CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 20 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 20 Hz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 31.5 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 31.5 Hz
Di  / dB re 1 V/Pa
Participating institutes

This is the possibility to place a comment.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 63 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 63 Hz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 125 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 125 Hz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 250 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 250 Hz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 500 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 500 Hz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 1 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 1 kHz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 2 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 2 kHz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 4 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 4 kHz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 8 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 8 kHz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 12.5 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 12.5 kHz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 16 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 16 kHz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 25 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

Pressure sensitivity level at 25 kHz
Di  / dB re 1 V/Pa
Participating institutes

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 20 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.007 0.029
HBK-DPLA -0.034 0.053
METAS -0.032 0.057
KRISS -0.018 0.072
NMIJ -0.004 0.063
GUM 0.021 0.044
UME 0.021 0.063
NMIA 0.014 0.091
CENAM 0.024 0.082
NRC 0.036 0.044
VNIIFTRI -0.009 0.053
NMISA 0.003 0.044
INMETRO -0.011 0.047
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 31.5 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.009 0.027
HBK-DPLA -0.026 0.044
METAS -0.025 0.052
KRISS -0.011 0.073
NMIJ -0.004 0.044
GUM 0.021 0.035
UME 0.021 0.044
NMIA 0.011 0.073
CENAM 0.022 0.063
NRC 0.031 0.035
VNIIFTRI -0.009 0.054
NMISA -0.001 0.044
INMETRO -0.015 0.043
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 63 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.014 0.028
HBK-DPLA -0.020 0.031
METAS -0.017 0.047
KRISS -0.010 0.045
NMIJ -0.005 0.036
GUM 0.010 0.026
UME 0.015 0.026
NMIA 0.007 0.054
CENAM 0.019 0.045
NRC 0.015 0.026
VNIIFTRI -0.010 0.054
NMISA -0.007 0.026
INMETRO -0.007 0.044
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 125 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.014 0.028
HBK-DPLA -0.011 0.026
METAS -0.009 0.044
KRISS 0.001 0.045
NMIJ -0.013 0.036
GUM 0.007 0.026
UME 0.017 0.026
NMIA 0.006 0.036
CENAM 0.018 0.045
NRC 0.012 0.026
VNIIFTRI -0.013 0.055
NMISA -0.008 0.026
INMETRO -0.007 0.044
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 250 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.016 0.029
HBK-DPLA -0.006 0.021
METAS -0.003 0.042
KRISS 0.008 0.036
NMIJ -0.012 0.036
GUM 0.008 0.026
UME 0.018 0.026
NMIA 0.004 0.036
CENAM 0.013 0.045
NRC 0.003 0.026
VNIIFTRI -0.012 0.055
NMISA -0.009 0.026
INMETRO -0.005 0.044
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 500 Hz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.017 0.029
HBK-DPLA -0.002 0.016
METAS -0.001 0.041
KRISS 0.013 0.036
NMIJ -0.010 0.036
GUM 0.010 0.026
UME 0.015 0.026
NMIA 0.002 0.036
CENAM 0.012 0.045
NRC 0.005 0.026
VNIIFTRI -0.015 0.055
NMISA -0.011 0.026
INMETRO -0.007 0.044
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 1 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.018 0.029
HBK-DPLA 0.001 0.011
METAS 0.000 0.041
KRISS 0.015 0.036
NMIJ -0.011 0.036
GUM 0.004 0.026
UME 0.014 0.026
NMIA 0.002 0.036
CENAM 0.010 0.045
NRC 0.004 0.026
VNIIFTRI -0.016 0.055
NMISA -0.013 0.026
INMETRO -0.009 0.044
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 2 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.019 0.029
HBK-DPLA 0.002 0.011
METAS 0.002 0.040
KRISS 0.019 0.036
NMIJ -0.013 0.036
GUM 0.007 0.026
UME 0.012 0.026
NMIA 0.001 0.036
CENAM 0.010 0.045
NRC 0.002 0.026
VNIIFTRI -0.018 0.055
NMISA -0.013 0.026
INMETRO -0.013 0.044
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 4 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.021 0.029
HBK-DPLA 0.004 0.011
METAS 0.003 0.039
KRISS 0.023 0.036
NMIJ -0.010 0.036
GUM 0.005 0.026
UME 0.015 0.036
NMIA -0.001 0.036
CENAM 0.008 0.055
NRC 0.005 0.026
VNIIFTRI -0.025 0.055
NMISA -0.013 0.026
INMETRO -0.023 0.044
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 8 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.029 0.031
HBK-DPLA 0.004 0.011
METAS 0.003 0.040
KRISS 0.023 0.036
NMIJ -0.021 0.036
GUM -0.016 0.045
UME 0.004 0.045
NMIA 0.002 0.036
CENAM 0.001 0.055
NRC 0.014 0.026
VNIIFTRI -0.026 0.055
NMISA -0.011 0.036
INMETRO -0.059 0.046
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 12.5 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.032 0.036
HBK-DPLA 0.005 0.010
METAS 0.008 0.058
KRISS 0.015 0.054
NMIJ -0.030 0.036
GUM -0.035 0.073
UME 0.000 0.054
NMIA 0.012 0.054
CENAM -0.007 0.092
NRC 0.025 0.036
VNIIFTRI -0.035 0.082
NMISA -0.004 0.036
INMETRO -0.117 0.046
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 16 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.033 0.044
HBK-DPLA 0.011 0.018
METAS 0.021 0.084
KRISS -0.007 0.063
NMIJ -0.037 0.044
GUM -0.057 0.072
UME -0.012 0.072
NMIA 0.026 0.072
CENAM -0.008 0.119
NRC 0.028 0.053
VNIIFTRI -0.047 0.119
NMISA 0.000 0.053
INMETRO -0.132 0.053
Participating institutes
CCAUV.A-K6

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 25 kHz

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui  at a 95 % level of confidence, expressed in dB re 1 V/Pa.

LABi Di Uneg,i Upos,i
dB
dB
dB
LNE -0.041 0.120
HBK-DPLA -0.003 0.049
METAS 0.065 0.382
KRISS 0.150 0.227
NMIJ 0.008 0.189
GUM -0.137 0.217
UME 0.023 0.180
NMIA 0.022 0.132
CENAM 0.021 0.366
NRC 0.013 0.339
VNIIFTRI 0.048 0.459
NMISA 0.006 0.132
INMETRO -0.014 0.153
Participating institutes
CCAUV.A-K6
Comparison
Comparison close
CC comparison
Linked comparison
CCAUV.A-K6
Metrology area, Sub-field Acoustics, Ultrasound, Vibration, Acoustics
Description Primary pressure calibration of laboratory standard microphones LS2P
Time of measurements 2019 - 2020
Status Approved for equivalence
Final Reports of the comparisons
References

CCAUV.A-K6 Technical Protocol

CCAUV.A-K6 Registration and progress form

Measurand Sensitivity level (dB ref 1V/Pa):
2 Hz - 20 Hz (optional), 20 Hz - 25 kHz (mandatory)
Sensitivity phase (degree):
2 Hz - 25 kHz (optional)

Transfer device Microphones Brüel & Kjær type 4180
serial numbers 2124385 and 2412872
Comparison type Key Comparison
Consultative Committee CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration)
Conducted by CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration)
Comments This comparison is a repeat of CCAUV.A-K3
Pilot institute LNE
Laboratoire national de métrologie et d'essais
France
Contact person Dominique Rodrigues

+33 1 30 69 14 87
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
LNE

Laboratoire national de métrologie et d'essais, France, EURAMET

CENAM

Centro Nacional de Metrologia, Mexico, SIM

GUM

Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET

HBK-DPLA

Hottinger Brüel & Kjaer A/S - Danish Primary Laboratory for Acoustics Section, Denmark, EURAMET

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

METAS

Federal Institute of Metrology, Switzerland, EURAMET

NMIA

National Measurement Institute, Australia, Australia, APMP

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

NRC

National Research Council, Canada, SIM

UME

TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET

VNIIFTRI

Institute of Physical Technical and Radiotechnical Measurements, Rosstandart, Russian Federation, COOMET

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 20 Hz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.076 0.016
  4180.241287 -37.806 0.016

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 31.5 Hz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.092 0.014
  4180.241287 -37.829 0.014

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 63 Hz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.113 0.011
  4180.241287 -37.847 0.011

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 125 Hz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.130 0.010
  4180.241287 -37.864 0.010

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 250 Hz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.142 0.010
  4180.241287 -37.875 0.010

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 500 Hz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.148 0.009
  4180.241287 -37.882 0.009

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 1 kHz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.146 0.009
  4180.241287 -37.882 0.009

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 2 kHz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.122 0.009
  4180.241287 -37.862 0.009

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 4 kHz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -38.017 0.009
  4180.241287 -37.772 0.009

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 8 kHz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -37.651 0.010
  4180.241287 -37.477 0.010

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 12.5 kHz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -37.315 0.011
  4180.241287 -37.294 0.011

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 16 kHz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Microphone xR UR
  dB dB
  4180.212439 -37.603 0.016
  4180.241287 -37.693 0.015

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 25 kHz

 

For each microphone, the key comparison reference value xR and its expanded uncertainty U(k = 2) are computed with exclusions as explained in section 7 of the CCAUV.A-K6 Final report (unit is dB re 1 V/Pa).

 

  Mic xR UR
  dB dB
  4180.212439 -41.939 0.045
  4180.241287 -42.057 0.046

 

The degree of equivalence of laboratory i relative to the reference value is given by a pair of terms expressed in dB re 1 V/Pa:

Di and Uits expanded uncertainty (k = 2). The degrees of equivalence are computed as explained in section 7 of CCAUV.A-K6 Final report.

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 20 Hz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.085 0.035
  HBK-DPLA -38.100 0.060
  METAS -38.100 0.061
  KRISS -38.10 0.08
  NMIJ -38.08 0.07
  GUM -38.06 0.05
  UME -38.05 0.07
  NMIA -38.06 0.10
  CENAM -38.053 0.09
  NRC -38.04 0.05
  VNIIFTRI -38.09 0.06
  NMISA -38.070 0.050
  INMETRO -38.087 0.050

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.811 0.035
  HBK-DPLA -37.849 0.060
  METAS -37.846 0.066
  KRISS -37.82 0.08
  NMIJ -37.81 0.07
  GUM -37.78 0.05
  UME -37.79 0.07
  NMIA -37.79 0.10
  CENAM -37.781 0.09
  NRC -37.77 0.05
  VNIIFTRI -37.81 0.06
  NMISA -37.806 0.050
  INMETRO -37.745 0.050

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 31.5 Hz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.105 0.035
  HBK-DPLA -38.112 0.050
  METAS -38.111 0.055
  KRISS -38.11 0.08
  NMIJ -38.10 0.05
  GUM -38.08 0.04
  UME -38.07 0.05
  NMIA -38.08 0.08
  CENAM -38.074 0.07
  NRC -38.06 0.04
  VNIIFTRI -38.11 0.06
  NMISA -38.094 0.050
  INMETRO -38.107 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.835 0.030
  HBK-DPLA -37.862 0.050
  METAS -37.860 0.061
  KRISS -37.83 0.08
  NMIJ -37.83 0.05
  GUM -37.80 0.04
  UME -37.81 0.05
  NMIA -37.82 0.08
  CENAM -37.803 0.07
  NRC -37.80 0.04
  VNIIFTRI -37.83 0.06
  NMISA -37.829 0.050
  INMETRO -37.763 0.045

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 63 Hz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.129 0.035
  HBK-DPLA -38.125 0.035
  METAS -38.124 0.049
  KRISS -38.13 0.05
  NMIJ -38.12 0.04
  GUM -38.11 0.03
  UME -38.09 0.03
  NMIA -38.11 0.06
  CENAM -38.096 0.05
  NRC -38.10 0.03
  VNIIFTRI -38.13 0.06
  NMISA -38.120 0.030
  INMETRO -38.12 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.860 0.030
  HBK-DPLA -37.875 0.035
  METAS -37.871 0.055
  KRISS -37.85 0.05
  NMIJ -37.85 0.04
  GUM -37.83 0.03
  UME -37.84 0.03
  NMIA -37.84 0.06
  CENAM -37.827 0.05
  NRC -37.83 0.03
  VNIIFTRI -37.85 0.06
  NMISA -37.855 0.030
  INMETRO -37.785 0.045

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 125 Hz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.145 0.035
  HBK-DPLA -38.134 0.030
  METAS -38.133 0.045
  KRISS -38.14 0.05
  NMIJ -38.14 0.04
  GUM -38.13 0.03
  UME -38.11 0.03
  NMIA -38.13 0.04
  CENAM -38.114 0.05
  NRC -38.12 0.03
  VNIIFTRI -38.15 0.06
  NMISA -38.138 0.030
  INMETRO -38.137 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.877 0.030
  HBK-DPLA -37.883 0.030
  METAS -37.880 0.052
  KRISS -37.86 0.05
  NMIJ -37.88 0.04
  GUM -37.85 0.03
  UME -37.85 0.03
  NMIA -37.86 0.04
  CENAM -37.845 0.05
  NRC -37.85 0.03
  VNIIFTRI -37.87 0.06
  NMISA -37.872 0.030
  INMETRO -37.798 0.045

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 250 Hz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.159 0.035
  HBK-DPLA -38.141 0.025
  METAS -38.138 0.043
  KRISS -38.14 0.04
  NMIJ -38.15 0.04
  GUM -38.14 0.03
  UME -38.12 0.03
  NMIA -38.14 0.04
  CENAM -38.130 0.05
  NRC -38.14 0.03
  VNIIFTRI -38.16 0.06
  NMISA -38.151 0.030
  INMETRO -38.147 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.889 0.030
  HBK-DPLA -37.888 0.025
  METAS -37.886 0.051
  KRISS -37.86 0.04
  NMIJ -37.89 0.04
  GUM -37.86 0.03
  UME -37.86 0.03
  NMIA -37.87 0.04
  CENAM -37.860 0.05
  NRC -37.87 0.03
  VNIIFTRI -37.88 0.06
  NMISA -37.884 0.030
  INMETRO -37.808 0.045

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 500 Hz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.167 0.035
  HBK-DPLA -38.144 0.020
  METAS -38.143 0.042
  KRISS -38.14 0.04
  NMIJ -38.16 0.04
  GUM -38.15 0.03
  UME -38.13 0.03
  NMIA -38.15 0.04
  CENAM -38.138 0.05
  NRC -38.14 0.03
  VNIIFTRI -38.17 0.06
  NMISA -38.159 0.030
  INMETRO -38.155 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.896 0.030
  HBK-DPLA -37.890 0.020
  METAS -37.888 0.050
  KRISS -37.86 0.04
  NMIJ -37.89 0.04
  GUM -37.86 0.03
  UME -37.87 0.03
  NMIA -37.88 0.04
  CENAM -37.868 0.05
  NRC -37.88 0.03
  VNIIFTRI -37.89 0.06
  NMISA -37.892 0.030
  INMETRO -37.815 0.045

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 1 kHz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.167 0.035
  HBK-DPLA -38.140 0.015
  METAS -38.141 0.041
  KRISS -38.14 0.04
  NMIJ -38.16 0.04
  GUM -38.15 0.03
  UME -38.13 0.03
  NMIA -38.15 0.04
  CENAM -38.138 0.05
  NRC -38.14 0.03
  VNIIFTRI -38.17 0.06
  NMISA -38.160 0.030
  INMETRO -38.155 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.897 0.030
  HBK-DPLA -37.886 0.015
  METAS -37.887 0.049
  KRISS -37.86 0.04
  NMIJ -37.89 0.04
  GUM -37.87 0.03
  UME -37.87 0.03
  NMIA -37.88 0.04
  CENAM -37.869 0.05
  NRC -37.88 0.03
  VNIIFTRI -37.89 0.06
  NMISA -37.893 0.030
  INMETRO -37.817 0.045

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 2 kHz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.144 0.035
  HBK-DPLA -38.115 0.015
  METAS -38.115 0.040
  KRISS -38.11 0.04
  NMIJ -38.14 0.04
  GUM -38.12 0.03
  UME -38.11 0.03
  NMIA -38.13 0.04
  CENAM -38.115 0.05
  NRC -38.12 0.03
  VNIIFTRI -38.15 0.06
  NMISA -38.136 0.030
  INMETRO -38.135 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.878 0.030
  HBK-DPLA -37.864 0.015
  METAS -37.865 0.049
  KRISS -37.83 0.04
  NMIJ -37.87 0.04
  GUM -37.85 0.03
  UME -37.85 0.03
  NMIA -37.86 0.04
  CENAM -37.849 0.05
  NRC -37.86 0.03
  VNIIFTRI -37.87 0.06
  NMISA -37.873 0.030
  INMETRO -37.801 0.045

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 4 kHz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -38.042 0.035
  HBK-DPLA -38.008 0.015
  METAS -38.010 0.039
  KRISS -38.00 0.04
  NMIJ -38.03 0.04
  GUM -38.02 0.03
  UME -38.00 0.04
  NMIA -38.02 0.04
  CENAM -38.012 0.06
  NRC -38.01 0.03
  VNIIFTRI -38.05 0.06
  NMISA -38.031 0.030
  INMETRO -38.04 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.789 0.030
  HBK-DPLA -37.773 0.015
  METAS -37.773 0.047
  KRISS -37.74 0.04
  NMIJ -37.78 0.04
  GUM -37.76 0.03
  UME -37.76 0.04
  NMIA -37.77 0.04
  CENAM -37.762 0.06
  NRC -37.77 0.03
  VNIIFTRI -37.79 0.06
  NMISA -37.784 0.030
  INMETRO -37.723 0.045

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 8 kHz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.688 0.040
  HBK-DPLA -37.644 0.015
  METAS -37.645 0.041
  KRISS -37.64 0.04
  NMIJ -37.67 0.04
  GUM -37.67 0.05
  UME -37.65 0.05
  NMIA -37.65 0.04
  CENAM -37.650 0.06
  NRC -37.64 0.03
  VNIIFTRI -37.68 0.06
  NMISA -37.662 0.040
  INMETRO -37.71 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.497 0.030
  HBK-DPLA -37.475 0.015
  METAS -37.477 0.048
  KRISS -37.45 0.04
  NMIJ -37.50 0.04
  GUM -37.49 0.05
  UME -37.47 0.05
  NMIA -37.47 0.04
  CENAM -37.476 0.06
  NRC -37.46 0.03
  VNIIFTRI -37.50 0.06
  NMISA -37.487 0.040
  INMETRO -37.475 0.045

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 12.5 kHz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.365 0.045
  HBK-DPLA -37.312 0.015
  METAS -37.308 0.060
  KRISS -37.31 0.06
  NMIJ -37.34 0.04
  GUM -37.35 0.08
  UME -37.32 0.06
  NMIA -37.30 0.06
  CENAM -37.316 0.10
  NRC -37.29 0.04
  VNIIFTRI -37.35 0.09
  NMISA -37.318 0.040
  INMETRO -37.432 0.045

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.308 0.030
  HBK-DPLA -37.288 0.015
  METAS -37.285 0.067
  KRISS -37.27 0.06
  NMIJ -37.33 0.04
  GUM -37.33 0.08
  UME -37.29 0.06
  NMIA -37.28 0.06
  CENAM -37.308 0.10
  NRC -37.27 0.04
  VNIIFTRI -37.33 0.09
  NMISA -37.300 0.040
  INMETRO -37.379 0.045

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 16 kHz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.666 0.055
  HBK-DPLA -37.595 0.025
  METAS -37.586 0.090
  KRISS -37.61 0.07
  NMIJ -37.63 0.05
  GUM -37.66 0.08
  UME -37.62 0.08
  NMIA -37.57 0.08
  CENAM -37.601 0.13
  NRC -37.57 0.06
  VNIIFTRI -37.64 0.13
  NMISA -37.603 0.060
  INMETRO -37.735 0.050

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -37.696 0.035
  HBK-DPLA -37.679 0.025
  METAS -37.668 0.095
  KRISS -37.70 0.07
  NMIJ -37.74 0.05
  GUM -37.75 0.08
  UME -37.70 0.08
  NMIA -37.67 0.08
  CENAM -37.712 0.13
  NRC -37.67 0.06
  VNIIFTRI -37.75 0.13
  NMISA -37.693 0.060
  INMETRO -37.827 0.050

 

CCAUV.A-K6

 

MEASURAND   Pressure sensitivity level

NOMINAL FREQUENCY 25 kHz

 

xi  measured sensitivity reported by laboratory iLabi

Ui  expanded uncertainty (k = 2) of xi 

Note: The values in bold have been excluded from the calculation of the Key Comparison Reference Values

 

  Microphone 4180.2124385  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -42.013 0.145
  HBK-DPLA -41.934 0.070
  METAS -41.916 0.414
  KRISS -41.78 0.25
  NMIJ -41.96 0.21
  GUM -42.08 0.24
  UME -41.90 0.20
  NMIA -41.90 0.15
  CENAM -41.938 0.40
  NRC -41.94 0.37
  VNIIFTRI -41.90 0.50
  NMISA -41.929 0.150
  INMETRO -41.953 0.160

 

  Microphone 4180.2412872  
  Labi xi Ui
  dB re 1 V/Pa dB
  LNE -42.066 0.130
  HBK-DPLA -42.069 0.070
  METAS -41.950 0.419
  KRISS -41.91 0.25
  NMIJ -42.02 0.21
  GUM -42.19 0.24
  UME -42.05 0.20
  NMIA -42.05 0.15
  CENAM -42.017 0.40
  NRC -42.03 0.37
  VNIIFTRI -42.00 0.50
  NMISA -42.056 0.150
  INMETRO -42.168 0.160