CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 10 MHz

 

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

Degrees of equivalence represented by Di and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

Calibration factor at 10 MHz
Di / dimensionless
Participating institutes

* TM3 was not measured

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 4 GHz

 

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

Degrees of equivalence represented by Di and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

Calibration factor at 4 GHz
Di / dimensionless
Participating institutes

* TM3 was not measured

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 18 GHz

 

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

Degrees of equivalence represented by Di and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

Calibration factor at 4 GHz
Di / dimensionless
Participating institutes

* TM3 was not measured

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 10 MHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

Graph of equivalence : Di and expanded uncertainty Ui, both dimensionless
Click on the graph for a closer view

Red diamonds : CCEM.RF-K8.CL participants
Blue squares : EUROMET.EM.RF-K8.CL participants

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 4 GHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

Graph of equivalence : Di and expanded uncertainty Ui, both dimensionless
Click on the graph for a closer view

Red diamonds : CCEM.RF-K8.CL participants
Blue squares : EUROMET.EM.RF-K8.CL participants

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 18 GHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

Graph of equivalence : Di and expanded uncertainty Ui, both dimensionless
Click on the graph for a closer view

Red diamonds : CCEM.RF-K8.CL participants
Blue squares : EUROMET.EM.RF-K8.CL participants

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 10 MHz

 

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

Degrees of equivalence represented by Di and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
INRIM 0.0070 0.0255
INTA 0.0027 0.0159
METAS * -0.0035 0.0106
CMI -0.0021 0.0108
BFKH 0.0048 0.0110
PTB * 0.0043 0.0047
NPL 0.0035 0.0109
VSL -0.0057 0.0091
NMISA -0.0015 0.0173
UME -0.0058 0.0186
MIRS/SIQ/Metrology 0.0050 0.0095
LNE * 0.0000 0.0083
NIST -0.0105 0.0084
NMIA 0.0040 0.0069
KRISS * 0.0083 0.0049
NMIJ * -0.0049 0.0104
NRC -0.0365 0.0104
NIM -0.0021 0.0112
NMC, A*STAR * 0.0029 0.0145
NPL 0.0002 0.0206
INRIM 0.0236 0.0207
METAS 0.0064 0.0236
VSL -0.0019 0.0225
NRC -0.0258 0.0208
SMU 0.0053 0.0268
MIKES -0.0001 0.0125
NMIJ 0.0012 0.0059
SCL 0.0004 0.0108
NMIA 0.0024 0.0055
NMIM -0.0233 0.0232
NMISA 0.0014 0.0146
NMC, A*STAR 0.0029 0.0103
NPLI 0.0021 0.0084
KRISS 0.0071 0.0052
NIM -0.0017 0.0079
NIMT -0.0016 0.0124
Participating institutes
CCEM.RF-K8.CL
EUROMET.EM.RF-K8.CL
EUROMET.EM.RF-K8.1.CL
APMP.EM.RF-K8.CL

* TM3 was not measured

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 4 GHz

 

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

Degrees of equivalence represented by Di and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
INRIM 0.0016 0.0059
INTA 0.0010 0.0091
METAS * -0.0094 0.0122
CMI -0.0028 0.0049
BFKH 0.0046 0.0112
PTB * -0.0036 0.0037
NPL 0.0036 0.0176
VSL 0.0003 0.0083
NMISA 0.0071 0.0196
UME 0.0016 0.0081
MIRS/SIQ/Metrology 0.0086 0.0107
LNE * -0.0021 0.0059
NIST -0.0004 0.0031
NMIA -0.0004 0.0042
KRISS * 0.0013 0.0047
NMIJ * -0.0028 0.0101
NRC -0.0062 0.0046
NIM 0.0016 0.0064
NMC, A*STAR * 0.0052 0.0073
NPL 0.0023 0.0029
INRIM -0.0013 0.0162
METAS 0.0045 0.0141
VSL 0.0024 0.0069
NRC 0.0013 0.0039
SMU 0.0011 0.0151
RISE 0.0021 0.0067
MIKES 0.0021 0.0121
NMIJ 0.0006 0.0046
SCL -0.0016 0.0097
NMIA 0.0035 0.0041
MSL 0.0004 0.006
NMIM 0.0016 0.0264
NMISA -0.0031 0.0175
NMC, A*STAR -0.0017 0.0046
NPLI 0.0038 0.0086
KRISS 0.0004 0.0041
NIM 0.0016 0.0059
NIMT 0.0044 0.013
Participating institutes
CCEM.RF-K8.CL
EUROMET.EM.RF-K8.CL
EUROMET.EM.RF-K8.1.CL
APMP.EM.RF-K8.CL

* TM3 was not measured

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 18 GHz

 

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

Degrees of equivalence represented by Di and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
INRIM 0.0216 0.0382
INTA -0.0081 0.0222
METAS * -0.0423 0.0244
CMI -0.0098 0.0190
BFKH 0.0007 0.0292
PTB * -0.0078 0.0067
NPL -0.0059 0.0159
VSL 0.0003 0.0187
NMISA -0.0004 0.0405
UME -0.0010 0.0276
MIRS/SIQ/Metrology -0.0014 0.0230
LNE * -0.0024 0.0070
NIST 0.0027 0.0088
NMIA 0.0001 0.0065
KRISS * 0.0006 0.0094
NMIJ * -0.0047 0.0107
NRC -0.0015 0.0099
NIM 0.0039 0.0115
NMC, A*STAR * 0.0199 0.0108
NPL -0.0014 0.0121
INRIM 0.0284 0.0414
METAS 0.0009 0.0229
VSL 0.0045 0.0159
NRC -0.0091 0.0127
SMU -0.0031 0.0264
RISE -0.0026 0.0155
MIKES -0.0128 0.0225
NMIJ 0.0013 0.0075
SCL 0.0182 0.024
NMIA 0.0066 0.0082
MSL 0.0038 0.0129
NMIM -0.0214 0.0397
NMISA 0.0031 0.0408
NMC, A*STAR 0.0066 0.0083
NPLI 0.0011 0.0196
KRISS 0.0039 0.0082
NIM -0.0016 0.0098
NIMT 0.0088 0.0218
Participating institutes
CCEM.RF-K8.CL
EUROMET.EM.RF-K8.CL
EUROMET.EM.RF-K8.1.CL
APMP.EM.RF-K8.CL

* TM3 was not measured

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 10 MHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

Degrees of equivalence, Di and expanded uncertainty Ui, both dimensionless

- : no measurements at this frequency
N/A : not applicable (no measurements at all on the specified device)
In blue : participants in EUROMET.EM.RF-K8.CL

Results are presented under A4 printable format in Summary Results (.PDF file).

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 4 GHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

Degrees of equivalence, Di and expanded uncertainty Ui, both dimensionless

N/A : not applicable (no measurements at all on the specified device)
In blue : participants in EUROMET.EM.RF-K8.CL

Results are presented under A4 printable format in Summary Results (.PDF file).

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 18 GHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

Degrees of equivalence, Di and expanded uncertainty Ui, both dimensionless

- : no measurements at this frequency
N/A : not applicable (no measurements at all on the specified device)
In blue : participants in EUROMET.EM.RF-K8.CL

Results are presented under A4 printable format in Summary Results (.PDF file).

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

APMP.EM.RF-K8.CL
Metrology area, Sub-field Electricity and Magnetism, Radio frequencies
Description Power in 50 ohm coaxial lines
Time of measurements 2012 - 2013
Status Approved for equivalence
Final Reports of the comparisons
References

APMP.EM.RF-K8.CL Technical Protocol

Measurand Calibration factor
Parameters Power in 50 Ω coaxial lines
Frequency: 10 MHz, 50 MHz, 1 GHz, 4 GHz, 8 GHz, 12 GHz, 15 GHz, and 18 GHz
Transfer device Two thermocouple power sensors and a power meter
Comparison type Key Comparison
Consultative Committee CCEM (Consultative Committee for Electricity and Magnetism)
Conducted by APMP (Asia Pacific Metrology Programme)
Comments Results published 16 October 2026
Pilot institute NMIJ AIST
National Metrology Institute of Japan
Japan
Contact person Kazuhiro Shimaoka

+81 29 861 6828
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

MSL

Measurement Standards Laboratory, New Zealand, APMP

NIM

National Institute of Metrology, China, APMP

NIMT

National Institute of Metrology (Thailand), Thailand, APMP

NMC, A*STAR

National Metrology Centre, Agency for Science, Technology and Research, Singapore, APMP

NMIA

National Measurement Institute, Australia, Australia, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

NML-SIRIM (became NMIM)

National Metrology Laboratory, SIRIM Berhad (became National Metrology Institute of Malaysia), Malaysia, APMP

NPLI

CSIR National Physical Laboratory of India, India, APMP

SCL

Standards and Calibration Laboratory, Hong Kong, China, APMP

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 10 MHz

 

CCEM.RF-K8.CL

TRAVELLING STANDARDS  : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector

For each travelling standard, the key comparison reference value, xR, is obtained as the unweighted average of the results of a number of selected laboratories (see section 7.2 on page 17 of the CCEM.RF-K8.CL Final Report). The standard uncertainty, uR, of the key comparison reference value is the standard deviation of the mean. xR and uR are dimensionless quantities.

For TM1 at 10 MHz, xR = 0.9681 and uR = 0.0018
For TM3 at 10 MHz, xR = 0.9640 and uR = 0.0030

For each travelling standard of type-N connector (TM1 and TM3), the degree of equivalence of laboratory i with respect to the reference value is given by a pair of terms, both dimensionless:
DTMn i = the difference between laboratory i result and xR, and UTMn i, its expanded uncertainty taking into account uR (n = 1 or 3).
These are combined to obtain the degree of equivalence of laboratory i for type-N connector, using the following expressions:
Di = avg(DTM1 i, DTM3 i) and
Ui = [[avg(UTM1 i, UTM3 i)]2 + [abs(DTM1 i - DTM3 i)/2]2]1/2.

The degree of equivalence between two laboratories i and j is given by a pair of terms, both dimensionless:
Dij = (Di - Dj) and Uij, its expanded uncertainty (k = 2), Uij
= (Ui2 + Uj2)1/2.
The pair-wise degrees of equivalence are not explicitely calculated.

 

Linking EUROMET.EM.RF-K8.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

The laboratories that are participants in both comparisons and which have an independent realisation of the quantity power (NPL, INRIM, NRC, and NMi-VSL) are used for linking the results obtained in EUROMET.EM.RF-K8.CL to the results obtained in CCEM.RF-K8.CL. The linking process is explained in section 5 of the Linkage Report.
 

Linking EUROMET.EM.RF-K8.1.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector

The NMi-VSL is the pilot laboratory of the bilateral comparison EUROMET.EM.RF-K8.1.CL by which the results of MIKES are linked to EUROMET.EM.RF-K8.CL and thereby CCEM.RF-K8.CL (see EUROMET.EM.RF-K8.1.CL Final report).
 

Linking APMP.EM.RF-K8.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

NMIJ, NMIA, NMC, KRISS and NIM provide the link between the key comparison CCEM.RF-K8.CL and APMP.EM.RF-K8.CL, having participated in both comparisons. The linking process is described in Section 9 of the APMP.EM.RF-K8.CL Final report.

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 4 GHz

 

CCEM.RF-K8.CL

TRAVELLING STANDARDS  : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector

For each travelling standard, the key comparison reference value, xR, is obtained as the unweighted average of the results of a number of selected laboratories (see section 7.2 on page 17 of the CCEM.RF-K8.CL Final Report). The standard uncertainty, uR, of the key comparison reference value is the standard deviation of the mean. xR and uR are dimensionless quantities.

For TM1 at 4 GHz, xR = 0.9818 and uR = 0.0010
For TM3 at 4 GHz, xR = 0.9790 and uR = 0.0004

For each travelling standard of type-N connector (TM1 and TM3), the degree of equivalence of laboratory i with respect to the reference value is given by a pair of terms, both dimensionless:
DTMn i = the difference between laboratory i result and xR, and UTMn i, its expanded uncertainty taking into account uR (n = 1 or 3).
These are combined to obtain the degree of equivalence of laboratory i for type-N connector, using the following expressions:
Di = avg(DTM1 i, DTM3 i) and
Ui = [[avg(UTM1 i, UTM3 i)]2 + [abs(DTM1 i - DTM3 i)/2]2]1/2.

The degree of equivalence between two laboratories i and j is given by a pair of terms, both dimensionless:
Dij = (Di - Dj) and Uij, its expanded uncertainty (k = 2), Uij
= (Ui2 + Uj2)1/2.
The pair-wise degrees of equivalence are not explicitely calculated.

 

Linking EUROMET.EM.RF-K8.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

The laboratories that are participants in both comparisons and which have an independent realisation of the quantity power (NPL, INRIM, NRC, and NMi-VSL) are used for linking the results obtained in EUROMET.EM.RF-K8.CL to the results obtained in CCEM.RF-K8.CL. The linking process is explained in section 5 of the Linkage Report.
 

Linking EUROMET.EM.RF-K8.1.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector

The NMi-VSL is the pilot laboratory of the bilateral comparison EUROMET.EM.RF-K8.1.CL by which the results of MIKES are linked to EUROMET.EM.RF-K8.CL and thereby CCEM.RF-K8.CL (see EUROMET.EM.RF-K8.1.CL Final report).
 

Linking APMP.EM.RF-K8.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

NMIJ, NMIA, NMC, KRISS and NIM provide the link between the key comparison CCEM.RF-K8.CL and APMP.EM.RF-K8.CL, having participated in both comparisons. The linking process is described in Section 9 of the APMP.EM.RF-K8.CL Final report.

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 18 GHz

 

CCEM.RF-K8.CL

TRAVELLING STANDARDS  : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector

For each travelling standard, the key comparison reference value, xR, is obtained as the unweighted average of the results of a number of selected laboratories (see section 7.2 on page 17 of the CCEM.RF-K8.CL Final Report). The standard uncertainty, uR, of the key comparison reference value is the standard deviation of the mean. xR and uR are dimensionless quantities.

For TM1 at 18 GHz, xR = 0.9391 and uR = 0.0021
For TM3 at 18 GHz, xR = 0.9307 and uR = 0.0016

For each travelling standard of type-N connector (TM1 and TM3), the degree of equivalence of laboratory i with respect to the reference value is given by a pair of terms, both dimensionless:
DTMn i = the difference between laboratory i result and xR, and UTMn i, its expanded uncertainty taking into account uR (n = 1 or 3).
These are combined to obtain the degree of equivalence of laboratory i for type-N connector, using the following expressions:
Di = avg(DTM1 i, DTM3 i) and
Ui = [[avg(UTM1 i, UTM3 i)]2 + [abs(DTM1 i - DTM3 i)/2]2]1/2.

The degree of equivalence between two laboratories i and j is given by a pair of terms, both dimensionless:
Dij = (Di - Dj) and Uij, its expanded uncertainty (k = 2), Uij
= (Ui2 + Uj2)1/2.
The pair-wise degrees of equivalence are not explicitely calculated.

 

Linking EUROMET.EM.RF-K8.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

The laboratories that are participants in both comparisons and which have an independent realisation of the quantity power (NPL, INRIM, NRC, and NMi-VSL) are used for linking the results obtained in EUROMET.EM.RF-K8.CL to the results obtained in CCEM.RF-K8.CL. The linking process is explained in section 5 of the Linkage Report.
 

Linking EUROMET.EM.RF-K8.1.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector

The NMi-VSL is the pilot laboratory of the bilateral comparison EUROMET.EM.RF-K8.1.CL by which the results of MIKES are linked to EUROMET.EM.RF-K8.CL and thereby CCEM.RF-K8.CL (see EUROMET.EM.RF-K8.1.CL Final report).
 

Linking APMP.EM.RF-K8.CL to CCEM.RF-K8.CL

TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

NMIJ, NMIA, NMC, KRISS and NIM provide the link between the key comparison CCEM.RF-K8.CL and APMP.EM.RF-K8.CL, having participated in both comparisons. The linking process is described in Section 9 of the APMP.EM.RF-K8.CL Final report.

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00
FREQUENCY : 10 MHz
TRAVELLING STANDARDS : thermistor mounts with PC7-connector

  • Key comparison CCEM.RF-K8.CL

For each travelling standard, the key comparison reference value, xR, is obtained as the unweighted average of the results of a number of selected laboratories (see section 7.2 on page 17 of the CCEM.RF-K8.CL Final Report). The standard uncertainty, uR, of the key comparison reference value is the standard deviation of the mean. xR and uR are dimensionless quantities.

For TM2 at 10 MHz, xR = 0.9645 and uR = 0.0023

The degree of equivalence of laboratory i with respect to the reference value is given by a pair of terms, both dimensionless:
Di = the difference between laboratory i result and xR, and Ui, its expanded uncertainty taking into account uR.
The degree of equivalence between two laboratories i and j is given by a pair of terms, both dimensionless:
Dij = (Di - Dj) and Uij, its expanded uncertainty (k = 2), Uij = (Ui2 + Uj2)1/2.
The pair-wise degrees of equivalence are not explicitely calculated.
  • Linking EUROMET.EM.RF-K8.CL key comparison to CCEM.RF-K8.CL

The laboratories that are participants in both comparisons and which have an independent realisation of the quantity power (NMi-VSL and NRC) are used for linking the results obtained in EUROMET.EM.RF-K8.CL to the results obtained in CCEM.RF-K8.CL. The linking process is explained in section 6 of the Linkage Report.

It follows that the table of degrees of equivalence and the graph of equivalence obtained for CCEM.RF-K8.CL are extended with results obtained in EUROMET.EM.RF-K8.CL.

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00
FREQUENCY : 4 GHz
TRAVELLING STANDARDS : thermistor mounts with PC7-connector

  • Key comparison CCEM.RF-K8.CL

For each travelling standard, the key comparison reference value, xR, is obtained as the unweighted average of the results of a number of selected laboratories (see section 7.2 on page 17 of the CCEM.RF-K8.CL Final Report). The standard uncertainty, uR, of the key comparison reference value is the standard deviation of the mean. xR and uR are dimensionless quantities.

For TM2 at 4 GHz, xR = 0.9829 and uR = 0.0011

The degree of equivalence of laboratory i with respect to the reference value is given by a pair of terms, both dimensionless:
Di = the difference between laboratory i result and xR, and Ui, its expanded uncertainty taking into account uR.
The degree of equivalence between two laboratories i and j is given by a pair of terms, both dimensionless:
Dij = (Di - Dj) and Uij, its expanded uncertainty (k = 2), Uij = (Ui2 + Uj2)1/2.
The pair-wise degrees of equivalence are not explicitely calculated.
  • Linking EUROMET.EM.RF-K8.CL key comparison to CCEM.RF-K8.CL

The laboratories that are participants in both comparisons and which have an independent realisation of the quantity power (NMi-VSL and NRC) are used for linking the results obtained in EUROMET.EM.RF-K8.CL to the results obtained in CCEM.RF-K8.CL. The linking process is explained in section 6 of the Linkage Report.

It follows that the table of degrees of equivalence and the graph of equivalence obtained for CCEM.RF-K8.CL are extended with results obtained in EUROMET.EM.RF-K8.CL.

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00
FREQUENCY : 18 GHz
TRAVELLING STANDARDS : thermistor mounts with PC7-connector

  • Key comparison CCEM.RF-K8.CL

For each travelling standard, the key comparison reference value, xR, is obtained as the unweighted average of the results of a number of selected laboratories (see section 7.2 on page 17 of the CCEM.RF-K8.CL Final Report). The standard uncertainty, uR, of the key comparison reference value is the standard deviation of the mean. xR and uR are dimensionless quantities.

For TM2 at 18 GHz, xR = 0.9434 and uR = 0.0017

The degree of equivalence of laboratory i with respect to the reference value is given by a pair of terms, both dimensionless:
Di = the difference between laboratory i result and xR, and Ui, its expanded uncertainty taking into account uR.
The degree of equivalence between two laboratories i and j is given by a pair of terms, both dimensionless:
Dij = (Di - Dj) and Uij, its expanded uncertainty (k = 2), Uij = (Ui2 + Uj2)1/2.
The pair-wise degrees of equivalence are not explicitely calculated.
  • Linking EUROMET.EM.RF-K8.CL key comparison to CCEM.RF-K8.CL

The laboratories that are participants in both comparisons and which have an independent realisation of the quantity power (NMi-VSL and NRC) are used for linking the results obtained in EUROMET.EM.RF-K8.CL to the results obtained in CCEM.RF-K8.CL. The linking process is explained in section 6 of the Linkage Report.

It follows that the table of degrees of equivalence and the graph of equivalence obtained for CCEM.RF-K8.CL are extended with results obtained in EUROMET.EM.RF-K8.CL.

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 10 MHz

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

The actual calibration factors of the devices under test are not relevant for the quality of the measurement results, which are given here as the difference between the laboratory result and the relevant key comparison reference value.

Please see Equivalence statements.

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 4 GHz

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

The actual calibration factors of the devices under test are not relevant for the quality of the measurement results, which are given here as the difference between the laboratory result and the relevant key comparison reference value.

Please see Equivalence statements.

CCEM.RF-K8.CL, EUROMET.EM.RF-K8.CL, EUROMET.EM.RF-K8.1.CL and APMP.EM.RF-K8.CL

 

MEASURAND   Calibration factor in coaxial 7 mm transmission line

NOMINAL VALUE 1.00

FREQUENCY 18 GHz

TRAVELLING STANDARDS FOR CCEM.RF-K8.CL : Thermistor mounts, TM1 and TM3 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.CL : Thermistor mounts, TM4 and TM6 both having a male type-N 50 ohm connector

TRAVELLING STANDARDS FOR EUROMET.EM.RF-K8.1.CL : Thermistor mounts, DUT1 and DUT2 both having a male type-N 50 ohm connector
TRAVELLING STANDARDS FOR APMP.EM.RF-K8.CL : Thermocouple power sensors, Sensor 1 and Sensor 2 both having a male type-N 50 ohm connector

 

The actual calibration factors of the devices under test are not relevant for the quality of the measurement results, which are given here as the difference between the laboratory result and the relevant key comparison reference value.

Please see Equivalence statements.

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 10 MHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

The actual calibration factors of the devices under test are not relevant for the quality of the measurement results, which are given here as the difference between the laboratory result and the relevant key comparison reference value.

Please see Equivalence statements.

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 4 GHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

The actual calibration factors of the devices under test are not relevant for the quality of the measurement results, which are given here as the difference between the laboratory result and the relevant key comparison reference value.

Please see Equivalence statements.

CCEM.RF-K8.CL and EUROMET.EM.RF-K8.CL

MEASURAND : Calibration factor in coaxial 7 mm transmission line
NOMINAL VALUE : 1.00

FREQUENCY : 18 GHz

TRAVELLING STANDARD FOR CCEM.RF-K8.CL : TM2 with a PC7-connector
TRAVELLING STANDARD FOR EUROMET.EM.RF-K8.CL : TM5 with a PC7-connector

The actual calibration factors of the devices under test are not relevant for the quality of the measurement results, which are given here as the difference between the laboratory result and the relevant key comparison reference value.

Please see Equivalence statements.