Comparison
Comparison close
Results

EURAMET.RI(II)-K2.Ho166

 

MEASURAND        Activity concentration of a Ho-166 solution

NOMINAL VALUE      2600 kBq/g

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/g.

Ho-166
Di /(kBq/g)
Participating institutes
Comparison
Comparison close
Results

EURAMET.RI(II)-K2.Ho166

 

MEASURAND        Activity concentration of a Ho-166 solution

NOMINAL VALUE      2600 kBq/g

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/g.

LABi Di Uneg,i Upos,i
kBq/g
kBq/g
kBq/g
CMI -0.8 5.2
ENEA 46.5 74.1
LNE-LNHB 2.4 16.3
NPL 0.5 15.4
Participating institutes
EURAMET.RI(II)-K2.Ho-166
Comparison
Comparison close
RMO comparison
Linked comparison
EURAMET.RI(II)-K2.Ho-166
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity concentration of the same Ho-166 solution
Time of measurements 2017 - 2018
Status Approved
Final Reports of the comparisons
References

EURAMET RI(II)-K2 Ho-166 Technical Protocol

Measurand Activity concentration (Bq/g)
Transfer device Samples of Ho-166 as Ho(NO33)3 in 1 M HNO3 in flame-sealed SIR-type ampoules.
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET 1435
Pilot institute CMI
Czech Metrology Institute
Czechia
Contact person J Solc

+420 266 020 487
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
CMI

Czech Metrology Institute, Czechia, EURAMET

ENEA-INMRI

Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, Italy, EURAMET

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

NPL

National Physical Laboratory, United Kingdom, EURAMET

Comparison
Comparison close
Results

EURAMET.RI(II)-K2.Ho166

 

MEASURAND        Activity concentration of a Ho-166 solution

NOMINAL VALUE      2600 kBq/g

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from a power-moderated mean method based on the reported results, ENEA excluded.

 

        xR = 2623.6 kBq/g     uR = 3.6 kBq/g

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in kBq/g:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

Comparison
Comparison close
Results

EURAMET.RI(II)-K2.Ho166

 

MEASURAND        Activity concentration of a Ho-166 solution

NOMINAL VALUE      2600 kBq/g

 

xi   activité per mass as reported by laboratory iLabi

u   combined standard uncertainty of x

       
  Labi xi (xi)
  kBq / g kBq / g
  CMI 2622.8 4.4
  ENEA 2670.1 37
  LNE-LNHB 2626.0 8.9
  NPL 2624.1 8.5