Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ce-139 and APMP.RI(II)-K2.Ce-139

MEASURAND   SIR equivalent activity of Ce-139

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Ce-139
Di / MBq
Participating institutes

Results given in absolute values.

 

 

BIPM.RI(II)-K1.Ce-139

MEASURAND   SIR equivalent activity of Ce-139

 

Degrees of equivalence represented by Di = (xI - xR) / xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Ce-139
Di / (kBq/MBq)
Participating institutes

Results given in relative values.

 

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ce-139 and APMP.RI(II)-K2.Ce-139

MEASURAND   SIR equivalent activity of Ce-139

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
NMIJ -0.03 0.65
PTB -0.16 0.63
BEV -1.2 2.4
NMISA 1.0 1.4
LNE-LNHB -0.03 0.98
INER 0.1 1.0
KRISS -0.9 1.1
NIM 2.0 1.5
VNIIM 0.29 0.63
Participating institutes
BIPM.RI(II)-K1.Ce-139
APMP.RI(II)-K2.Ce-139

Results given in absolute values.

 

 

BIPM.RI(II)-K1.Ce-139

MEASURAND   SIR equivalent activity of Ce-139

 

Degrees of equivalence represented by Di = (xI - xR) / xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
NMIJ -0.2 4.9
PTB -1.2 4.7
BEV -9 18
NMISA 8 11
LNE-LNHB -0.2 7.4
INER 0.8 7.5
KRISS -6.6 8.3
NIM 15 11
VNIIM 2.2 4.7
Participating institutes
BIPM.RI(II)-K1.Ce-139
APMP.RI(II)-K2.Ce-139

This is the possibility to place a comment.

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K1.Ce-139
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Ce-139
Time of measurements 1976 -
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
Most recent results published on 27 September 2022
 
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

BEV

Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET

BFKH

Government Office of the Capital City Budapest, Hungary, EURAMET

Országos Mérésügyi Hivatal (OMH), now BFKH

CMI

Czech Metrology Institute, Czechia, EURAMET

Czech Metrology Institute - Inspectorate for Ionizing Radiation (CMI-IIR), now represented by CMI

IRA

Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

Korea Atomic Energy (KAE), now represented by KRISS.

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

LNMRI/IRD

Laboratorio Nacional de Metrologia das Radiacoes Ionizantes, IRD, Brazil, SIM

NIST

National Institute of Standards and Technology, United States, SIM

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS


Council for Scientific and Industrial Research, National Metrology Laboratory (CSIR-NML), now represented by NMISA

NPL

National Physical Laboratory, United Kingdom, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ce-139 and APMP.RI(II)-K2.Ce-139

MEASURAND   SIR equivalent activity of Ce-139

 

The key comparison reference value xR for this radionuclide is estimated to be

 

xR = 132.77 MBq with a standard uncertainty uR = 0.14 MBq

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence are both expressed in MBq, where xi is the equivalent activity for laboratory i.

The Degrees of equivalence are also given in relative terms by Di = (xI - xR) /xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

 

Linking APMP.RI(II)-K2.Ce-139 to BIPM.RI(II)-K1.Ce-139

The results of APMP.RI(II)-K2.Ce-139 were linked to BIPM.RI(II)-K1.Ce-139 via SIR.

 

 

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ce-139 and APMP.RI(II)-K2.Ce-139

MEASURAND   SIR equivalent activity of Ce-139

 

xi  measurement value submitted by laboratory iLabi

ui   combined standard uncertainty of xi

 

BIPM.RI(II)-K1.Ce-139

         
  Labi xi u(xi) Date of measurement
  MBq MBq
  NMIJ 132.74 0.35 2004-03-16
  PTB 132.61 0.34 2008-03-14
  BEV 131.6 1.2 2008-12-02
  NMISA 133.81 0.73 2019-03-07
  LNE-LNHB 132.74 0.51 2022-03-16
         

 

 

APMP.RI(II)-K2.Ce-139

         
  Labi xi u(xi) Date of measurement
  MBq MBq
  INER 132.88 0.48 2004
  KRISS 131.89 0.53 2004
  NIM 134.74 0.74 2004
  VNIIM 133.06 0.28 2004