BIPM.RI(I)-K4, SIM.RI(I)-K4, EUROMET.RI(I)-K4, EURAMET.RI(I)-K4.2, EURAMET.RI(I)-K4.1 and APMP.RI(I)-K4
MEASURAND Absorbed dose to water
Radiation quality Co-60 beam
Degrees of equivalence, Di and expanded uncertainty Ui (k = 2)
BIPM.RI(I)-K4, SIM.RI(I)-K4, EUROMET.RI(I)-K4, EURAMET.RI(I)-K4.2, EURAMET.RI(I)-K4.1 and APMP.RI(I)-K4
MEASURAND Absorbed dose to water
Radiation quality Co-60 beam
Degrees of equivalence, Di and expanded uncertainty Ui (k = 2)
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mGy/Gy |
mGy/Gy |
mGy/Gy |
||
MKEH | -0.7 | 9.6 | ||
ENEA | -0.1 | 8.8 | ||
VNIIFTRI | -1.4 | 8.6 | ||
NMIJ | -3.0 | 9.2 | ||
LNE-LNHB | -1.9 | 7.8 | ||
METAS | 1.1 | 10 | ||
PTB | -1.3 | 7.6 | ||
NPL | 2.3 | 14 | ||
VSL | -3.0 | 9.6 | ||
BEV | -0.3 | 8.2 | ||
GUM | 3.0 | 7.0 | ||
ARPANSA | -0.5 | 10 | ||
NRC | -0.5 | 6.8 | ||
CMI | -4.0 | 24 | ||
RMTC | -5.3 | 12 | ||
SSM | -1.4 | 10 | ||
STUK | -3.9 | 8.5 | ||
NRPA | 3.2 | 8.8 | ||
SMU | -4.7 | 25 | ||
IAEA | -0.4 | 10 | ||
HIRCL | 3.0 | 12 | ||
ITN | -7.1 | 13 | ||
NIST | -0.6 | 11 | ||
LNMRI | 1.0 | 15 | ||
CNEA | 12 | 18 | ||
SCK-CEN | -1.5 | 16 | ||
CIEMAT | 2.3 | 11 | ||
VINS | 0.0 | 14 | ||
Nuc. Malaysia | -1.7 | 14 | ||
AECS-NRLM | 0.7 | 24 | ||
KRISS | -3.3 | 8.0 | ||
NMISA | -0.1 | 11 | ||
BRMD, DMSc | 1.8 | 11 | ||
BARC | 3.2 | 11 | ||
NIM | 1.0 | 10 | ||
PTKMR-BATAN | 2.3 | 19 | ||
NRSL/INER | -1.9 | 8.2 | ||
NIS | 0.3 | 14 |
Metrology area, Sub-field | Ionizing Radiation, Section I (x and gamma rays, electrons) |
Description | Measurement of absorbed dose to water for Cobalt 60 |
Time of measurements | 2005 - 2008 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
References | |
Measurand | Absorbed dose rate to water in mGy/nC |
Parameters | Radiation type: Co-60 gamma rays |
Transfer device | Two ionization chambers and electrometers |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | EURAMET (European Association of National Metrology Institutes) |
RMO Internal Identifier | EUROMET Project No 813 |
Comments | Resuts published on 27 April 2010 Last results update: 07 November 2018 SIM.RI(I)-K4, EUROMET.RI(I)-K4 and EURAMET.RI(I)-K4.2 results are linked to those of BIPM.RI(I)-K4 Absorbed dose to water, radiation type: Co-60
|
Pilot institute |
IAEA
International Atomic Energy Agency IAEA - International Organization |
Contact person | I. Csete +43 2600 28328 |
Pilot laboratory | |
---|---|
IAEA |
International Atomic Energy Agency, IAEA - International Organization, EURAMET |
ARPANSA |
Australian Radiation Protection and Nuclear Safety Agency, Australia, APMP |
BEV |
Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET |
BIM |
Bulgarian Institute of Metrology, Bulgaria, EURAMET |
CIEMAT |
Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Spain, EURAMET |
CMI |
Czech Metrology Institute, Czechia, EURAMET |
CNEA |
National Commission of Atomic Energy, Argentina, SIM |
ENEA-INMRI |
Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, Italy, EURAMET |
GUM |
Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET |
HIRCL |
Helenic Ionizing Radiation Calibration Laboratory, Greece, EURAMET |
IRB |
Rudjer Boskovic Institute; became: Secondary Standard Dosimetry Laboratory of the Ruder Boskovic Institute, HMI/IRB-SSDL, Croatia, EURAMET |
ITN-LMRIR |
Institute for Nuclear Technology, Portugal, EURAMET |
LNE-LNHB |
Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET |
LNMRI |
Laboratorio Nacional de Metrologia das Radiaçoes Ionizantes, IRD, Brazil, SIM |
METAS |
Federal Institute of Metrology, Switzerland, EURAMET |
MKEH |
Hungarian Trade Licensing Office; now BFKH, Hungary, EURAMET |
NIST |
National Institute of Standards and Technology, United States, SIM |
NPL |
National Physical Laboratory, United Kingdom, EURAMET |
NRC |
National Research Council, Canada, SIM |
NRPA |
Norvegian Radiation Protection Authority, Norway, EURAMET |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
RMTC |
Radiation Metrology and Testing Centre, Latvia, EURAMET |
SMU |
Slovensky Metrologicky Ustav, Slovakia, EURAMET |
SSM |
Swedish Radiation Safety Authority, Sweden, EURAMET |
STUK |
Radiation and Nuclear Safety Authority, Finland, EURAMET |
VSL |
VSL, Netherlands, EURAMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
BIPM.RI(I)-K4, SIM.RI(I)-K4, EUROMET.RI(I)-K4, EURAMET.RI(I)-K4.2, EURAMET.RI(I)-K4.1 and APMP.RI(I)-K4
MEASURAND Absorbed dose to water
Radiation quality Co-60 beam
The key comparison reference value is the BIPM evaluation of absrobed dose to water.
BIPM.RI(I)-K4
The degree of equivalence of each laboratory i with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy:
the relative difference Di and Ui, its expanded uncertainty (k = 2), with Ui = 2ui.
When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms both expressed in mGy/Gy:
Dij = Di - Dj and Uij , its expanded uncertainty (k = 2). The approximation for Uij is explained in section 5 of the Final Report.
Linking SIM.RI(I)-K4 to BIPM.RI(I)-K4
The relative value xi is the comparison result for laboratory i participant in SIM.RI(I)-K4 and was normalized to the value of NRC acting as the linking laboratory (see SIM.RI(I)-K4 Final Report).
The degree of equivalence of each laboratory i participant in SIM.RI(I)-K4 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).
Linking EUROMET.RI(I)-K4 to BIPM.RI(I)-K4
The relative value xi is the comparison result for laboratory i participant in EUROMET.RI(I)-K4 and is linked to the BIPM.RI(I)-K4 via the nine linking laboratories ARPANSA, BEV, ENEA-INMRI, LNE-LNHB, METAS, MKEH, NRC, PTB and VSL (see EUROMET.RI(I)-K4 Final Report).
The degree of equivalence of each laboratory i participant in EUROMET.RI(I)-K4 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).
Linking EURAMET.RI(I)-K4.1 to BIPM.RI(I)-K4
The linking of EURAMET.RI(I)-K4.1 to BIPM.RI(I)-K4 was made via BEV, LNE-LNHB and METAS, having participated in both comparisons.
The degree of equivalence of each laboratory i participant in EURAMET.RI(I)-K4.1 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).
Linking EURAMET.RI(I)-K4.2 to BIPM.RI(I)-K4
The linking of EURAMET.RI(I)-K4.2 to BIPM.RI(I)-K4 was made via BEV having participated in both comparisons.
The degree of equivalence of each laboratory i participant in EURAMET.RI(I)-K4.2 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).
Linking APMP.RI(I)-K4 to BIPM.RI(I)-K4
The linking of APMP.RI(I)-K4 to BIPM.RI(I)-K4 was made via ARPANSA, NMIJ-AIST and PTB having participated in both comparisons.
The degree of equivalence of each laboratory i participant in APMP.RI(I)-K4 with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy: Di = (xi - 1) x 1000 and Ui, its expanded uncertainty (k = 2).
Following decision of the CCRI in 2011, the pair-wise degrees of equivalence are not explicitely published in the KCDB.
BIPM.RI(I)-K4, SIM.RI(I)-K4, EUROMET.RI(I)-K4, EURAMET.RI(I)-K4.2, EURAMET.RI(I)-K4.1 and APMP.RI(I)-K4
MEASURAND Absorbed dose to water
Radiation quality Co-60 beam
BIPM.RI(I)-K4 - Laboratory measurements
xi : ratio of the measurement made at the BIPM by laboratory i to that of the BIPM
ui : combined standard uncertainty of the ratio xi
SIM.RI(I)-K4 - Laboratory measurements
xi : result of measurement carried out at laboratory i converted to absorbed dose through the NRC measurements at the BIPM (See SIM.RI(I)-K4 Final Report)
ui : combined standard uncertainty of xi
EUROMET.RI(I)-K4 - Laboratory measurements
xi : result of measurement carried out at laboratory i converted to absorbed dose through nine NMI measurements at the BIPM (See EUROMET.RI(I)-K4 Final Report)
The linking laboratories are MKEH, LNE-LNHB, PTB, ENEA, ARPANSA, NRC, BEV, METAS and VSL.
ui : combined standard uncertainty of xi
EURAMET.RI(I)-K4.2 - Laboratory measurements
xi : result of measurement carried out at laboratory i linked via BEV, LNE-LNHB and METAS.
ui : combined standard uncertainty of xi
EURAMET.RI(I)-K4.1 - Laboratory measurements
xi : result of measurement carried out at laboratory i linked via BEV.
ui : combined standard uncertainty of xi
APMP.RI(I)-K4 - Laboratory measurements
xi : result of measurement carried out at laboratory i linked via ARPANSA, NMIJ-AIST and PTB.
ui : combined standard uncertainty of xi