMET.M.D-K2, having participated in both comparisons. The linking process is described in Section 4 of the EURAMET.M.D-K2 Final Report.

 

Linking EURAMET.M.D-K2.1 to CCM.D-K2

The BEV and PTB provide the link between the key comparison CCM.D-K2 and EURAMET.M.D-K2.1. BEV's results are linked to CCM.D-K2 through its participation in EURAMET.M.D-K2, while PTB participated in CCM.D-K2 and provided the link for EURAMET.M.D-K2. The linking process is described in Section 5 of the EURAMET.M.D-K2.1 Final Report.

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Tetrachloroethylene at 5 °C

 

CCM.D-K2

The key comparison reference value xR and its associated expanded uncertainty at a 95% level of confidence, UR are calculated according to Procedure A explained in section 5 of the CCM.D-K2 Final Report. It is the weighted mean of the results.

        xR = 1647.5491 kg/m3     UR = 0.0119 kg/m3 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in kg/m3:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 (ui2 – uR2)1/2  where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

 

Linking EURAMET.M.D-K2 to CCM.D-K2

The PTB and MKEH provides the link between the key comparison CCM.D-K2 and EURAMET.M.D-K2, having participated in both comparisons. The linking process is described in Section 4 of the EURAMET.M.D-K2 Final Report.

 

Linking EURAMET.M.D-K2.1 to CCM.D-K2

The BEV and PTB provide the link between the key comparison CCM.D-K2 and EURAMET.M.D-K2.1. BEV's results are linked to CCM.D-K2 through its participation in EURAMET.M.D-K2, while PTB participated in CCM.D-K2 and provided the link for EURAMET.M.D-K2. The linking process is described in Section 5 of the EURAMET.M.D-K2.1 Final Report.

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Viscosity oil at 20 °C

 

CCM.D-K2

The key comparison reference value xR and its associated expanded uncertainty at a 95% level of confidence, UR are calculated according to Procedure A explained in section 5 of the CCM.D-K2 Final Report. It is the weighted mean of the results.

        xR = 845.6817 kg/m3     UR = 0.0052 kg/m3 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in kg/m3:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 (ui2 – uR2)1/2  where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

 

Linking EURAMET.M.D-K2 to CCM.D-K2

The PTB and MKEH provides the link between the key comparison CCM.D-K2 and EURAMET.M.D-K2, having participated in both comparisons. The linking process is described in Section 4 of the EURAMET.M.D-K2 Final Report.

 

Linking EURAMET.M.D-K2.1 to CCM.D-K2

The BEV and PTB provide the link between the key comparison CCM.D-K2 and EURAMET.M.D-K2.1. BEV's results are linked to CCM.D-K2 through its participation in EURAMET.M.D-K2, while PTB participated in CCM.D-K2 and provided the link for EURAMET.M.D-K2. The linking process is described in Section 5 of the EURAMET.M.D-K2.1 Final Report.

MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure B explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report).
The key comparison reference value is computed by a Monte Carlo method using the median. The distribution is usually non-symmetric and the confidence interval may not be equally spaced around the reference value.

xR = 772.0696kg/m3, UR lower = 0.0047 kg/m3 and UR upper = 0.0053 kg/m3

In Procedure B, the degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated by the Monte Carlo method described in Metrologia, 2012, 39, 589-595.

MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 768.5652 kg/m3, UR = 0.0027 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 754.5871 kg/m3, UR = 0.0040 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C

The key comparison reference value, xR , and associated expanded uncertainty at a 95% level of confidence, UR, are calculated according to Procedure A explained in Metrologia, 2012, 39, 589-595 (see page 17 of the Final Report). It is the weighted mean of the results.

xR = 740.6078 kg/m3, UR = 0.0048 kg/m3

The degree of equivalence, Di, of laboratory i with respect to the key comparison reference value is calculated as the offset to xR together with its expanded uncertainty Ui = 2(ui2uR2)1/2, where ui and uR are the standard uncertainties of the laboratory result and of the key comparison reference value, respectively.

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Water at 20 °C

 

CCM.D-K2

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2004.

  Labi xi Ui
  kg m-3 kg m-3
  NRC* 998.3122 0.0149
  PTB 998.3260 0.0025
  OMH* 998.3215 0.0049
  NMIJ 998.3235 0.0151
  KRISS 998.3165 0.0042
  CENAM 998.3215 0.0046
  VNIIM 998.3340 0.0060

*The results shown are corrected values (see Section 5 of the CCM.D-K2 Final Report ).

 

EURAMET.M.D-K2

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2008.

  Labi xi Ui
  kg m-3 kg m-3
  BEV 998.5215 0.0035
  CEM 998.4151 0.0536
  GUM 998.5206 0.0021
  INM 998.5539 0.0112
  INRIM 998.5322 0.0042
  IPQ 1 997.6420 0.0686
  IPQ 2 998.4891 0.0343
  LNE 998.5094 0.0084
  MIKES 998.5220 0.0064
  MKEH 998.5235 0.0047
  NPL 998.5224 0.0055
  PTB 998.5207 0.0031
  SMU 998.5300 0.0120
  UME 998.5358 0.0103

 

 

EURAMET.M.D-K2.1

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2021.

  Labi xi Ui
  kg m-3 kg m-3
  BEV 998.425 0.011
  INM 998.4336 0.0070
  CMI 998.428 0.022
  DMDM 998.4188 0.0084
  GUM 998.410 0.015
  IMBIH 998.429 0.012
  IPQ 2 998.4309 0.0056
  JV 998.392 0.080
  PTB 998.4240 0.0081
  UME 998.4275 0.0082

 

 

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Tetrachloroethylene at 20 °C

 

CCM.D-K2

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2004.

  Labi xi Ui
  kg m-3 kg m-3
  NRC 1622.751 0.0941
  PTB 1622.7198 0.0230
  OMH 1622.7260 0.0224
  NMIJ 1622.7041 0.0289
  KRISS 1622.7343 0.0237
  CENAM 1622.7414 0.0386
  VNIIM 1622.728 0.0276

Uncertainties are corrected to include the inhomegeneity of the liquid. See section 5.3 of CCM.D-K2 Final Report.

 

EURAMET.M.D-K2

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2008.

  Labi xi Ui
  kg m-3 kg m-3
  BEV 1622.6741 0.0051
  CEM - -
  GUM 1622.6577 0.0102
  INM 1622.7040 0.0189
  INRIM 1622.6723 0.0194
  IPQ 1 1620.3625 0.1306
  IPQ 2 1622.7234 0.0649
  LNE 1622.6280 0.0267
  MIKES - -
  MKEH 1622.6700 0.0100
  NPL 1622.5102 0.0247
  PTB 1622.6777 0.0037
  SMU 1622.6720 0.0200
  UME 1622.6839 0.0266

- Did not measure

 

EURAMET.M.D-K2.1

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2021.

  Labi xi Ui
  kg m-3 kg m-3
  BEV 1622.595 0.075
  INM 1622.2640 0.0075
  CMI 1622.654 0.080
  DMDM 1622.521 0.028
  GUM 1622.587 0.012
  IMBIH - -
  IPQ 2 1622.487 0.013
  JV 1622.194 0.058
  PTB 1622.593 0.074
  UME 1622.609 0.019

-Did not measure

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Tetrachloroethylene at 5 °C

 

CCM.D-K2

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2004.

  Labi xi Ui
  kg m-3 kg m-3
  NRC 1647.574 0.1206
  PTB 1647.5543 0.0230
  OMH 1647.5423 0.0229
  NMIJ 1647.5348 0.0304
  KRISS 1647.5561 0.0237
  CENAM 1647.564 0.0556
  VNIIM - -

- Did not measure

Uncertainties are corrected to include the inhomegeneity of the liquid. (See section 5.3 of CCM.D-K2 Final Report)

 

EURAMET.M.D-K2

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2008.

  Labi xi Ui
  kg m-3 kg m-3
  BEV 1647.5042 0.0063
  CEM - -
  GUM 1647.4896 0.0215
  INM * *
  INRIM - -
  IPQ 1 - -
  IPQ 2 - -
  LNE 1647.4940 0.0342
  MIKES - -
  MKEH - -
  NPL - -
  PTB 1647.5099 0.0038
  SMU 1647.5160 0.0200
  UME 1647.5457 0.0297

- Did not measure

* Measurements were withdrawn

 

EURAMET.M.D-K2.1

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2021.

  Labi xi Ui
  kg m-3 kg m-3
  BEV 1647.425 0.077
  INM 1647.0747 0.0075
  CMI 1647.724 0.080
  DMDM 1647.290 0.028
  GUM 1647.427 0.011
  IMBIH - -
  IPQ 2 1647.377 0.013
  JV - -
  PTB 1647.4280 0.074
  UME 1647.445 0.019

- Did not measure

CCM.D-K2, EURAMET.M.D-K2 and EURAMET.M.D-K2.1

MEASURAND: Density of liquids

           LIQUID:  Viscosity oil at 20 °C

 

CCM.D-K2

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2004.

  Labi xi Ui
  kg m-3 kg m-3
  NRC 845.6630 0.0587
  PTB 845.6896 0.0101
  OMH 845.6787 0.0118
  NMIJ - -
  KRISS 845.6739 0.0103
  CENAM* 845.6797 0.0152
  VNIIM 845.6792 0.0130

- Did not measure

*This value was corrected for atmospheric pressure

Uncertainties are corrected to include the inhomegeneity of the liquid. See section 5.3 of CCM.D-K2 Final Report.

 

EURAMET.M.D-K2

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2008.

  Labi xi Ui
  kg m-3 kg m-3
  BEV 831.9436 0.0035
  CEM 831.7340 0.0395
  GUM 831.9477 0.0043
  INM 831.9550 0.0104
  INRIM 831.9485 0.0092
  IPQ 1 831.8688 0.0738
  IPQ 2 831.7228 0.0369
  LNE 831.9593 0.0117
  MIKES - -
  MKEH 831.9447 0.0049
  NPL 831.9421 0.0063
  PTB 831.9409 0.0021
  SMU 831.9770 0.0140
  UME 831.9629 0.0117

- Did not measure

 

EURAMET.M.D-K2.1

xi   measured value submitted by laboratory iLabi

Ui   expanded uncertainty of xi at 95 % level of confidence

Measurements were carried out in 2021.

  Labi xi Ui
  kg m-3 kg m-3
  BEV 830.337 0.064
  INM 830.4131 0.0075
  CMI 830.369 0.030
  DMDM 830.555 0.010
  GUM 830.3484 0.0078
  IMBIH 830.299 0.040
  IPQ 2 830.3783 0.0096
  JV 830.368 0.032
  PTB 830.391 0.064
  UME 830.3586 0.0080

MEASURAND : Density of liquids
LIQUID : Pentadecane at 15 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 12).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 20 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Tables 10 and 11).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 40 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 13).

MEASURAND : Density of liquids
LIQUID : Pentadecane at 60 °C

Measurements were carried from 27 April 2004 to 28 June 2004.

The results of the participants are given in Section 5 of the CCM.D-K2 Final Report (see Table 14).