CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (SE)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   40 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (SE)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   80 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (SE)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   160 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (SE)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   800 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (SE)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   2 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (SE)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   5 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (SE)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   20 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (SE)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   40 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   80 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   160 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   800 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   2 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   5 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   20 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

Magnitude (BB)
Di / fC/(m/s2)
Participating institutes

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.07 0.09
HBK-DPLA -0.17 0.50
METAS -0.02 0.28
CEM -0.11 0.50
NIST -0.06 0.37
CENAM 0.43 0.37
NMIJ 0.19 0.88
NIM 0.27 0.50
NMIA 0.07 0.37
INMETRO 0.10 0.24
NMISA 0.08 0.37
VNIIM -0.03 0.63
NMC A*Star 0.00 0.51
SE UkrMet -0.19 0.63
INRIM -0.05 0.51
UME 0.48 0.64
CMI 0.24 0.64
GUM 0.17 0.63
LNE -0.03 0.38
SASO-NMCC 0.16 0.76
RISE 0.33 0.51
NIMT 0.01 0.53
NPLI -0.12 1.32
INTI 0.34 1.57
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   40 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.08 0.10
HBK-DPLA -0.16 0.50
METAS 0.01 0.23
CEM -0.02 0.50
NIST 0.00 0.37
CENAM 0.05 0.37
NMIJ 0.06 0.50
NIM 0.34 0.50
NMIA 0.10 0.37
INMETRO 0.11 0.24
NMISA 0.08 0.37
VNIIM 0.03 0.63
NMC A*Star 0.13 0.37
SE UkrMet -0.09 0.63
INRIM -0.09 0.51
UME 0.21 0.38
CMI 0.35 0.38
GUM 0.07 0.65
LNE 0.07 0.38
SASO-NMCC 0.06 0.65
RISE 0.17 0.52
NIMT 0.11 0.53
NPLI -0.09 1.32
INTI 0.04 1.57
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   80 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.06 0.10
HBK-DPLA -0.16 0.50
METAS 0.02 0.23
CEM -0.05 0.50
NIST 0.02 0.37
CENAM -0.02 0.37
NMIJ 0.06 0.50
NIM 0.29 0.50
NMIA 0.09 0.37
INMETRO 0.12 0.24
NMISA 0.10 0.37
VNIIM 0.02 0.63
NMC A*Star 0.10 0.37
SE UkrMet -0.14 0.63
INRIM -0.04 0.51
UME 0.17 0.38
CMI 0.27 0.38
GUM 0.07 0.65
LNE 0.06 0.38
SASO-NMCC 0.02 0.65
RISE 0.15 0.52
NIMT 0.08 0.53
NPLI -0.10 1.32
INTI -0.06 0.74
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   160 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.08 0.10
HBK-DPLA -0.13 0.50
METAS 0.02 0.25
CEM -0.04 0.50
NIST -0.13 0.37
CENAM 0.00 0.37
NMIJ 0.14 0.50
NIM 0.35 0.50
NMIA 0.10 0.37
INMETRO 0.15 0.24
NMISA -0.03 0.37
VNIIM 0.13 0.63
NMC A*Star 0.24 0.37
SE UkrMet -0.06 0.63
INRIM -0.23 0.51
UME 0.15 0.38
CMI 0.25 0.38
GUM 0.05 0.65
LNE 0.09 0.38
SASO-NMCC 0.10 0.65
RISE 0.10 0.52
NIMT -0.18 0.53
NPLI -0.37 1.32
INTI -0.30 0.74
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   800 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.09 0.10
HBK-DPLA -0.14 0.50
METAS -0.05 0.28
CEM -0.06 0.50
NIST -0.03 0.37
CENAM 0.03 0.37
NMIJ 0.07 0.37
NIM 0.36 0.50
NMIA 0.11 0.37
INMETRO 0.15 0.24
NMISA 0.07 0.37
VNIIM 0.09 0.63
NMC A*Star 0.15 0.37
SE UkrMet -0.15 0.63
INRIM -0.08 0.51
UME 0.12 0.38
CMI 0.27 0.38
GUM 0.03 0.65
LNE 0.03 0.38
SASO-NMCC 0.00 0.65
RISE 0.12 0.52
NIMT 0.25 0.53
NPLI 0.01 1.32
INTI 0.32 0.85
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   2 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.08 0.09
HBK-DPLA -0.10 0.50
METAS -0.11 0.32
CEM -0.08 0.50
NIST -0.01 0.50
CENAM 0.07 0.37
NMIJ 0.09 0.37
NIM 0.38 0.50
NMIA 0.08 0.37
INMETRO 0.15 0.24
NMISA 0.05 0.63
VNIIM 0.29 0.63
NMC A*Star 0.11 0.51
SE UkrMet -0.24 0.88
INRIM -0.07 0.64
UME 0.05 0.65
CMI 0.26 0.65
GUM -0.02 0.65
LNE 0.02 0.38
SASO-NMCC -0.02 0.78
RISE 0.23 0.52
NIMT 0.18 0.76
NPLI -0.58 1.41
INTI 0.02 2.01
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   5 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB 0.00 0.09
HBK-DPLA -0.09 0.77
METAS -0.17 0.30
CEM 0.02 0.51
NIST 0.12 0.51
CENAM -0.52 0.38
NMIJ 0.23 0.64
NIM 0.49 0.51
NMIA 0.32 0.38
INMETRO 0.12 0.77
NMISA 0.00 1.03
VNIIM -0.28 1.29
NMC A*Star 0.15 0.70
SE UkrMet -0.27 0.90
INRIM -0.08 0.78
UME 0.25 0.67
CMI 0.37 0.67
GUM -0.17 0.80
LNE -0.12 0.80
SASO-NMCC 0.24 0.80
RISE 0.27 0.80
NIMT 0.30 1.12
NPLI -2.84 1.63
INTI -0.14 2.21
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   20 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB 0.31 1.86
HBK-DPLA -4.52 3.94
METAS -7.51 3.64
CEM 1.29 4.06
NIST 2.07 4.08
CENAM 2.08 1.88
NMIJ 1.71 5.13
NIM 0.63 2.98
NMIA 4.27 3.29
INMETRO -0.67 2.30
NMISA -1.20 3.59
VNIIM -3.85 3.96
NMC A*Star -0.16 2.79
SE UkrMet -5.31 5.98
INRIM 7.93 2.15
UME 9.09 5.53
CMI 2.82 5.35
SASO-NMCC 9.28 5.53
NIMT 5.74 4.86
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.01 0.10
HBK-DPLA -0.10 0.50
METAS -0.21 0.28
CEM 0.15 0.51
NIST -0.04 0.37
CENAM 0.20 0.38
NMIJ 0.01 0.50
NIM 0.26 0.51
NMIA -0.10 0.37
INMETRO 0.12 0.24
NMISA 0.04 0.37
VNIIM 0.29 0.64
NMC A*Star -0.18 0.52
SE UkrMet -0.25 0.63
INRIM -0.23 0.51
UME 0.20 0.64
CMI 0.35 0.64
GUM 0.08 0.77
LNE -0.29 0.38
SASO-NMCC 0.09 0.77
RISE 0.01 0.51
NIMT 0.00 0.52
NPLI -0.34 1.33
INTI -0.17 1.58
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   40 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.04 0.10
HBK-DPLA -0.01 0.51
METAS 0.07 0.23
CEM 0.05 0.51
NIST 0.07 0.38
CENAM 0.05 0.38
NMIJ -0.04 0.51
NIM 0.22 0.51
NMIA 0.02 0.38
INMETRO -0.03 0.24
NMISA -0.03 0.37
VNIIM 0.20 0.64
NMC A*Star -0.01 0.38
SE UkrMet -0.29 0.63
INRIM -0.12 0.51
UME 0.15 0.38
CMI 0.22 0.39
GUM 0.01 0.64
LNE 0.00 0.38
SASO-NMCC 0.20 0.64
RISE 0.00 0.51
NIMT 0.06 0.53
NPLI -0.42 1.33
INTI -0.29 1.58
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   80 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.04 0.10
HBK-DPLA -0.07 0.50
METAS 0.10 0.23
CEM 0.06 0.51
NIST 0.12 0.38
CENAM 0.06 0.38
NMIJ -0.05 0.51
NIM 0.21 0.51
NMIA 0.01 0.37
INMETRO -0.01 0.24
NMISA 0.01 0.37
VNIIM 0.16 0.64
NMC A*Star -0.04 0.37
SE UkrMet -0.24 0.63
INRIM -0.18 0.51
UME 0.03 0.38
CMI 0.18 0.38
GUM -0.03 0.64
LNE -0.01 0.38
SASO-NMCC 0.12 0.64
RISE 0.00 0.51
NIMT 0.01 0.53
NPLI -0.36 1.33
INTI -0.25 0.74
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   160 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.03 0.10
HBK-DPLA -0.06 0.50
METAS 0.07 0.24
CEM 0.07 0.51
NIST 0.10 0.38
CENAM 0.04 0.37
NMIJ 0.00 0.51
NIM 0.21 0.51
NMIA 0.04 0.37
INMETRO -0.03 0.24
NMISA -0.14 0.37
VNIIM 0.16 0.64
NMC A*Star -0.01 0.37
SE UkrMet -0.23 0.63
INRIM -0.24 0.51
UME 0.10 0.38
CMI 0.27 0.39
GUM 0.00 0.64
LNE 0.02 0.38
SASO-NMCC 0.23 0.64
RISE -0.02 0.51
NIMT -0.01 0.53
NPLI -0.43 1.33
INTI -0.26 0.74
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   800 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.02 0.10
HBK-DPLA -0.03 0.51
METAS 0.04 0.24
CEM -0.01 0.51
NIST 0.05 0.38
CENAM -0.04 0.38
NMIJ -0.06 0.38
NIM 0.23 0.51
NMIA 0.01 0.38
INMETRO -0.01 0.24
NMISA -0.04 0.38
VNIIM 0.38 0.64
NMC A*Star -0.01 0.38
SE UkrMet -0.23 0.63
INRIM -0.06 0.51
UME 0.08 0.38
CMI 0.20 0.39
GUM 0.08 0.64
LNE 0.04 0.38
SASO-NMCC 0.16 0.64
RISE -0.01 0.51
NIMT 0.09 0.53
NPLI -0.15 1.34
INTI 0.71 0.86
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   2 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.10 0.09
HBK-DPLA 0.06 0.51
METAS 0.13 0.31
CEM 0.02 0.51
NIST 0.05 0.51
CENAM 0.10 0.38
NMIJ 0.02 0.38
NIM 0.40 0.51
NMIA 0.04 0.38
INMETRO 0.09 0.24
NMISA -0.01 0.64
VNIIM 0.32 0.64
NMC A*Star 0.01 0.52
SE UkrMet -0.11 0.89
INRIM -0.16 0.64
UME -0.02 0.64
CMI 0.20 0.64
GUM 0.09 0.64
LNE 0.09 0.39
SASO-NMCC 0.10 0.77
RISE -0.01 0.51
NIMT 0.08 0.76
NPLI -0.10 1.43
INTI 0.17 2.03
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   5 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB -0.41 0.21
HBK-DPLA -0.01 0.77
METAS 0.31 0.58
CEM -0.08 0.50
NIST 0.00 0.50
CENAM -0.49 0.35
NMIJ 0.09 0.63
NIM 0.58 0.50
NMIA 0.14 0.36
INMETRO 0.07 0.77
NMISA 0.06 1.03
VNIIM -0.02 1.30
NMC A*Star -0.16 0.69
SE UkrMet -0.21 0.90
INRIM -0.22 0.78
UME -0.13 0.65
CMI 0.22 0.65
GUM 0.20 0.79
LNE 0.16 0.79
SASO-NMCC 0.05 0.78
RISE -0.05 0.78
NIMT 0.04 1.13
NPLI -0.41 1.67
INTI 0.32 2.23
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB 0.97 0.53
HBK-DPLA 0.82 0.77
METAS 1.05 2.03
CEM 0.24 1.07
NIST -0.56 0.61
CENAM -0.66 0.61
NIM 0.85 1.36
NMIA -0.44 1.49
INMETRO 0.76 1.07
NMISA 1.05 1.65
VNIIM -0.56 2.05
NMC A*Star -0.35 1.62
SE UkrMet -0.06 2.06
INRIM -0.55 0.83
UME -0.19 1.39
CMI -0.23 1.39
GUM 1.37 1.54
LNE -0.81 1.38
SASO-NMCC 4.73 1.73
RISE -0.19 1.68
NIMT 0.15 2.02
NPLI 1.32 2.70
INTI 0.98 2.69
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   20 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in fC / (m/s2).

LABi Di Uneg,i Upos,i
fC/(m/s2)
fC/(m/s2)
fC/(m/s2)
PTB 0.30 1.73
HBK-DPLA -5.97 3.82
METAS 2.23 3.92
CEM -3.66 3.68
NIST -4.34 3.66
CENAM 1.58 1.75
NIM -0.69 2.75
NMIA -0.54 2.75
INMETRO 0.41 2.15
NMISA 1.79 3.39
VNIIM -9.76 3.74
NMC A*Star 0.10 2.60
SE UkrMet -5.18 5.55
INRIM 0.24 1.92
UME -0.38 4.80
CMI 1.16 4.83
SASO-NMCC 2.90 4.88
NIMT 5.36 4.73
Participating institutes
CCAUV.V-K5
EURAMET.AUV.V-K5
AFRIMETS.AUV.V-K5

For the degrees of equivalence for phase, please see Section 8.5 of the CCAUV.V-K5 Final Report, Section 8.8 of the EURAMET.AUV.V-K5 Final Report and Section 9.5 of the AFRIMETS.AUV.V-K5 Final Report.

EURAMET.AUV.V-K5.1
Metrology area, Sub-field Acoustics, Ultrasound, Vibration, Vibration
Description Primary calibration of magnitude and phase of the complex sensitivity of accelerometers
Time of measurements 2025 - 2026
Status Protocol complete
References

Technical protocol

Measurand Magnitude of the charge sensitivity
Parameters Frequency range from 5 Hz to 10 kHz.
Acceleration amplitudes: preferably 50 m/s² to 100 m/s², a range of 2 m/s² to 200 m/s² is admissible.
Ambient temperature and accelerometer temperature during the calibration: (23 ± 2) °C
Transfer device Piezoelectric accelerometers BK 8305 (SN: 705519) back to back (BB) type and BK 4382 (SN:1671819) single ended (SE) type supplied by INRIM
Comparison type Key Comparison
Consultative Committee CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration)
Conducted by EURAMET (European Association of National Metrology Institutes)
Comments This key comparison in which INRIM (Italy), NPLI (India) and LNE (France) will participate aims at providing supporting evidence for the registration of calibration and measurement capabilities (CMCs) in the framework of the CIPM MRA.
The specific task of this comparison is to measure the magnitude of the charge sensitivity of two accelerometers (back-to-back type and single ended type) at specified frequencies with primary means i.e. according to relevant standards.
The sensitivities of the piezoelectric accelerometer BK 8305 (SN: 705519) back to back (BB) type and BK 4382 (SN:1671819) single ended (SE) type supplied by INRIM, are to be reported by each participant at agreed frequencies and acceleration amplitudes. The sensitivity reported will exclude effects from the applicable conditioning amplifier/power supply unit (PSU) used. For all instances, the participating laboratory shall provide the amplifier to be used.
Pilot institute INRIM
Istituto Nazionale di Ricerca Metrologica
Italy
Contact person Alessandro Schiavi

First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
INRIM

Istituto Nazionale di Ricerca Metrologica, Italy, EURAMET

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 126.571 fC/(m/s2)     uR = 0.085 fC/(m/s2)

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   40 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 126.561 fC/(m/s2)     uR = 0.081 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

 

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   80 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 126.565 fC/(m/s2)     uR = 0.081 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   160 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 126.532 fC/(m/s2)     uR = 0.082 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   800 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 126.636 fC/(m/s2)     uR = 0.082 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   2 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 127.097 fC/(m/s2)     uR = 0.087 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   5 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 129.886 fC/(m/s2)     uR = 0.098 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   20 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 206.819 fC/(m/s2)     uR = 0.906 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 128.001 fC/(m/s2)     uR = 0.085 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   40 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 128.007 fC/(m/s2)     uR = 0.082 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   80 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 127.984 fC/(m/s2)     uR = 0.082 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   160 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 127.958 fC/(m/s2)     uR = 0.083 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   800 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 128.028 fC/(m/s2)     uR = 0.082 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   2 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 128.342 fC/(m/s2)     uR = 0.087 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   5 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 130.614 fC/(m/s2)     uR = 0.164 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 138.981 fC/(m/s2)     uR = 0.323 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   20 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

The key comparison reference value xR and its associated standard uncertainty uR are computed from a weighted mean to which institutes who presented outlier results, or no result, did not contribute. Information of the evaluation of xR and uR is given in Section 8.1 of the CCAUV.V-K5 Final Report.

        xR = 192.170 fC/(m/s2)     uR = 0.840 fC/(m/s2)

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in fC / (m/s2).

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence as described in Section 8 of the CCAUV.V-K5 Final Report.

 

Linking EURAMET.AUV.V-K5 to CCAUV.V-K5

HBK-DPLA, PTB, CEM and METAS provided the link between EURAMET.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 8 of the EURAMET.AUV.V-K5 Final Report.

 

Linking AFRIMETS.AUV.V-K5 to CCAUV.V-K5

NMISA provided the link between AFRIMETS.AUV.V-K5 and CCAUV.V-K5, having participated in both comparisons. The linking process is described in Section 9 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   40 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   80 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   160 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   800 Hz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   2 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   5 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   20 kHz

INSTRUMENT           Single-ended (SE) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   40 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   80 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   160 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   800 Hz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   2 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   5 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   10 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.

CCAUV.V-K5, EURAMET.AUV.V-K5 and AFRIMETS.AUV.V-K5

 

MEASURAND     Magnitude of complex sensitivity

NOMINAL VALUE   20 kHz

INSTRUMENT           Back-to-back (BB) accelerometer

 

CCAUV.V-K5

Measurement results are listed in Section 7 of the CCAUV.V-K5 Final Report.

 

EURAMET.AUV.V-K5

Measurement results are listed in Section 7 of the EURAMET.AUV.V-K5 Final Report.

 

AFRIMETS.AUV.V-K5

Measurement results are listed in Section 8 of the AFRIMETS.AUV.V-K5 Final Report.