Metrology area, Sub-field | Ionizing Radiation, Section I (x and gamma rays, electrons) |
Description | Comparison of air kerma standards |
Time of measurements | 2004 - 2005 |
Status | Approved |
Final Reports of the comparisons | |
Measurand | Air kerma |
Parameters | ISO 4037 radiation qualities |
Transfer device | Three ionisation chambers |
Comparison type | Supplementary Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | EURAMET (European Association of National Metrology Institutes) |
RMO Internal Identifier | EUROMET Project No 545 |
Comments | Radiation qualities: ISO 4037-1, Narrow spectra series N 30 to N 300. Minimum: 30 kV, 60 kV, 100 kV, 200 kV and 300 kV, preferentially all |
Pilot institute |
PTB
Physikalisch-Technische Bundesanstalt Germany |
Contact person | Ludwig BUERMANN +49 (0) 531 592 6620 |
Pilot laboratory | |
---|---|
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
ARPANSA |
Australian Radiation Protection and Nuclear Safety Agency, Australia, APMP |
BEV |
Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET |
NIST |
National Institute of Standards and Technology, United States, SIM |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
NMi-VSL |
Nederlands Meetinstituut - Van Swinden Laboratorium (became VSL in 2009), Netherlands, EURAMET |
NPL |
National Physical Laboratory, United Kingdom, EURAMET |
NRSL/INER |
Institute of Nuclear Energy Research, Chinese Taipei, APMP |
OMH |
Országos Mérésügyi Hivatal; Subsequently MKEH, now BKFH, Hungary, EURAMET |
STUK |
Radiation and Nuclear Safety Authority, Finland, EURAMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).