Comparison
Comparison close
Results

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Equivalent activity of 177Lu
Di / MBq
Participating institutes

Results in absolute values.

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

Degrees of equivalence represented by Di = (xi - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Equivalent activity of 177Lu
Di / (kBq/MBq)
Participating institutes

Results in relative values.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
JRC 8 19
NPL 0.7 3.3
IFIN-HH -8.5 10
CMI 0.6 4.7
LNE-LNHB -0.2 3.9
IRA 0.1 4.3
ANSTO -3.0 3.6
ENEA-INMRI 5.3 7.9
LNMRI/IRD -1.0 7.2
NIST 0.6 4.4
NMISA 1.8 3.7
POLATOM -0.6 6.6
PTB -0.2 3.3
Participating institutes
BIPM.RI(II)-K1.Lu-177
CCRI.RI(II)-K2.Lu-177

Results in absolute values.

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

Degrees of equivalence represented by Di = (xi - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
JRC 13 34
NPL 1.3 6.0
IFIN-HH -15 18
CMI 1.1 8.4
LNE-LNHB -0.3 7.0
IRA 0.2 7.7
ANSTO -5.3 6.4
ENEA-INMRI 10 14
LNMRI/IRD -2 13
NIST 1.1 7.9
NMISA 3.3 6.7
POLATOM -1 12
PTB -0.3 5.9
Participating institutes
BIPM.RI(II)-K1.Lu-177
CCRI(II)-K2.Lu-177

Results in relative values.

Comparison
Comparison close
BIPM comparison
Linked comparison
CCRI(II)-K2.Lu-177
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity concentration of the same Lu-177 solution
Time of measurements 2009
Status Approved for equivalence
Final Reports of the comparisons
References

CCRI(II)-K2.Lu-177 Technical Protocol

Measurand Activity concentration (15 MBq/g nominal)
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by CCRI (Consultative Committee for Ionizing Radiation)
Comments Results published on 20 December 2013
SIR equivalent activity of the radionuclide 177Lu
Results are linked to those of BIPM.RI(II)-K1.Lu-177
Pilot institute NIST
National Institute of Standards and Technology
United States
Contact person B. Zimmerman

+1 301 975 4338
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NIST

National Institute of Standards and Technology, United States, SIM

ANSTO

Australian Nuclear Science and Technology Organisation, Australia, APMP

ENEA-INMRI

Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, Italy, EURAMET

IFIN-HH

"Horia Hulubei" National Institute of Research and Development for Physics and Nuclear Engineering, Romania, EURAMET

IRMM

Institute for Reference Materials and Measurements, European Union, EURAMET

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

LNMRI/IRD

Laboratorio Nacional de Metrologia das Radiacoes Ionizantes, IRD, Brazil, SIM

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

NPL

National Physical Laboratory, United Kingdom, EURAMET

POLATOM

Radioisotope Centre Polatom, Poland, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177

 

MEASURAND   Equivalent activity of 177Lu

 

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is 558 MBq, with a standard uncertainty uR equal to 1.1 MBq 

(see Section 4.1 of the BIPM.RI(II)-K1.Lu-177 2026 Final report).  The value xi is taken as the equivalent activity for a laboratory i.

 

The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: 

Di = (xi - xR) and Ui  its expanded uncertainty (k = 2), both expressed in MBq, and 

Ui  = 2((1 - 2wi )ui2 + uR2 )1/2 , where wi  is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xi - xR) /xR and Ui /xR , both expressed in kBq/MBq.

 

Linking CCRI(II)-K2.Lu-177 to BIPM.RI(II)-K1.Lu-177

The value xi is the equivalent activity for laboratory i participant in CCRI(II)-K2.Lu-177 having been normalized using the value of the NPL as the linking laboratory.

The degree of equivalence of laboratory i participant in CCRI(II)-K2.Lu-177 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value.

These statements make it possible to extend the BIPM.RI(II)-K1.Lu-177 matrices of equivalence to all participants in the CCRI(II)-K2.Lu-177 comparison

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Lu-177 and CCRI(II)-K2.Lu-177 

 

MEASURAND   Equivalent activity of 177Lu

 

xi   measured mass fraction reported by laboratory i, Labi

ui   combined standard uncertainty of xi

 

BIPM.RI(II)-K1.Lu-177

 

  Labi xi ui Date of measurement
    kBq kBq  
  NPL 559.5 1.8 2009-04-28
  JRC 566.0 10.0 2009-04-30
  IFIN-HH 550.3 5.1 2013-07-11
  CMI 559.4 2.5 2023-09-12
  LNE-LNHB 558.6 2.1 2024-02-09
  IRA 558.9 2.3 2025-10-08

 

 

CCRI(II)-K2.Lu-177 

  Labi xi ui Date of measurement
  kBq kBq
  ANSTO 555.8 1.4 2009
  ENEA-INMRI 564.1 3.8 2009
  LNMRI/IRD 557.8 3.4 2009
  NIST 559.4 1.9 2009
  NMISA 560.6 1.5 2009
  POLATOM 558.2 3.1 2009
  PTB 558.6 1.2 2009