BIPM.RI(I)-K8 and APMP.RI(I)-K8
MEASURAND Air kerma
SOURCE HDR Ir-192 brachytherapy sources
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.
a) results obtained using a thimble-type chamber
b) results obtained using a well-type chamber
BIPM.RI(I)-K8 and APMP.RI(I)-K8
MEASURAND Air kerma
SOURCE HDR Ir-192 brachytherapy sources
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mGy/Gy.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mGy/Gy |
mGy/Gy |
mGy/Gy |
||
VSL (a) | -12.7 | 12.2 | ||
NPL (a) | -1.1 | 11.4 | ||
PTB (a) | 0.3 | 19.8 | ||
NRC (a) | -3.4 | 10.0 | ||
NMIJ (b) | 3.6 | 10.8 | ||
IAEA | 9.4 | 30.5 | ||
INER | 10.1 | 19.6 | ||
KRISS | 4.2 | 16.4 | ||
NMISA | 7.3 | 30.5 | ||
Nuclear Malaysia | -6.4 | 30.8 | ||
PTKMR-BATAN | 40.8 | 34.3 |
a) results obtained using a thimble-type chamber
b) results obtained using a well-type chamber
Metrology area, Sub-field | Ionizing Radiation, Section I (x and gamma rays, electrons) |
Description | Measurement of reference air kerma rate for HDR Ir-192 brachytherapy sources |
Time of measurements | 2009 - |
Status | Continuous, approved for equivalence |
References | |
Measurand | Reference air kerma rate (in mGy/h) relative to the BIPM evaluation |
Parameters | Radiation type: HDR Ir-192 sealed brachytherapy sources |
Transfer device | BIPM ionization chambers: well-type and thimble-type |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | BIPM (Bureau International des Poids et Mesures) |
Comments | The Final Report for each bi-lateral comparsion is listed in the tab "Participants" First results published on 13 September 2014 Most recent update: 20 October 2021 |
Pilot institute |
BIPM
Bureau International des Poids et Mesures BIPM - International Organization |
Contact person | C. Kessler +33 (0) 1 45 07 70 84 |
Pilot laboratory | |
---|---|
BIPM |
Bureau International des Poids et Mesures, BIPM - International Organization, N/A |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
2015 |
NPL |
National Physical Laboratory, United Kingdom, EURAMET |
2010 |
NRC |
National Research Council, Canada, SIM |
2014 |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
2011 |
VSL |
VSL, Netherlands, EURAMET |
2009 |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
BIPM.RI(I)-K8 and APMP.RI(I)-K8
MEASURAND Air kerma
SOURCE HDR Ir-192 brachytherapy sources
The key comparison reference value is the BIPM evaluation of reference air kerma rate.
The degree of equivalence of each laboratory i with respect to the key comparison refernce value is given by a pair of terms both expressed in mGy/Gy: the relative difference Di and Ui = 2 ui, its expanded uncertainty (k = 2).
When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms, both expressed in mGy/Gy: Dij = Di - Dj and Uij, its expanded uncertainty (k = 2).
Linking APMP.RI(I)-K8 to BIPM.RI(I)-K8
The INSTITUTE provides the link between the key comparison COMPARISON IDENTIFIER and COMPARISON IDENTIFIER, having participated in both comparisons. The linking process is described in Section X of the COMPARISON IDENTIFIER Final Report.
BIPM.RI(I)-K8 and APMP.RI(I)-K8
MEASURAND Air kerma
SOURCE HDR Ir-192 brachytherapy sources
xi measurement result obtained by laboratory i relative to the KCRV
ui combined standard uncertainty of xi
BIPM.RI(I)-K8
APMP.RI(I)-K8