CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter A: 92 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
Graphs of equivalence
Filter A - 380 nm: Di and Ui (k = 2)
Note: UIPQ = 9.762E-03; DUME = 1.466E-02 and UUME = 2.806E-03
Filter A - 400 nm: Di and Ui (k = 2)
Note: UIPQ = 8.374E-03; DUME = 1.515E-02 and UUME = 2.639E-03
Filter A - 500 nm: Di and Ui (k = 2)
Note: DINRIM = 5.285E-03 and UINRIM = 1.207E-03; DUME = 1.552E-02 and UUME = 3.540E-03
Filter A - 600 nm: Di and Ui (k = 2)
Note: DUME = 1.313E-02 and UUME = 2.161E-03
Filter A - 700 nm: Di and Ui (k = 2)
Note: DUME = 8.719E-03 and UUME = 9.100E-04
Filter A - 800 nm: Di and Ui (k = 2)
Filter A - 900 nm: Di and Ui (k = 2)
Filter A - 1000 nm: Di and Ui (k = 2)
Red diamonds : participants in CCPR-K6
Blue squares : participants in EUROMET.PR-K6
Green triangle : participant in EURAMET.PR-K6.1
Violet circle : participant in EURAMET.PR-K6.2
Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter B: 56 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
Graphs of equivalence
Filter B - 380 nm: Di and Ui (k = 2)
Note: DUME = 1.967E-02 and UUME = 4.210E-03
Filter B - 400 nm: Di and Ui (k = 2)
Filter B - 500 nm: Di and Ui (k = 2)
Filter B - 600 nm: Di and Ui (k = 2)
Filter B - 700 nm: Di and Ui (k = 2)
Filter B - 800 nm: Di and Ui (k = 2)
Filter B - 900 nm: Di and Ui (k = 2)
Filter B - 1000 nm: Di and Ui (k = 2)
Red diamonds : participants in CCPR-K6
Blue squares : participants in EUROMET.PR-K6
Green triangle : participant in EURAMET.PR-K6.1
Violet circle : participant in EURAMET.PR-K6.2
Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter C: 10 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
Graphs of equivalence
Filter C - 380 nm: Di and Ui (k = 2)
Note: DUME = 4.691E-03 and UUME = 7.527E-04
Filter C - 400 nm: Di and Ui (k = 2)
Note: UBEV = 2.002E-03
Filter C - 500 nm: Di and Ui (k = 2)
Note: DCMI = -2.259E-02 and UCMI = 1.087E-04
Filter C - 600 nm: Di and Ui (k = 2)
Note: UBEV = 2.402E-03; DSP = 6.872E-01 and USP = 4.260E-04
Filter C - 700 nm: Di and Ui (k = 2)
Note: UBEV = 2.401E-03
Filter C - 800 nm: Di and Ui (k = 2)
Note: UBEV = 2.201E-03
Filter C - 900 nm: Di and Ui (k = 2)
Note: DINRIM = 3.672E-03 and UINRIM = 4.112E-04; DSP = 2.226E-03 and USP = 1.183E-03; DUME = 1.594E-03 and UUME = 2.472E-04
Filter C - 1000 nm: Di and Ui (k = 2)
Note: DBEV = 1.533E-03 and UBEV = 2.200E-03; DINRIM = 2.689E-03 and UINRIM = 1.614E-04; DSP = 1.561E-03 and USP = 9.206E-04
Red diamonds : participants in CCPR-K6
Blue squares : participants in EUROMET.PR-K6
Green triangle : participant in EURAMET.PR-K6.1
Violet circle : participant in EURAMET.PR-K6.2
Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter D: 1 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
Graphs of equivalence
Filter D - 380 nm: Di and Ui (k = 2)
Note: UVNIIOFI = 6.540E-04; UBEV = 1.685E-03
Filter D - 400 nm: Di and Ui (k = 2)
Note: UVNIIOFI = 6.622E-04; UBEV = 1.542E-03
Filter D - 500 nm: Di and Ui (k = 2)
Note: UVNIIOFI = 5.983E-04; UBEV = 1.403E-03; DCMI = -2.016E-03 and UCMI = 9.717E-05
Filter D - 600 nm: Di and Ui (k = 2)
Note: UVNIIOFI = 6.281E-04; UBEV = 2.404E-03
Filter D - 700 nm: Di and Ui (k = 2)
Note: UBEV = 2.217E-03
Filter D - 800 nm: Di and Ui (k = 2)
Note: UVNIIOFI = 6.029E-04; UBEV = 2.002E-03
Filter D - 900 nm: Di and Ui (k = 2)
Note: UNMIJ = 1.178E-03; UVNIIOFI = 6.081E-04; UBEV = 2.000E-03; DINRIM = 9.023E-04 and UINRIM = 1.082E-04; DSP = 4.793E-04 and USP = 4.019E-04
Filter D - 1000 nm: Di and Ui (k = 2)
Note: UNMIJ = 1.030E-03; UVNIIOFI = 6.161E-04; UBEV = 1.801E-03; DINRIM = 6.069E-04 and UINRIM = 6.151E-05
Red diamonds : participants in CCPR-K6
Blue squares : participants in EUROMET.PR-K6
Green triangle : participant in EURAMET.PR-K6.1
Violet circle : participant in EURAMET.PR-K6.2
Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter A: 92 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
The tables below show the degrees of equivalence relative to the key comparison reference values
Blue indicates participants in EUROMET.PR-K6
Green indicates participant in EURAMET.PR-K6.1
Orange indicates participant in EURAMET.PR-K6.1
Results are available in Summary Results (PDF file).
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter B: 56 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
The tables below show the degrees of equivalence relative to the key comparison reference values
Blue indicates participants in EUROMET.PR-K6
Green indicates participant in EURAMET.PR-K6.1
Orange indicates participant in EURAMET.PR-K6.1
Results are available in Summary Results (PDF file).
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter C: 10 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
The tables below show the degrees of equivalence relative to the key comparison reference values
Blue indicates participants in EUROMET.PR-K6
Green indicates participant in EURAMET.PR-K6.1
Orange indicates participant in EURAMET.PR-K6.1
Results are available in Summary Results (PDF file).
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter D: 1 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
The tables below show the degrees of equivalence relative to the key comparison reference values
Blue indicates participants in EUROMET.PR-K6
Green indicates participant in EURAMET.PR-K6.1
Orange indicates participant in EURAMET.PR-K6.1
Results are available in Summary Results (PDF file).
Metrology area, Sub-field | Photometry and Radiometry, Properties of materials |
Description | Spectral regular transmittance (Bilateral comparison) |
Time of measurements | 2008 - 2009 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
Measurand | Spectral regular transmittance Nominal values: 92 %, 56 %, 10 % and 1 % |
Parameters | Wavelength from 380 nm to 700 nm (7 wavelengths) |
Transfer device | Coloured glass filters |
Comparison type | Key Comparison |
Consultative Committee | CCPR (Consultative Committee for Photometry and Radiometry) |
Conducted by | EURAMET (European Association of National Metrology Institutes) |
Comments | This is a bilateral key comparison between BIM and VSL. Results published on 12 May 2017 EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2 results are linked to those of CCPR-K6 Spectral regular transmittance
|
Pilot institute |
BIM
Bulgarian Institute of Metrology Bulgaria |
Contact person | Angela KUNOVA +359 2 974 31 61 |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter A: 92 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
The key comparison reference value, xR, and its standard uncertainty, uR, are deduced from the participants' results compared to the Pilot's results, using a weighted mean with application of a cut-off, outliers being excluded, as explained in Sections 5 and 8 of the Final Report. 380 nm: xR = 4.357E-04, uR = 1.058E-04 |
The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to equations 16 and 17 on page 52 of the Final Report (see also Section 9 on page 62). |
The degrees of equivalence between pairs of laboratories have not been computed for this key comparison. |
The linking process is described in section 6 of the EUROMET.PR-K6 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to the participants in EUROMET.PR-K6 only.
The linking process is described in section 6 of the EURAMET.PR-K6.1 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to DMDM participant in EURAMET.PR-K6.1 only.
The linking is made via the VSL who participated both in the CCPR-K6 and EURAMET.PR-K6.2 comparisons.
Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter B: 56 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
The key comparison reference value, xR, and its standard uncertainty, uR, are deduced from the participants' results compared to the Pilot's results, using a weighted mean with application of a cut-off, outliers being excluded, as explained in Sections 5 and 8 of the Final Report. 380 nm: xR = 1.540E-03, uR = 1.637E-04 |
The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to equations 16 and 17 on page 52 of the Final Report (see also Section 9 on page 62). |
The degrees of equivalence between pairs of laboratories have not been computed for this key comparison. |
The linking process is described in section 6 of the EUROMET.PR-K6 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to the participants in EUROMET.PR-K6 only.
The linking process is described in section 6 of the EURAMET.PR-K6.1 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to DMDM participant in EURAMET.PR-K6.1 only.
The linking is made via the VSL who participated both in the CCPR-K6 and EURAMET.PR-K6.2 comparisons.
Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter C: 10 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
The key comparison reference value, xR, and its standard uncertainty, uR, are deduced from the participants' results compared to the Pilot's results, using a weighted mean with application of a cut-off, outliers being excluded, as explained in Sections 5 and 8 of the Final Report. 380 nm: xR = 1.211E-04, uR = 3.114E-05 |
The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to equations 16 and 17 on page 52 of the Final Report (see also Section 9 on page 62). |
The degrees of equivalence between pairs of laboratories have not been computed for this key comparison. |
The linking process is described in section 6 of the EUROMET.PR-K6 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to the participants in EUROMET.PR-K6 only.
The linking process is described in section 6 of the EURAMET.PR-K6.1 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to DMDM participant in EURAMET.PR-K6.1 only.
The linking is made via the VSL who participated both in the CCPR-K6 and EURAMET.PR-K6.2 comparisons.
Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter D: 1 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm
The key comparison reference value, xR, and its standard uncertainty, uR, are deduced from the participants' results compared to the Pilot's results, using a weighted mean with application of a cut-off, outliers being excluded, as explained in Sections 5 and 8 of the Final Report. 380 nm: xR = -1.599E-06, uR = 3.689E-06 |
The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to equations 16 and 17 on page 52 of the Final Report (see also Section 9 on page 62). |
The degrees of equivalence between pairs of laboratories have not been computed for this key comparison. |
The linking process is described in section 6 of the EUROMET.PR-K6 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to the participants in EUROMET.PR-K6 only.
The linking process is described in section 6 of the EURAMET.PR-K6.1 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to DMDM participant in EURAMET.PR-K6.1 only.
The linking is made via the VSL who participated both in the CCPR-K6 and EURAMET.PR-K6.2 comparisons.
Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter A: 92 % nominal transmittance at 546 nm
The filter characteristics are detailed in Table 3 on page 7 of the Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the Final Report.
The filter standards used are detailed in section 3.1 of the EUROMET.PR-K6 Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the EUROMET.PR-K6 Final Report.
The filter standards used are detailed in section 3.1 of the EURAMET.PR-K6.1 Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of the participants are available in page 17 of the EURAMET.PR-K6.1 Final Report.
The filter standards used are listed on p. 2 in the EURAMET.PR-K6.2 Final Report. Measurements are carried out for five wavelengths: 380 nm, 400 nm, 500 nm, 600 nm and 700 nm. Measurement results are listed on pp. 4 to 10 in the EURAMET.PR-K6.2 Final Report.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter B: 56 % nominal transmittance at 546 nm
The filter characteristics are detailed in Table 3 on page 7 of the Final Report. Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the Final Report.
The filter standards used are detailed in section 3.1 of the EUROMET.PR-K6 Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the EUROMET.PR-K6 Final Report.
The filter standards used are detailed in section 3.1 of the EURAMET.PR-K6.1 Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of the participants are available in page 17 of the EURAMET.PR-K6.1 Final Report.
The filter standards used are listed on p. 2 in the EURAMET.PR-K6.2 Final Report. Measurements are carried out for five wavelengths: 380 nm, 400 nm, 500 nm, 600 nm and 700 nm. Measurement results are listed on pp. 4 to 10 in the EURAMET.PR-K6.2 Final Report.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter C: 10 % nominal transmittance at 546 nm
The filter characteristics are detailed in Table 3 on page 7 of the Final Report. Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the Final Report.
The filter standards used are detailed in section 3.1 of the EUROMET.PR-K6 Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the EUROMET.PR-K6 Final Report.
The filter standards used are detailed in section 3.1 of the EURAMET.PR-K6.1 Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of the participants are available in page 17 of the EURAMET.PR-K6.1 Final Report.
The filter standards used are listed on p. 2 in the EURAMET.PR-K6.2 Final Report. Measurements are carried out for five wavelengths: 380 nm, 400 nm, 500 nm, 600 nm and 700 nm. Measurement results are listed on pp. 4 to 10 in the EURAMET.PR-K6.2 Final Report.
CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2
MEASURAND: Spectral regular transmittance
Filter D: 1 % nominal transmittance at 546 nm
The filter characteristics are detailed in Table 3 on page 7 of the Final Report. Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the Final Report.
The filter standards used are detailed in section 3.1 of the EUROMET.PR-K6 Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the EUROMET.PR-K6 Final Report.
The filter standards used are detailed in section 3.1 of the EURAMET.PR-K6.1 Final Report.
Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of the participants are available in page 17 of the EURAMET.PR-K6.1 Final Report.
The filter standards used are listed on p. 2 in the EURAMET.PR-K6.2 Final Report. Measurements are carried out for five wavelengths: 380 nm, 400 nm, 500 nm, 600 nm and 700 nm. Measurement results are listed on pp. 4 to 10 in the EURAMET.PR-K6.2 Final Report.