BIPM.RI(I)-K4, SIM.RI(I)-K4, EUROMET.RI(I)-K4, EURAMET.RI(I)-K4.2, EURAMET.RI(I)-K4.1 and APMP.RI(I)-K4

 

MEASURAND                 Absorbed dose to water

Radiation quality        Co-60 beam

 

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2)

Absorbed dose to water
Di  (mGy/Gy)
Participating institutes

The SIM.RI(I)-K4 results data are no longer included for Degrees of Equivalence being older than 15 years.

BIPM.RI(I)-K4, SIM.RI(I)-K4, EUROMET.RI(I)-K4, EURAMET.RI(I)-K4.2, EURAMET.RI(I)-K4.1 and APMP.RI(I)-K4

 

MEASURAND                 Absorbed dose to water

Radiation quality        Co-60 beam

 

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2)

LABi Di Uneg,i Upos,i
mGy/Gy
mGy/Gy
mGy/Gy
ENEA -0.1 8.8
VNIIFTRI -1.4 8.6
NMIJ -3.0 9.2
LNE-LNHB -1.9 7.8
METAS 1.1 10
PTB -1.3 7.6
NPL 2.3 14
VSL -3.0 9.6
BEV -0.3 8.2
GUM 3.0 7.0
ARPANSA -0.5 10
NRC -0.5 6.8
NIM 2.7 7.2
CMI -4.0 24
RMTC -5.3 12
SSM -1.4 10
STUK -3.9 8.5
NRPA 3.2 8.8
SMU -4.7 25
IAEA -0.4 10
HIRCL 3.0 12
ITN -7.1 13
NIST -0.6 11
LNMRI 1.0 15
CNEA 12 18
SCK-CEN -1.5 16
CIEMAT 2.3 11
VINS 0.0 14
Nuc. Malaysia -1.7 14
AECS-NRLM 0.7 24
KRISS -3.3 8.0
NMISA -0.1 11
BRMD, DMSc 1.8 11
BARC 3.2 11
PTKMR-BATAN 2.3 19
NRSL/INER -1.9 8.2
NIS 0.3 14
Participating institutes
BIPM.RI(I)-K4 (older that 10 y)
BIPM.RI(I)-K4
EUROMET.RI(I)-K4
EUROMET.RI(I)-K4.1
EUROMET.RI(I)-K4.2
APMP.RI(I)-K4

The SIM.RI(I)-K4 results data are no longer included for Degrees of Equivalence being older than 15 years.

SIM.RI(I)-K4.2019
Metrology area, Sub-field Ionizing Radiation, Section I (x and gamma rays, electrons)
Description Absorbed dose to water
Time of measurements 2019 - 2021
Status Measurements in progress
References

SIM.RI(I)-K4 Registration and progress form

SIM.RI(I)-K4 Technical Protocol

Measurand Absorbed dose to water
Parameters Co-60 beam
Transfer device Ionization chambers: waterproof Farmer-type
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by SIM (Inter-American Metrology System)
RMO Internal Identifier SIM.RI(I).K1-K4
Comments
 
Pilot institute NRC
National Research Council
Canada
Contact person John Paul Archambault

+1-613-993-2715x252
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NRC

National Research Council, Canada, SIM

CNEA

National Commission of Atomic Energy, Argentina, SIM

ININ

Instituto Nacional de Investigaciones Nucleares, Mexico, SIM

LNMRI/IRD

Laboratorio Nacional de Metrologia das Radiacoes Ionizantes, IRD, Brazil, SIM

MIEM-LSMRI

Laboratorios Tecnogestión - Ministerio de Industria, Energía y Minería, Uruguay, SIM

NIST

National Institute of Standards and Technology, United States, SIM

BIPM.RI(I)-K4, SIM.RI(I)-K4, EUROMET.RI(I)-K4, EURAMET.RI(I)-K4.2, EURAMET.RI(I)-K4.1 and APMP.RI(I)-K4

 

MEASURAND                 Absorbed dose to water

Radiation quality        Co-60 beam

 

The key comparison reference value is the BIPM evaluation of absrobed dose to water.

 

BIPM.RI(I)-K4

The degree of equivalence of each laboratory i with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy:
the relative difference Di and Ui, its expanded uncertainty (k = 2), with Ui = 2ui.

When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms both expressed in mGy/Gy:

Dij = Di - Dj  and Uij  , its expanded uncertainty (k = 2). The approximation for Uij is explained in section 5 of the Final Report.

 

 

Linking SIM.RI(I)-K4 to BIPM.RI(I)-K4

The relative value xi is the comparison result for laboratory i participant in SIM.RI(I)-K4 and was normalized to the value of NRC acting as the linking laboratory (see SIM.RI(I)-K4 Final Report).

The degree of equivalence of each laboratory i participant in SIM.RI(I)-K4 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).

 

 

Linking EUROMET.RI(I)-K4 to BIPM.RI(I)-K4

The relative value xi is the comparison result for laboratory i participant in EUROMET.RI(I)-K4 and is linked to the BIPM.RI(I)-K4 via the nine linking laboratories ARPANSA, BEV, ENEA-INMRI, LNE-LNHB, METAS, MKEH, NRC, PTB and VSL (see EUROMET.RI(I)-K4 Final Report).

The degree of equivalence of each laboratory i participant in EUROMET.RI(I)-K4 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).

 


Linking EURAMET.RI(I)-K4.1 to BIPM.RI(I)-K4

The linking of EURAMET.RI(I)-K4.1 to BIPM.RI(I)-K4 was made via BEV, LNE-LNHB and METAS, having participated in both comparisons.

The degree of equivalence of each laboratory i participant in EURAMET.RI(I)-K4.1 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).

 

Linking EURAMET.RI(I)-K4.2 to BIPM.RI(I)-K4

The linking of EURAMET.RI(I)-K4.2 to BIPM.RI(I)-K4 was made via BEV having participated in both comparisons.

The degree of equivalence of each laboratory i participant in EURAMET.RI(I)-K4.2 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).

 

Linking APMP.RI(I)-K4 to BIPM.RI(I)-K4

The linking of APMP.RI(I)-K4 to BIPM.RI(I)-K4 was made via ARPANSA, NMIJ-AIST and PTB having participated in both comparisons.

The degree of equivalence of each laboratory i participant in APMP.RI(I)-K4 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).


Following decision of the CCRI in 2011, the pair-wise degrees of equivalence are not explicitely published in the KCDB.

BIPM.RI(I)-K4, SIM.RI(I)-K4, EUROMET.RI(I)-K4, EURAMET.RI(I)-K4.2, EURAMET.RI(I)-K4.1 and APMP.RI(I)-K4

 

MEASURAND                 Absorbed dose to water

Radiation quality        Co-60 beam

 

BIPM.RI(I)-K4 - Laboratory measurements

xi : ratio of the measurement made at the BIPM by laboratory i to that of the BIPM
ui : combined standard uncertainty of the ratio xi

 

 

SIM.RI(I)-K4 - Laboratory measurements

xi : result of measurement carried out at laboratory i converted to absorbed dose through the NRC measurements at the BIPM (See SIM.RI(I)-K4 Final Report)
ui : combined standard uncertainty of xi

 

EUROMET.RI(I)-K4 - Laboratory measurements

xi : result of measurement carried out at laboratory i converted to absorbed dose through nine NMI measurements at the BIPM (See EUROMET.RI(I)-K4 Final Report)
The linking laboratories are MKEH, LNE-LNHB, PTB, ENEA, ARPANSA, NRC, BEV, METAS and VSL.
ui : combined standard uncertainty of xi

 

EURAMET.RI(I)-K4.2 - Laboratory measurements

xi : result of measurement carried out at laboratory i  linked via BEV, LNE-LNHB and METAS.
ui : combined standard uncertainty of xi

 

EURAMET.RI(I)-K4.1 - Laboratory measurements

xi : result of measurement carried out at laboratory i  linked via BEV.
ui : combined standard uncertainty of xi

 

 

APMP.RI(I)-K4 - Laboratory measurements

xi : result of measurement carried out at laboratory i  linked via ARPANSA, NMIJ-AIST and PTB.
ui : combined standard uncertainty of xi