MEASURAND : Mass fraction of nitrogen in dry milk powder

Degrees of equivalence : offset Di and expanded uncertainty (k = 2) Ui expressed in %

MEASURAND : Mass fraction of nitrogen in dry milk powder

Degrees of equivalence : offset Di and expanded uncertainty (k = 2) Ui expressed in %

Results are presented under A4 printable format in Summary Results (PDF file).

Comparison
Comparison close
CC comparison
Linked comparison
CCQM-K149
Metrology area, Sub-field Chemistry and Biology, Inorganics
Description Nitrogen in milk powder
Time of measurements 2017 - 2018
Status Approved for equivalence
Final Reports of the comparisons
Measurand Mass fraction of nitrogen (1 % to 6 %)
Transfer device Sample of dry milk powder with mass 100 g that is packed into special water-proof double polyethylene packages hermetically sealed with double seam.
Comparison type Key Comparison
Consultative Committee CCQM (Consultative Committee for Amount of Substance)
Conducted by CCQM (Consultative Committee for Amount of Substance)
Comments

Results published on 01 April 2019

Mass fraction of nitrogen in dry milk powder

Pilot institute UNIIM
Ural Scientific and Research Institute for Metrology, Rosstandart
Russian Federation
Contact person M. Krasheninina

+7 (343) 350 60 63, 355 02 63
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
CENAM

Centro Nacional de Metrologia, Mexico, SIM

GL

Government Laboratory, Hong Kong, China, APMP

INACAL

National Institute for Quality, Peru, SIM

INM (CO)

Instituto Nacional de Metrología de Colombia, Colombia, SIM

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

INTI

Instituto Nacional de Tecnologia Industrial, Argentina, SIM

KazStandard

Kazakhstan Institute of Standardization and Metrology, Kazakhstan, COOMET

RISE

Research Institutes of Sweden AB, Sweden, EURAMET

SE "Ukrmetrteststandard"

State Enterprise "Ukrmetrteststandard", Ukraine, EURAMET

MEASURAND : Mass fraction of nitrogen in dry milk powder

The key comparison reference value xR was calculated as the median of the participants' results excluding KazInMetr's results, and its associated uncertainty uR is calculated according to Equation 8 on the page 15 of the Final report.

xR = 2.887 %, uR = 0.011 %
 

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms both expressed in %: Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), where Ui = 2(ui2 + uR2)1/2.
No pair-wise degrees of equivalence are computed for this comparison.

MEASURAND : Mass fraction of nitrogen in dry milk powder

xi : result of measurement carried out by laboratory i
ui : combined standard uncertainty of xi