Comparison
Comparison close
Results
Comparison
Comparison close
Results
Comparison
Comparison close
RMO comparison
Linked comparison
SIM.M.D-S6
Metrology area, Sub-field Mass and related quantities, Density
Description High-accuracy hydrometers
Time of measurements 2017 - 2018
Status Approved
Final Reports of the comparisons
References

SIM.M.D-S6 Technical Protocol

Measurand Density from 600 kg/m3 to 1300 kg/m3
Parameters Temperature at 20 °C
Transfer device Four high-accuracy hydrometers (0.1 kg/m3 resolution).
Nominal values: (600 - 610; 990 - 1000; 1090 - 1100; 1290 - 1300) kg/m3
Comparison type Supplementary Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by SIM (Inter-American Metrology System)
Pilot institute CENAM
Centro Nacional de Metrologia
Mexico
Contact person Luis Omar BECERRA

+52 442 211 05 73
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
CENAM

Centro Nacional de Metrologia, Mexico, SIM

CESMEC

Centro de Estudios, Medición y Certificación de Calidad, Chile, SIM

IBMETRO

Instituto Boliviano de Metrología, Bolivia, SIM

INEN

Servicio Ecuatoriano de Normalización, Ecuador, SIM

INM (CO)

Instituto Nacional de Metrología de Colombia, Colombia, SIM

INTI

Instituto Nacional de Tecnologia Industrial, Argentina, SIM

LACOMET

Laboratorio Costarricense de Metrologia, Costa Rica, SIM

LATU

Laboratorio Tecnologico del Uruguay, Uruguay, SIM

NIMT

National Institute of Metrology (Thailand), Thailand, APMP

SASO-NMCC

National Measurement and Calibration Center - Saudi Standards, Metrology and Quality Organization, Saudi Arabia, GULFMET

Comparison
Comparison close
Results
Comparison
Comparison close
Results