CCM.P-K4.2012
MEASURAND : Pressure
NOMINAL PRESSURE : 1 Pa, 3 Pa, 10 Pa, 30 Pa,100 Pa, 300 Pa, 1 000 Pa,
3 000 Pa, 7 000 Pa and 10 000 Pa
MEASURAND : Pressure
Degrees of equivalence Di and expanded uncertainty Ui (95% level of confidence) expressed in Pa
CCM.P-K4.2012
MEASURAND : Pressure
NOMINAL PRESSURE : 1 Pa, 3 Pa, 10 Pa, 30 Pa,100 Pa, 300 Pa, 1 000 Pa,
3 000 Pa, 7 000 Pa and 10 000 Pa
Results are presented under A4 printable format in Summary Results (PDF file).
Metrology area, Sub-field | Mass and related quantities, Pressure |
Description | Pressure measurements in gas (absolute mode) |
Time of measurements | 2012 - 2013 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
References | |
Measurand | Pressure: 1 Pa to 10 kPa |
Transfer device | Transfer standard packages |
Comparison type | Key Comparison |
Consultative Committee | CCM (Consultative Committee for Mass and Related Quantities) |
Conducted by | CCM (Consultative Committee for Mass and Related Quantities) |
Comments | Results published on 10 July 2019 Pressure measurements in gas (absolute mode)
|
Pilot institute |
NIST
National Institute of Standards and Technology United States |
Contact person | Jay HENDRICKS +1 301 975 4836 |
Pilot laboratory | |
---|---|
NIST |
National Institute of Standards and Technology, United States, SIM |
CENAM |
Centro Nacional de Metrologia, Mexico, SIM |
CMI |
Czech Metrology Institute, Czechia, EURAMET |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
NMISA |
National Metrology Institute of South Africa, South Africa, AFRIMETS |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
VNIIM |
D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCM.P-K4.2012
MEASURAND : Pressure
NOMINAL PRESSURE : 1 Pa, 3 Pa, 10 Pa, 30 Pa,100 Pa, 300 Pa, 1 000 Pa,
3 000 Pa, 7 000 Pa and 10 000 Pa
For each nominal pressure value, the key comparison reference value, pR, is obtained from an unweighted mean of the measurement method means, excluding CENAM for the determination of pR at 10 Paz and 30 Pa.
The combined standard uncertainty of pR is uR .
The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: Di = (pi - pR) and Ui, its expanded uncertainty at a (k = 2), both expressed in Pa. |
The degree of equivalence between two laboratories is given at p. 29 of the Final Report. |
CCM.P-K4.2012
MEASURAND : Pressure
NOMINAL PRESSURE : 1 Pa, 3 Pa, 10 Pa, 30 Pa,100 Pa, 300 Pa, 1 000 Pa,
3 000 Pa, 7 000 Pa and 10 000 Pa
pi : corrected mean gauge reading obtained by laboratory i
ui : combined standard uncertainty of pi
(see CCM.P-K4.2012 Final Report p.15)