Comparison
Comparison close
Results

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Equivalent activity of 177Lu
Di / MBq
Participating institutes

Results in absolute values.

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

Degrees of equivalence represented by Di = (xi - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Equivalent activity of 177Lu
Di / (kBq/MBq)
Participating institutes

Results in relative values.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
JRC 8 19
NPL 0.7 3.3
IFIN-HH -8.5 10
CMI 0.6 4.7
LNE-LNHB -0.2 3.9
IRA 0.1 4.3
ANSTO -3.0 3.6
ENEA-INMRI 5.3 7.9
LNMRI/IRD -1.0 7.2
NIST 0.6 4.4
NMISA 1.8 3.7
POLATOM -0.6 6.6
PTB -0.2 3.3
Participating institutes
BIPM.RI(II)-K1.Lu-177
CCRI.RI(II)-K2.Lu-177

Results in absolute values.

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

Degrees of equivalence represented by Di = (xi - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
JRC 13 34
NPL 1.3 6.0
IFIN-HH -15 18
CMI 1.1 8.4
LNE-LNHB -0.3 7.0
IRA 0.2 7.7
ANSTO -5.3 6.4
ENEA-INMRI 10 14
LNMRI/IRD -2 13
NIST 1.1 7.9
NMISA 3.3 6.7
POLATOM -1 12
PTB -0.3 5.9
Participating institutes
BIPM.RI(II)-K1.Lu-177
CCRI(II)-K2.Lu-177

Results in relative values.

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K1.Lu-177
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Lu-177
Time of measurements 2000 -
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System

SIR equivalent activity of the radionuclide 177Lu
Most recent results published on 27 October 2026
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Romain Coulon

First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

CMI

Czech Metrology Institute, Czechia, EURAMET


IFIN-HH

"Horia Hulubei" National Institute of Research and Development for Physics and Nuclear Engineering, Romania, EURAMET

IRA

Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET


JRC

Joint Research Centre, European Union, EURAMET

Comparison carried out under the institute name "IRMM"

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET


NIST

National Institute of Standards and Technology, United States, SIM

NPL

National Physical Laboratory, United Kingdom, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177

 

MEASURAND   Equivalent activity of 177Lu

 

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is 558 MBq, with a standard uncertainty uR equal to 1.1 MBq 

(see Section 4.1 of the BIPM.RI(II)-K1.Lu-177 2026 Final report).  The value xi is taken as the equivalent activity for a laboratory i.

 

The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: 

Di = (xi - xR) and Ui  its expanded uncertainty (k = 2), both expressed in MBq, and 

Ui  = 2((1 - 2wi )ui2 + uR2 )1/2 , where wi  is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xi - xR) /xR and Ui /xR , both expressed in kBq/MBq.

 

Linking CCRI(II)-K2.Lu-177 to BIPM.RI(II)-K1.Lu-177

The value xi is the equivalent activity for laboratory i participant in CCRI(II)-K2.Lu-177 having been normalized using the value of the NPL as the linking laboratory.

The degree of equivalence of laboratory i participant in CCRI(II)-K2.Lu-177 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value.

These statements make it possible to extend the BIPM.RI(II)-K1.Lu-177 matrices of equivalence to all participants in the CCRI(II)-K2.Lu-177 comparison

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

xi   measured mass fraction reported by laboratory i, Labi

ui   combined standard uncertainty of xi

 

BIPM.RI(II)-K1.Lu-177

 

  Labi xi ui Date of measurement
    kBq kBq  
  NPL 559.5 1.8 2009-04-28
  JRC 566.0 10.0 2009-04-30
  IFIN-HH 550.3 5.1 2013-07-11
  CMI 559.4 2.5 2023-09-12
  LNE-LNHB 558.6 2.1 2024-02-09
  IRA 558.9 2.3 2025-10-08

 

 

CCRI(II)-K2.Lu-177 

  Labi xi ui Date of measurement
  kBq kBq
  ANSTO 555.8 1.4 2009
  ENEA-INMRI 564.1 3.8 2009
  LNMRI/IRD 557.8 3.4 2009
  NIST 559.4 1.9 2009
  NMISA 560.6 1.5 2009
  POLATOM 558.2 3.1 2009
  PTB 558.6 1.2 2009