CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter A: 92 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

Graphs of equivalence

Filter A - 380 nm: Di and Ui (k = 2)
 

Note: UIPQ = 9.762E-03; DUME = 1.466E-02 and UUME = 2.806E-03

Filter A - 400 nm: Di and Ui (k = 2)

Note: UIPQ = 8.374E-03; DUME = 1.515E-02 and UUME = 2.639E-03

Filter A - 500 nm: Di and Ui (k = 2)

Note: DINRIM = 5.285E-03 and UINRIM = 1.207E-03; DUME = 1.552E-02 and UUME = 3.540E-03

 

Filter A - 600 nm: Di and Ui (k = 2)

Note: DUME = 1.313E-02 and UUME = 2.161E-03

 

Filter A - 700 nm: Di and Ui (k = 2)

Note: DUME = 8.719E-03 and UUME = 9.100E-04

 

Filter A - 800 nm: Di and Ui (k = 2)

 

 

Filter A - 900 nm: Di and Ui (k = 2)

 

Filter A - 1000 nm: Di and Ui (k = 2)

 

Red diamonds : participants in CCPR-K6
Blue squares : participants in EUROMET.PR-K6
Green triangle : participant in EURAMET.PR-K6.1
Violet circle : participant in EURAMET.PR-K6.2

 

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter B: 56 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

Graphs of equivalence

Filter B - 380 nm: Di and Ui (k = 2)

Note: DUME = 1.967E-02 and UUME = 4.210E-03

Filter B - 400 nm: Di and Ui (k = 2)

Filter B - 500 nm: Di and Ui (k = 2)

 

 

Filter B - 600 nm: Di and Ui (k = 2)

 

 

Filter B - 700 nm: Di and Ui (k = 2)

 

Filter B - 800 nm: Di and Ui (k = 2)

 

 

 

Filter B - 900 nm: Di and Ui (k = 2)

 

Filter B - 1000 nm: Di and Ui (k = 2)

 

 

Red diamonds : participants in CCPR-K6
Blue squares : participants in EUROMET.PR-K6
Green triangle : participant in EURAMET.PR-K6.1
Violet circle : participant in EURAMET.PR-K6.2

 

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter C: 10 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

Graphs of equivalence

Filter C - 380 nm: Di and Ui (k = 2)

Note: DUME = 4.691E-03 and UUME = 7.527E-04

Filter C - 400 nm: Di and Ui (k = 2)

Note: UBEV = 2.002E-03

Filter C - 500 nm: Di and Ui (k = 2)

Note: DCMI = -2.259E-02 and UCMI = 1.087E-04

 

Filter C - 600 nm: Di and Ui (k = 2)

Note: UBEV = 2.402E-03; DSP = 6.872E-01 and USP = 4.260E-04

 

Filter C - 700 nm: Di and Ui (k = 2)

Note: UBEV = 2.401E-03

 

Filter C - 800 nm: Di and Ui (k = 2)

 

 

 

Note: UBEV = 2.201E-03

Filter C - 900 nm: Di and Ui (k = 2)

Note: DINRIM = 3.672E-03 and UINRIM = 4.112E-04; DSP = 2.226E-03 and USP = 1.183E-03; DUME = 1.594E-03 and UUME = 2.472E-04

Filter C - 1000 nm: Di and Ui (k = 2)

Note: DBEV = 1.533E-03 and UBEV = 2.200E-03; DINRIM = 2.689E-03 and UINRIM = 1.614E-04; DSP = 1.561E-03 and USP = 9.206E-04

 

 

 

Red diamonds : participants in CCPR-K6
Blue squares : participants in EUROMET.PR-K6
Green triangle : participant in EURAMET.PR-K6.1
Violet circle : participant in EURAMET.PR-K6.2

 

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter D: 1 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

Graphs of equivalence

Filter D - 380 nm: Di and Ui (k = 2)

Note: UVNIIOFI = 6.540E-04; UBEV = 1.685E-03

Filter D - 400 nm: Di and Ui (k = 2)

 

Note: UVNIIOFI = 6.622E-04; UBEV = 1.542E-03

Filter D - 500 nm: Di and Ui (k = 2)

Note: UVNIIOFI = 5.983E-04; UBEV = 1.403E-03; DCMI = -2.016E-03 and UCMI = 9.717E-05

 

Filter D - 600 nm: Di and Ui (k = 2)

Note: UVNIIOFI = 6.281E-04; UBEV = 2.404E-03

 

Filter D - 700 nm: Di and Ui (k = 2)

Note: UBEV = 2.217E-03

 

Filter D - 800 nm: Di and Ui (k = 2)

Note: UVNIIOFI = 6.029E-04; UBEV = 2.002E-03

Filter D - 900 nm: Di and Ui (k = 2)

Note: UNMIJ = 1.178E-03; UVNIIOFI = 6.081E-04; UBEV = 2.000E-03; DINRIM = 9.023E-04 and UINRIM = 1.082E-04; DSP = 4.793E-04 and USP = 4.019E-04

 

Filter D - 1000 nm: Di and Ui (k = 2)

Note: UNMIJ = 1.030E-03; UVNIIOFI = 6.161E-04; UBEV = 1.801E-03; DINRIM = 6.069E-04 and UINRIM = 6.151E-05

Red diamonds : participants in CCPR-K6
Blue squares : participants in EUROMET.PR-K6
Green triangle : participant in EURAMET.PR-K6.1
Violet circle : participant in EURAMET.PR-K6.2

 

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter A: 92 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

The tables below show the degrees of equivalence relative to the key comparison reference values

 

Blue indicates participants in EUROMET.PR-K6
Green indicates participant in EURAMET.PR-K6.1
Orange indicates participant in EURAMET.PR-K6.1

Results are available in Summary Results (PDF file).

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter B: 56 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

The tables below show the degrees of equivalence relative to the key comparison reference values

 

 

Blue indicates participants in EUROMET.PR-K6
Green indicates participant in EURAMET.PR-K6.1
Orange indicates participant in EURAMET.PR-K6.1

Results are available in Summary Results (PDF file).

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter C: 10 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

The tables below show the degrees of equivalence relative to the key comparison reference values

 

Blue indicates participants in EUROMET.PR-K6
Green indicates participant in EURAMET.PR-K6.1
Orange indicates participant in EURAMET.PR-K6.1

Results are available in Summary Results (PDF file).

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter D: 1 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

The tables below show the degrees of equivalence relative to the key comparison reference values

 

Blue indicates participants in EUROMET.PR-K6
Green indicates participant in EURAMET.PR-K6.1
Orange indicates participant in EURAMET.PR-K6.1

Results are available in Summary Results (PDF file).

Comparison
Comparison close
RMO comparison
Linked comparison
EURAMET.PR-K6.2
Metrology area, Sub-field Photometry and Radiometry, Properties of materials
Description Spectral regular transmittance (Bilateral comparison)
Time of measurements 2008 - 2009
Status Approved for equivalence
Final Reports of the comparisons
Measurand Spectral regular transmittance
Nominal values: 92 %, 56 %, 10 % and 1 %

Parameters Wavelength from 380 nm to 700 nm (7 wavelengths)
Transfer device Coloured glass filters
Comparison type Key Comparison
Consultative Committee CCPR (Consultative Committee for Photometry and Radiometry)
Conducted by EURAMET (European Association of National Metrology Institutes)
Comments

This is a bilateral key comparison between BIM and VSL.

Results published on 12 May 2017

EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2 results are linked to those of CCPR-K6

Spectral regular transmittance

Pilot institute BIM
Bulgarian Institute of Metrology
Bulgaria
Contact person Angela KUNOVA

+359 2 974 31 61
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIM

Bulgarian Institute of Metrology, Bulgaria, EURAMET

VSL

VSL, Netherlands, EURAMET

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

  • Key comparison CCPR-K6

MEASURAND: Spectral regular transmittance
Filter A: 92 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

 

The key comparison reference value, xR, and its standard uncertainty, uR, are deduced from the participants' results compared to the Pilot's results, using a weighted mean with application of a cut-off, outliers being excluded, as explained in Sections 5 and 8 of the Final Report.

380 nm: xR = 4.357E-04, uR = 1.058E-04
400 nm: xR = 3.701E-04, uR = 8.210E-05
500 nm: xR = 3.234E-04, uR = 6.338E-05
600 nm: xR = 3.075E-04, uR = 6.302E-05
700 nm: xR = 4.225E-04, uR = 6.926E-05
800 nm: xR = 6.383E-04, uR = 4.380E-05
900 nm: xR = 5.228E-04, uR = 6.047E-05
1000 nm: xR = 4.211E-04, uR = 6.443E-05

 

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to equations 16 and 17 on page 52 of the Final Report (see also Section 9 on page 62).

 

The degrees of equivalence between pairs of laboratories have not been computed for this key comparison.

 

  • Linking key comparison EUROMET.PR-K6 to CCPR-K6

The linking process is described in section 6 of the EUROMET.PR-K6 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to the participants in EUROMET.PR-K6 only.

  • Linking key comparison EURAMET.PR-K6.1 to CCPR-K6

The linking process is described in section 6 of the EURAMET.PR-K6.1 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to DMDM participant in EURAMET.PR-K6.1 only.

  • Linking key comparison EURAMET.PR-K6.2 to CCPR-K6

The linking is made via the VSL who participated both in the CCPR-K6 and EURAMET.PR-K6.2 comparisons.

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

  • Key comparison CCPR-K6

MEASURAND: Spectral regular transmittance
Filter B: 56 % nominal transmittance at 546 nm

Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

 

The key comparison reference value, xR, and its standard uncertainty, uR, are deduced from the participants' results compared to the Pilot's results, using a weighted mean with application of a cut-off, outliers being excluded, as explained in Sections 5 and 8 of the Final Report.

380 nm: xR = 1.540E-03, uR = 1.637E-04
400 nm: xR = 1.093E-03, uR = 1.187E-04
500 nm: xR = 9.756E-04, uR = 1.139E-04
600 nm: xR = 7.611E-04, uR = 6.358E-05
700 nm: xR = 3.424E-04, uR = 5.251E-05
800 nm: xR = 2.010E-04, uR = 5.879E-05
900 nm: xR = 7.263E-05, uR = 5.819E-05
1000 nm: xR = 1.051E-04, uR = 6.384E-05

 

 

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to equations 16 and 17 on page 52 of the Final Report (see also Section 9 on page 62).

 

 

The degrees of equivalence between pairs of laboratories have not been computed for this key comparison.

 

  • Linking key comparison EUROMET.PR-K6 to CCPR-K6

The linking process is described in section 6 of the EUROMET.PR-K6 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to the participants in EUROMET.PR-K6 only.

  • Linking key comparison EURAMET.PR-K6.1 to CCPR-K6

The linking process is described in section 6 of the EURAMET.PR-K6.1 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to DMDM participant in EURAMET.PR-K6.1 only.

  • Linking key comparison EURAMET.PR-K6.2 to CCPR-K6

The linking is made via the VSL who participated both in the CCPR-K6 and EURAMET.PR-K6.2 comparisons.

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

  • Key comparison CCPR-K6

MEASURAND: Spectral regular transmittance
Filter C: 10 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

The key comparison reference value, xR, and its standard uncertainty, uR, are deduced from the participants' results compared to the Pilot's results, using a weighted mean with application of a cut-off, outliers being excluded, as explained in Sections 5 and 8 of the Final Report.

380 nm: xR = 1.211E-04, uR = 3.114E-05
400 nm: xR = -3.117E-05, uR = 2.835E-05
500 nm: xR = 1.156E-04, uR = 2.193E-05
600 nm: xR = 1.777E-04, uR = 1.765E-05
700 nm: xR = 9.633E-05, uR = 2.987E-05
800 nm: xR = 1.455E-05, uR = 2.755E-05
900 nm: xR = 8.838E-06, uR = 1.978E-05
1000 nm: xR = -1.840E-06, uR = 1.555E-05

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to equations 16 and 17 on page 52 of the Final Report (see also Section 9 on page 62).

The degrees of equivalence between pairs of laboratories have not been computed for this key comparison.
  • Linking key comparison EUROMET.PR-K6 to CCPR-K6

The linking process is described in section 6 of the EUROMET.PR-K6 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to the participants in EUROMET.PR-K6 only.

  • Linking key comparison EURAMET.PR-K6.1 to CCPR-K6

The linking process is described in section 6 of the EURAMET.PR-K6.1 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to DMDM participant in EURAMET.PR-K6.1 only.

  • Linking key comparison EURAMET.PR-K6.2 to CCPR-K6

The linking is made via the VSL who participated both in the CCPR-K6 and EURAMET.PR-K6.2 comparisons.

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter D: 1 % nominal transmittance at 546 nm
Wavelength: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm

  • Key comparison CCPR-K6

The key comparison reference value, xR, and its standard uncertainty, uR, are deduced from the participants' results compared to the Pilot's results, using a weighted mean with application of a cut-off, outliers being excluded, as explained in Sections 5 and 8 of the Final Report.

380 nm: xR = -1.599E-06, uR = 3.689E-06
400 nm: xR = -1.185E-05, uR = 4.792E-06
500 nm: xR = 8.555E-06, uR = 4.187E-06
600 nm: xR = 4.101E-05, uR = 4.668E-06
700 nm: xR = 3.374E-05, uR = 1.090E-05
800 nm: xR = 6.546E-06, uR = 8.918E-06
900 nm: xR = -4.104E-06, uR = 6.348E-06
1000 nm: xR = -3.835E-06, uR = 5.411E-06

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to equations 16 and 17 on page 52 of the Final Report (see also Section 9 on page 62).

The degrees of equivalence between pairs of laboratories have not been computed for this key comparison.
  • Linking key comparison EUROMET.PR-K6 to CCPR-K6

The linking process is described in section 6 of the EUROMET.PR-K6 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to the participants in EUROMET.PR-K6 only.

  • Linking key comparison EURAMET.PR-K6.1 to CCPR-K6

The linking process is described in section 6 of the EURAMET.PR-K6.1 Final Report. For each wavelength, the Matrix and the Graph of equivalence obtained in CCPR-K6 are extended to DMDM participant in EURAMET.PR-K6.1 only.

  • Linking key comparison EURAMET.PR-K6.2 to CCPR-K6

The linking is made via the VSL who participated both in the CCPR-K6 and EURAMET.PR-K6.2 comparisons.

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO-31-0 Annex B Rule B.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter A: 92 % nominal transmittance at 546 nm

  • Key comparison CCPR-K6

The filter characteristics are detailed in Table 3 on page 7 of the Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the Final Report.

  • Key comparison EUROMET.PR-K6

The filter standards used are detailed in section 3.1 of the EUROMET.PR-K6 Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the EUROMET.PR-K6 Final Report.

  • Key comparison EURAMET.PR-K6.1

The filter standards used are detailed in section 3.1 of the EURAMET.PR-K6.1 Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of the participants are available in page 17 of the EURAMET.PR-K6.1 Final Report.

  • Key comparison EURAMET.PR-K6.2

The filter standards used are listed on p. 2 in the EURAMET.PR-K6.2 Final Report. Measurements are carried out for five wavelengths: 380 nm, 400 nm, 500 nm, 600 nm and 700 nm. Measurement results are listed on pp. 4 to 10 in the EURAMET.PR-K6.2 Final Report.
 

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter B: 56 % nominal transmittance at 546 nm

  • Key comparison CCPR-K6

The filter characteristics are detailed in Table 3 on page 7 of the Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the Final Report.

  • Key comparison EUROMET.PR-K6

The filter standards used are detailed in section 3.1 of the EUROMET.PR-K6 Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the EUROMET.PR-K6 Final Report.

  • Key comparison EURAMET.PR-K6.1

The filter standards used are detailed in section 3.1 of the EURAMET.PR-K6.1 Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of the participants are available in page 17 of the EURAMET.PR-K6.1 Final Report.

  • Key comparison EURAMET.PR-K6.2

The filter standards used are listed on p. 2 in the EURAMET.PR-K6.2 Final Report. Measurements are carried out for five wavelengths: 380 nm, 400 nm, 500 nm, 600 nm and 700 nm. Measurement results are listed on pp. 4 to 10 in the EURAMET.PR-K6.2 Final Report.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter C: 10 % nominal transmittance at 546 nm

  • Key comparison CCPR-K6

The filter characteristics are detailed in Table 3 on page 7 of the Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the Final Report.

  • Key comparison EUROMET.PR-K6

The filter standards used are detailed in section 3.1 of the EUROMET.PR-K6 Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the EUROMET.PR-K6 Final Report.

  • Key comparison EURAMET.PR-K6.1

The filter standards used are detailed in section 3.1 of the EURAMET.PR-K6.1 Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of the participants are available in page 17 of the EURAMET.PR-K6.1 Final Report.

  • Key comparison EURAMET.PR-K6.2

The filter standards used are listed on p. 2 in the EURAMET.PR-K6.2 Final Report. Measurements are carried out for five wavelengths: 380 nm, 400 nm, 500 nm, 600 nm and 700 nm. Measurement results are listed on pp. 4 to 10 in the EURAMET.PR-K6.2 Final Report.

CCPR-K6, EUROMET.PR-K6, EURAMET.PR-K6.1 and EURAMET.PR-K6.2

MEASURAND: Spectral regular transmittance
Filter D: 1 % nominal transmittance at 546 nm

  • Key comparison CCPR-K6

The filter characteristics are detailed in Table 3 on page 7 of the Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the Final Report.

  • Key comparison EUROMET.PR-K6

The filter standards used are detailed in section 3.1 of the EUROMET.PR-K6 Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of all participants are available in the Annex B of the EUROMET.PR-K6 Final Report.

  • Key comparison EURAMET.PR-K6.1

The filter standards used are detailed in section 3.1 of the EURAMET.PR-K6.1 Final Report.

Measurements are carried out for eight wavelengths: 380 nm, 400 nm, 500 nm, 600 nm, 700 nm, 800 nm, 900 nm, and 1000 nm.
Absolute transmission measurements and associated uncertainties of the participants are available in page 17 of the EURAMET.PR-K6.1 Final Report.

  • Key comparison EURAMET.PR-K6.2

The filter standards used are listed on p. 2 in the EURAMET.PR-K6.2 Final Report. Measurements are carried out for five wavelengths: 380 nm, 400 nm, 500 nm, 600 nm and 700 nm. Measurement results are listed on pp. 4 to 10 in the EURAMET.PR-K6.2 Final Report.