BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         100 kV

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.

Air kerma at 100 kV
Di  (mGy/Gy)
Participating institutes

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         135 kV

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.

Air kerma at 135 kV
Di  (mGy/Gy)
Participating institutes

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         180 kV

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.

Air kerma at 180 kV
Di  (mGy/Gy)
Participating institutes

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         250 kV

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.

Air kerma at 250 kV
Di  (mGy/Gy)
Participating institutes

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         100 kV

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.

LABi Di Uneg,i Upos,i
mGy/Gy
mGy/Gy
mGy/Gy
MKEH -0.4 7.0
PTB 2.7 5.2
ENEA 3.9 6.2
BEV 3.2 6.4
NRC 3.1 6.6
NMIJ -0.8 6.2
VSL -1.0 6.4
NIST -2.2 7.8
NIM 7.2 6.2
NPL 0.4 6.8
KRISS 4.1 4.4
LNE-LNHB 0.5 7.6
VNIIM 0.5 3.8
GUM 6.9 6.0
ARPANSA 3.8 8.2
INER 3.7 4.6
Nuc. Malaysia 14.2 12.0
DMSc -3.1 11.8
BARC 8.5 13.8
NMISA 4.5 5.6
KRISS -8.4 5.2
IAEA 4.3 7.4
CNEA -6.0 14.3
LMNRI/IRD -9.5 12.1
ININ -9.3 16.1
Participating institutes
BIPM.RI(I)-K3
APMP.RI(I)-K3
SIM.RI(I)-K3

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         135 kV

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.

LABi Di Uneg,i Upos,i
mGy/Gy
mGy/Gy
mGy/Gy
MKEH 0.9 7.0
PTB 4.5 5.2
ENEA 4.2 6.2
BEV 4.7 6.4
NRC 2.3 6.6
NMIJ -1.4 6.2
VSL -0.4 6.4
NIST -3.3 7.8
NIM 5.4 6.2
NPL 0.0 6.8
KRISS 5.1 4.4
LNE-LNHB -0.5 7.6
VNIIM 1.0 3.8
GUM 3.2 6.0
ARPANSA 5.2 8.2
INER 4.3 4.6
Nuc. Malaysia 16.2 12.0
DMSc 4.2 11.8
NMISA 2.0 5.6
KRISS 1.1 5.2
IAEA 9.2 7.4
CNEA 1.1 14.3
LMNRI/IRD -9.4 12.1
ININ -12.1 16.1
Participating institutes
BIPM.RI(I)-K3
APMP.RI(I)-K3
SIM.RI(I)-K3

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         180 kV

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.

LABi Di Uneg,i Upos,i
mGy/Gy
mGy/Gy
mGy/Gy
MKEH 0.4 7.0
PTB 4.9 5.2
ENEA 7.3 6.2
BEV 4.1 6.4
NRC 1.3 6.6
NMIJ -2.4 6.2
VSL 0.0 6.4
NIST -2.7 7.8
NIM 6.1 6.2
NPL -2.5 6.8
KRISS 4.3 4.4
LNE-LNHB -1.0 7.6
VNIIM 1.7 3.8
GUM 3.6 6.0
ARPANSA 5.4 8.2
INER 6.0 4.6
Nuc. Malaysia 15.0 12.0
DMSc 9.6 11.8
NMISA 4.8 5.6
KRISS 6.6 5.2
IAEA 13.1 7.4
CNEA 2.1 14.3
LMNRI/IRD -8.0 12.1
ININ -11.1 16.1
Participating institutes
BIPM.RI(I)-K3
APMP.RI(I)-K3
SIM.RI(I)-K3

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         250 kV

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.

LABi Di Uneg,i Upos,i
mGy/Gy
mGy/Gy
mGy/Gy
MKEH 0.4 7.0
PTB 5.5 5.2
ENEA 5.6 6.2
BEV 1.1 6.4
NRC 0.4 6.6
NMIJ -3.7 6.2
VSL -2.1 6.4
NIST -5.8 7.8
NIM 6.0 6.2
NPL -4.4 6.8
KRISS 2.8 4.4
LNE-LNHB -2.8 7.6
VNIIM 2.2 3.8
GUM 2.7 6.0
ARPANSA 4.5 8.2
INER 5.5 4.6
Nuc. Malaysia 15.9 12.0
DMSc 13.0 11.8
BARC 14.8 13.8
NMISA 7.5 5.6
KRISS 7.6 5.2
IAEA 14.0 7.4
CNEA 1.4 14.3
LMNRI/IRD -8.5 12.1
ININ -12.00 16.1
Participating institutes
BIPM.RI(I)-K3
APMP.RI(I)-K3
SIM.RI(I)-K3
Comparison
Comparison close
BIPM comparison
Linked comparison
APMP.RI(I)-K3
Metrology area, Sub-field Ionizing Radiation, Section I (x and gamma rays, electrons)
Description Measurement of air kerma for medium energy X-rays
Time of measurements 2003
Status Approved for equivalence
Final Reports of the comparisons
Measurand Air kerma
Parameters Radiation type: X-rays
Radiation quality: 100 kV to 250 kV
Transfer device Two commercial cavity ionization chambers
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by APMP (Asia Pacific Metrology Programme)
Comments

Results published on 02 September 2008
APMP.RI(I)-K3 results are linked to those of BIPM.RI(I)-K3

Pilot institute NRSL/INER
Institute of Nuclear Energy Research
Chinese Taipei
Contact person Jeng-Hung Lee

+886 3 471 1400 ext 7672
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NRSL/INER

Institute of Nuclear Energy Research, Chinese Taipei, APMP

AECS-NRML

Atomic Energy Commission of Syria, Syrian Arab Republic, APMP

ARPANSA

Australian Radiation Protection and Nuclear Safety Agency, Australia, APMP

BARC

Bhabha Atomic Research Centre, India, APMP

DMSc

Department of Medical Sciences, Thailand, APMP

IAEA

International Atomic Energy Agency, IAEA - International Organization, EURAMET

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

Nuclear Malaysia

Malaysian Nuclear Agency, Malaysia, APMP

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         100 kV

 

For a comparison with laboratory , the key comparison reference value is the BIPM air kerma determination xR,i and its standard uncertainty uR,.

 

BIPM.RI(I)-K3

The degree of equivalence of each laboratory i with respect to the key comparison reference value xR,i  is given by a pair of terms: 
Di and Ui = 2ui , its expanded uncertainty (k = 2), both expressed in mGy/Gy:

When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms:

Dij = Di - Dj  and Uij = 2uij , its expanded uncertainty (k = 2), both expressed in mGy/Gy. In evaluating uij , account should be taken of correlation between uij and uj (see the BIPM.RI(I)-K3 Summary Report Metrologia 2003 40 06036). For regional comparisons involving secondary standards, this includes correlation related to traceability (see the APMP.RI(I)-K3 Final Report, and SIM.RI(I)-K3 Final Report).

 

Linking APMP.RI(I)-K3 to BIPM.RI(I)-K3

The results of APMP.RI(I)-K3 were linked to those of BIPM.RI(I)-K3 via PTB (Germany) and ARPANSA (Australia) who participated in both comparisons.

 

Linking SIM.RI(I)-K3 to BIPM.RI(I)-K3

The results of SIM.RI(I)-K3 were linked to those of BIPM.RI(I)-K3 via NIST (United States) who participated in both comparisons.

 

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         135 kV

 

For a comparison with laboratory , the key comparison reference value is the BIPM air kerma determination xR,i and its standard uncertainty uR,.

 

BIPM.RI(I)-K3

The degree of equivalence of each laboratory i with respect to the key comparison reference value xR,i  is given by a pair of terms: 
Di and Ui = 2ui , its expanded uncertainty (k = 2), both expressed in mGy/Gy:

When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms:

Dij = Di - Dj  and Uij = 2uij , its expanded uncertainty (k = 2), both expressed in mGy/Gy. In evaluating uij , account should be taken of correlation between uij and uj (see the BIPM.RI(I)-K3 Summary Report Metrologia 2003 40 06036). For regional comparisons involving secondary standards, this includes correlation related to traceability (see the APMP.RI(I)-K3 Final Report, and SIM.RI(I)-K3 Final Report).

 

Linking APMP.RI(I)-K3 to BIPM.RI(I)-K3

The results of APMP.RI(I)-K3 were linked to those of BIPM.RI(I)-K3 via PTB (Germany) and ARPANSA (Australia) who participated in both comparisons.

 

Linking SIM.RI(I)-K3 to BIPM.RI(I)-K3

The results of SIM.RI(I)-K3 were linked to those of BIPM.RI(I)-K3 via NIST (United States) who participated in both comparisons.

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         180 kV

 

For a comparison with laboratory , the key comparison reference value is the BIPM air kerma determination xR,i and its standard uncertainty uR,.

 

BIPM.RI(I)-K3

The degree of equivalence of each laboratory i with respect to the key comparison reference value xR,i  is given by a pair of terms: 
Di and Ui = 2ui , its expanded uncertainty (k = 2), both expressed in mGy/Gy:

When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms:

Dij = Di - Dj  and Uij = 2uij , its expanded uncertainty (k = 2), both expressed in mGy/Gy. In evaluating uij , account should be taken of correlation between uij and uj (see the BIPM.RI(I)-K3 Summary Report Metrologia 2003 40 06036). For regional comparisons involving secondary standards, this includes correlation related to traceability (see the APMP.RI(I)-K3 Final Report, and SIM.RI(I)-K3 Final Report).

 

Linking APMP.RI(I)-K3 to BIPM.RI(I)-K3

The results of APMP.RI(I)-K3 were linked to those of BIPM.RI(I)-K3 via PTB (Germany) and ARPANSA (Australia) who participated in both comparisons.

 

Linking SIM.RI(I)-K3 to BIPM.RI(I)-K3

The results of SIM.RI(I)-K3 were linked to those of BIPM.RI(I)-K3 via NIST (United States) who participated in both comparisons.

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         250 kV

 

For a comparison with laboratory , the key comparison reference value is the BIPM air kerma determination xR,i and its standard uncertainty uR,.

 

BIPM.RI(I)-K3

The degree of equivalence of each laboratory i with respect to the key comparison reference value xR,i  is given by a pair of terms: 
Di and Ui = 2ui , its expanded uncertainty (k = 2), both expressed in mGy/Gy:

When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms:

Dij = Di - Dj  and Uij = 2uij , its expanded uncertainty (k = 2), both expressed in mGy/Gy. In evaluating uij , account should be taken of correlation between uij and uj (see the BIPM.RI(I)-K3 Summary Report Metrologia 2003 40 06036). For regional comparisons involving secondary standards, this includes correlation related to traceability (see the APMP.RI(I)-K3 Final Report, and SIM.RI(I)-K3 Final Report).

 

Linking APMP.RI(I)-K3 to BIPM.RI(I)-K3

The results of APMP.RI(I)-K3 were linked to those of BIPM.RI(I)-K3 via PTB (Germany) and ARPANSA (Australia) who participated in both comparisons.

 

Linking SIM.RI(I)-K3 to BIPM.RI(I)-K3

The results of SIM.RI(I)-K3 were linked to those of BIPM.RI(I)-K3 via NIST (United States) who participated in both comparisons.

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         100 kV

 

uLabi  relative standard uncertainty of air kerma determined by Lab

xi   air kerma determined by Labi at the BIPM, relative to the reference value determined by the BIPM standard

ui   combined standard uncertainty of xi taking correlation into account

 

BIPM.RI(I)-K3

 

APMP.RI(I)-K3

 

SIM.RI(I)-K3

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         135 kV

 

uLabi  relative standard uncertainty of air kerma determined by Lab

xi   air kerma determined by Labi at the BIPM, relative to the reference value determined by the BIPM standard

ui   combined standard uncertainty of xi taking correlation into account

 

BIPM.RI(I)-K3

 

APMP.RI(I)-K3

 

SIM.RI(I)-K3

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         180 kV

 

uLabi  relative standard uncertainty of air kerma determined by Lab

xi   air kerma determined by Labi at the BIPM, relative to the reference value determined by the BIPM standard

ui   combined standard uncertainty of xi taking correlation into account

 

BIPM.RI(I)-K3

 

 

APMP.RI(I)-K3

 

SIM.RI(I)-K3

BIPM.RI(I)-K3, APMP.RI(I)-K3 and SIM.RI(I)-K3

 

MEASURAND                 Air kerma

Radiation quality         250 kV

 

uLabi  relative standard uncertainty of air kerma determined by Lab

xi   air kerma determined by Labi at the BIPM, relative to the reference value determined by the BIPM standard

ui   combined standard uncertainty of xi taking correlation into account

 

BIPM.RI(I)-K3

 

APMP.RI(I)-K3

SIM.RI(I)-K3