CCQM-K11 and CCQM-K11.2

MEASURAND : Mass fraction of glucose in human serum
 

Degrees of equivalence Di and expanded uncertainty Ui (95% level of confidence) expressed in mg/g.


Red diamonds: CCQM-K11
Blue squares: CCQM-K11.2

CCQM-K11 and CCQM-K11.2

MEASURAND : Mass fraction of glucose in human serum

Degrees of equivalence Di and expanded uncertainty Ui (95% level of confidence) expressed in mg/g.


 

Information is available in Summary Results 

Comparison
Comparison close
CC comparison
Linked comparison
CCQM-K11.2
Metrology area, Sub-field Chemistry and Biology, Organics
Description Glucose in human serum
Time of measurements 2012
Status Approved for equivalence
Final Reports of the comparisons
Measurand Mass fraction
Transfer device Samples from frozen human serum
Comparison type Key Comparison
Consultative Committee CCQM (Consultative Committee for Amount of Substance)
Conducted by CCQM (Consultative Committee for Amount of Substance)
Comments

This is a subsequent key comparison to CCQM-K11.

Results published on 15 June 2018

Glucose in human serum

The CCQM-K11.2 results are linked to the CCQM-K11 results.

Pilot institute NIST
National Institute of Standards and Technology
United States
Contact person M.J. Welch

+1 301 975 3100
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NIST

National Institute of Standards and Technology, United States, SIM

CENAM

Centro Nacional de Metrologia, Mexico, SIM

HSA

Health Science Authority, Singapore, APMP

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

LNE

Laboratoire national de métrologie et d'essais, France, EURAMET

NIM

National Institute of Metrology, China, APMP

NIMT

National Institute of Metrology (Thailand), Thailand, APMP

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

UME

TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET

VNIIM

D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET

CCQM-K11 and CCQM-K11.2

MEASURAND : Mass fraction of glucose in human serum

CCQM-K11

The key comparison reference value, xR, is calculated as the mean of the participant results; xR = 0.7762 mg/g.
The expanded uncertainty of the key comparison reference value at a 95% level of confidence is: UR = 0.0088 mg/g.

The degree of equivalence of each laboratory with respect to the key comparison reference value is given by a pair of terms: Di = (xi - xR) and Ui, its expanded uncertainty corresponding to a 95% confidence interval, both expressed in mg/g.

Linking CCQM-K11.2 to CCQM-K11

An independent key comparison reference value was assigned from the mean, standard deviation, and pooled standard uncertainty of the results reported by the three reference laboratories and the nine participants. It was agreed that although all results would be used to calculate the KCRV, degrees of equivalence would not be estimated for the three reference laboratories KRISS, NIST and PTB.

 

 

CCQM-K11 and CCQM-K11.2

MEASURAND : Mass fraction of glucose in human serum
 

xi : result of measurement carried out by laboratory i
ui
: combined standard uncertainty of xi

CCQM-K11

CCQM-K11.2